Research Article

Determination of Platycodin D and Platycodin D3 in Rat Plasma Using Liquid Chromatography-Tandem Mass Spectrometry

Table 1

Calculated concentrations of PD in calibration standards prepared in rat plasma ( at each level).

Theoretical concentration (ng/mL)SlopeIntercept
5010020050010002000500010000

Mean50.998.7191.1493.51000.22031.55027.210309.10.00008−0.0019050.9982
CV (%)3.26.15.15.14.56.33.02.7
RE (%)1.8−1.3 −4.5 −1.3 0.0 1.6 0.5 3.1

CV: coefficient of variation; RE: relative error.