Perioperative Intravascular Fluid Assessment and Monitoring: A Narrative Review of Established and Emerging Techniques

1Department of Anesthesiology and Critical Care, Ronald Reagan UCLA Medical Center, David Geffen School of Medicine at UCLA, 757 Westwood Plaza, Suite 2231, Los Angeles, CA, 90095, USA
2Division of Pulmonary and Critical Care Medicine, VA Palo Alto Health Care System, 3801 Miranda Avenue, Palo Alto, CA 94304, USA
3Department of Anesthesia, Stanford University School of Medicine, 3801 Miranda Avenue, Palo Alto, CA, 94304, USA

Received 14 December 2010; Revised 29 March 2011; Accepted 4 May 2011

Academic Editor: Maurizio Cecconi

Copyright © 2011 Sumit Singh et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Accurate assessments of intravascular fluid status are an essential part of perioperative care and necessary in the management of the hemodynamically unstable patient. Goal-directed fluid management can facilitate resuscitation of the hypovolemic patient, reduce the risk of fluid overload, reduce the risk of the injudicious use of vasopressors and inotropes, and improve clinical outcomes. In this paper, we discuss the strengths and limitations of a spectrum of noninvasive and invasive techniques for assessing and monitoring intravascular volume status and fluid responsiveness in the perioperative and critically ill patient.

1. Introduction

The ability to assess intravascular volume is an essential part of perioperative care and the management of perioperative hemodynamic instability. Insufficient intravascular volume can result in decreased oxygen delivery to tissues and organ dysfunction, while fluid overload states can contribute to the development of edema and organ dysfunction, including respiratory failure. The injudicious use of vasopressors and inotropes in the hypovolemic patient can be hazardous and increase the risk of a poor outcome.

Two concepts are relevant to assessments of fluid status in perioperative and critical care. Euvolemia describes a state of normal body fluid volume that allows adequate filling of the cardiac chambers and, in turn, makes it possible for the heart to produce a cardiac output that can meet the organism oxygen demand. In the setting of euvolemia, neither diuresis nor fluid administration is necessary. Fluid responsiveness describes the ability of the heart to respond to filling volume variations, modifying its stroke volume and consequently the cardiac output. From a patient management perspective, fluid responsiveness determines the extent to which circulatory homeostasis can be maintained with fluids alone versus the need for inotropes or vasopressors. An understanding of both concepts derives from the Frank-Starling relationship describing the changes in cardiac stroke volume in response to changes in cardiac preload. The ascending portion of the Frank-Starling curve will correspond to the fluid responsive phase of resuscitation, as seen with an increase in the cardiac output. Once the left ventricle reaches the plateau phase of the curve, fluid administration will not improve the cardiac output any further; it may lead to the adverse effects related to fluid overloading, such as hydrostatic pulmonary edema.

In a broad sense, if euvolemia is the goal of fluid use in resuscitation, then fluid responsiveness reflects the process of working toward establishing euvolemia. In evaluating the various techniques for analyzing fluid status, it is helpful to contrast their utility in predicting fluid responsiveness versus euvolemia and to consider how these relative strengths and weaknesses may be paired with different clinical situations to yield accurate and meaningful information. Methods of interpreting intravascular volume range from clinical assessments such as inspection of veins and passive leg raising, to more invasive methods such as central venous and pulmonary artery catheterization, to newer technically intensive methods such as echocardiography and analysis of flow parameters.

2. Assessment of Fluid Status by Physical Examination

Only extremes of body water can be predicted by clinical examination. Tachycardia and hypotension and loss of skin turgor are associated with frank hypovolemia in unanaesthetized adults, while tenting of skin and sunken eyes and fontanels can be found in infants. The ability to detect hypervolemia is more dependent upon cardiac pathology where smaller grades of fluid excess can produce findings such as rales, a third heart sound, oxygen desaturation, jugular vein distension, and peripheral edema. In operative patients, experienced clinicians can often predict hypovolemia from the clinical situation. Fluid deficits resulting from a patient’s nil per os (NPO) status and use of inhalational agents and positive pressure ventilation early in a case are usually indicative of relative hypovolemia, and anesthesiologists typically respond by increasing the volume of infused fluids. Blood loss and release of inflammatory mediators can also reduce blood volume, preload, and vascular tone in latter phases of an operation, when, again, additional boluses would likely be given with careful assessment of the blood pressure response.

Depending on patient position and access, examination of neck veins and passive leg raising test can yield useful information. The passive leg raising test (PLR) delivers a reversible endogenous fluid challenge by increasing venous return resulting from elevating the legs to 45 degrees in a supine patient and evaluating its effect on blood pressure and heart rate. One simple described way to perform PLR is by using the pivotal motion of the bed to transfer a 45 degree recumbent patient into a horizontal position with supine trunk and elevated legs [1]. This technique recruits the splanchnic blood in addition to the blood from lower extremities, in contrast to simply elevating legs in a supine patient. PLR test has shown good correspondence with other derived indices to predict fluid responsiveness in patients with sepsis and pancreatitis [2] and has been specifically evaluated with transthoracic echo [3] and esophageal Doppler [4] in mechanically ventilated patients. Most operative situations will preclude the use of PLR and likely prompt the clinician to assess the response to a fluid bolus. However, in evaluating patients postoperatively, this simple technique may be useful.

3. Invasive Pressure Monitoring

3.1. Central Venous Pressure (CVP)

CVP is the measurement of pressures within the thorax in the superior vena cava (SVC) and serves as a reasonable surrogate for the corresponding right atrial pressures. CVP is the most widely used technique for measuring intravascular volume in critically ill patients [5].

The CVP pressure tracing consists of three positive waves (a, c, and v) and two negative waves (x and y). The CVP is specifically measured at the “z” point of the CVP tracing which corresponds to the leading edge of the c wave (Figure 1). At this part of the cardiac cycle, the catheter tip is in continuity with the ventricle and hence affords the best estimate of cardiac preload [6]. As with the measurement of jugular venous distension, the reference point for measurement of the CVP is the midaxillary line in the fifth intercostals space. Numerical recordings of CVP are measured at end expiration, a time in the respiratory cycle where the opposing forces of lung elasticity and chest recoil are balanced and exert the least pressure on the central and pulmonary vasculature. In patients with forcible expiration, the true CVP may be better represented by a value at the start of expiration [6]. Failure to attend to these basic principles will lead to erroneous data with little reproducibility across multiple users and time points.

231493.fig.001
Figure 1: A typical CVP waveform (lower tracing) and accompanying electrocardiogram (upper). The a, c, and v waves are shown, along with the z point (arrow), indicating the appropriate time in the cardiac cycle for CVP measurement. All analyses need to occur at end expiration.

In our experience, single point estimates of CVP are of limited clinical value unless they are low (<5 mm Hg) and confirm an existing suspicion for hypovolemia. Trends and their correspondence to clinical evidence of organ function and perfusion help to create a more meaningful picture of fluid needs and euvolemia. The standard for testing volume responsiveness is to give a fluid challenge. This involves giving fluids to increase the CVP by 2 mm Hg and then determining whether it increased the CO [7]. In a study of 83 ICU patients, Magder and Bafaqeeh showed that patients who were able to increase the CVP by 2 mm Hg after receiving a bolus of approximately 500 mL of isotonic crystalloids over 10–30 minutes had a cardiac index increase of 300 mL/min/m2 . The two important findings of the study were that only 4.5% of the patients with a CVP more than 10 mm Hg responded to a fluid challenge, also of the patients who had increase in CO, 42% had a simultaneous increase in blood pressure. Therefore, the two conclusions from the study were that patients with a CVP of more than 10 mm Hg were poor responders to volume infusion and that 10 mm Hg may represent euvolemia in most individuals. Second, blood pressure increase was not a good indicator of cardiac response to a fluid challenge [8].

The Society of Critical Care Medicine’s Surviving Sepsis protocol recommends to fluid resuscitate a septic patient to a target CVP goal of 8–12 mm Hg and 12–15 mm Hg in a mechanically ventilated patient [9]. Though these recommendations are part of a sepsis care bundle that has been shown to improve survival, the CVP should not be interpreted in isolation, but rather in conjunction with the cardiac output. Depending on where the patient is on the Frank Starling curve, some patients may be adequately resuscitated at a CVP of 6-7 mm Hg, while others may still be hypovolemic at 10 mm Hg.

The discussion above applies to use of jugular and subclavian veins. If these sites are not available, femoral vein line pressures have been shown to be a good substitute for CVP. In various studies in spontaneously breathing and ventilated patients, inferior vena cava pressures have been shown to be around 0.5 mm Hg lower than the SVC pressures on average and rarely more than 3 mm Hg different [10]. Though the readings have been shown to be reliable in patients receiving mechanical ventilation and high positive end expiratory pressure, the same rule cannot be applied to patients with raised intra-abdominal pressures, as encountered during laparoscopy.

A healthy person can have a CVP less than zero in upright position and still have an adequate CO and be euvolemic. Conversely, CVP can be high in a patient with poor ventricular function and low cardiac output or with a good ventricular function and volume overload [6]. As illustrated by these common scenarios, values derived from pressure readings are most useful when used in conjunction with a dynamic clinical response such as blood pressure or urine output, or some measure of cardiac output. Echocardiography, analysis of venous oxyhemoglobin saturation, and immediate changes in blood pressure have all been used in this regard.

Another distinct advantage of central venous catheterization is the ability to obtain simultaneous measurements of pressure or pressure change and central venous saturations (ScvO2) with a single device. ScvO2 values of more than 70% are consistent with an adequate cardiac output and perfusion status, although this relationship is true only with adequate hemoglobin levels and stable oxygen requirements (VO2) of the body. The early goal-directed therapy for severe sepsis and septic shock has ScvO2 monitoring as an integral part of the resuscitation goals [11]. Intermittent blood sampling and catheters with special sensors are equally efficacious means of measuring ScvO2.

3.2. Pulmonary Artery Catheters (PACs) and Pulmonary Artery Occlusion (Wedge) Pressures

Aberrations in right heart compliance and pulmonary vascular resistance as may arise from cardiac and lung pathology can drastically alter the relationship between CVP and left heart preload. Pulmonary artery catheterization has therefore remained an attractive option to measure both right and left heart and pulmonary artery pressures in critically ill patients with such underlying pathology. The monitor is based on a balloon-tipped catheter that “floats” through the right atrium and ventricle and into the pulmonary artery as it is advanced. When “wedged” in one of the proximal branches of the pulmonary artery (PA), the catheter is in continuity with the left ventricle. With some minor gradients in between structures, as long as the catheter is the West Zone III of the lung and no mitral valve pathology is present, the following relationship exists.

3.3. Wedge Pressure Left Atrial Pressure Left Ventricular End-Diastolic Pressure (LVEDP)

When pulmonary vascular resistance is not elevated, the PA diastolic pressure approximates the pulmonary artery wedge pressure (PAWP) and provides a useful estimate of LVEDP without having to advance and wedge the catheter.

Pulmonary artery catheters can also be used to measure mixed venous saturations and CO via the Fick principle by analysis of blood aspirated from the catheter’s distal port. Typically, most CO measurements are made by electronic analysis of a thermal wash-out curve of fluid of a known temperature detected proximally at the point of injection and distally at a thermistor located 4 cm from the catheter tip. Continuous measurement of CO is accomplished with catheters containing an additional thermal filament and a computation module. The latter system requires occasional set-up time and calibration. Data from either of these catheters can be used to derive right ventricular ejection fraction (REF) and right ventricular end-diastolic volume (RVEDV). Thus, in addition to titration of fluids, the impact of inotropes, vasoconstrictors, and other interventions can be measured and followed. Low REF has been found to be helpful in identifying myocardial depression and as a marker of poor prognosis in septic patients [12]. In cardiac surgery patients, REF and RVEDV may be helpful in early detection of impaired right ventricular function secondary to right coronary artery stenosis [13].

As with the CVP, PAWP measurements are dependent on myocardial compliance. Multiple studies on ICU patients have shown that PAWP in acute illness correlates poorly or inconsistently with left ventricular end-diastolic volume (LVEDV) [1417]. Comparisons of PAWP and right RVEDV index found the latter to correlate more reliably with changes in cardiac index (CI) [18, 19]. Though RVEDV by PAC has been found to be consistent with different methods of measurement, such as ventriculography [20] and angiography [21, 22], other investigators have found the data less convincing [23, 24]. De Simone et al. compared 3D echocardiography data with PAC-derived RVEDV and found no significant correlation between the two [25].

For most of the past three decades, use of the PAC was based on the uncontested assumption that targeted improvement of physiologic parameters to supranormal endpoints is a desirable strategy. This approach began in the early 1970s when Shoemaker and colleagues described a pattern of higher ventricular performance, oxygen delivery, and oxygen consumption which predicted survival in trauma patients [26]. Subsequent studies in surgical patients seemed to confirm the survival benefit of using the PAC to facilitate increasings CO and oxygen delivery [27, 28]. In high-risk surgical patients, a significant reduction in mortality (4% versus 30%) followed early implementation of a protocol attempting to reach supranormal cardiac indices (CI ≥ 4.5 L/min/m2), [29]). A similar reduction in mortality (5.7% versus 22%) was seen in high-risk surgical patients randomized to a protocol where an oxygen delivery index of greater than 600 mL/min/m2 was achieved by dopexamine infusion [30], and a later study of trauma patients demonstrated a decrease in mortality from 37% to 18% with a similar protocol of PAC-facilitated endpoints [31].

Using the PAC to facilitate increased oxygen delivery in medical and surgical ICU patients was the next natural step. However, therapy designed to achieve such supranormal values had no effect in one study where therapy was targeted to achieve a supranormal CI and normal SvO2 [32], did not improve outcome in patients randomized to receive an O2 delivery index of ≥600 mL/min/m2 in another [33], did not improve outcomes in septic patients where supranormal indices of DO2 and VO2 were attempted [34], and, finally, decreased survival in a study using the PAC to guide supranormal hemodynamics [35]. In the studies above, use of the PAC was intertwined with a specific protocol for achieving key therapeutic endpoints, thus seeming to negate the value of the technology, when the flaw may have been the endpoint the technology was used to achieve. A more recent generation of controlled studies comparing pulmonary artery and CVP catheters with less stringent, and in some cases clinician-generated endpoints, have again shown no benefit of the catheter per se in high-risk surgical patients [36], patients with shock and sepsis [37], and adult patients with the acute respiratory distress syndrome (ARDS) [38]. In the Canadian study of surgical patients, the PAC was associated with a higher incidence of nonfatal pulmonary embolism [36].

Use of PACs has fallen over the last ten years due to the factors cited above, as well as due to higher complication rates [39], frequent misinterpretation of PAC data [40], and relative success with CVP-based methods for resuscitation in septic shock [11]. The American Society of Anesthesiologist Practice Guidelines recommend PAC for high-risk surgical patients only [41].

At present no study has demonstrated a positive association between PAC use for fluid management and survival [42]. Many of the randomized studies allowed physician exclusion of patients thought to benefit from the monitoring system, such as patients with heart failure and pulmonary hypertension where fluid management is only one of a handful of variables under active scrutiny and manipulation. Such use is likely to be highly individualized and unlikely to conform to a set protocol or study design that will allow a definitive statement of benefit. It is therefore prudent to be aware of the capabilities of the PAC and maintain an open mind about its potential value when a fuller physiologic picture may be necessary to make decisions regarding fluid responsiveness and volume status.

4. Cardiorespiratory Interactions and Dynamic Analysis of Fluid Status

Cardiac output and blood pressure interact with the respiratory system in a predictable manner according to the relationships indicated in Table 1. With positive pressure ventilation, venous return to the left ventricle is augmented, causing a rise in cardiac output and blood pressure during early inspiration. Later, the decrease in RV preload caused by positive intrathoracic pressure causes a drop in LV preload and systemic blood pressure [42]. Compared to the static hemodynamic parameters (CVP, PAWP), which are measured at one point in the cardiac cycle, dynamic parameters have been shown to be more accurate in assessment of intravascular volume and fluid responsiveness.

tab1
Table 1: Cardiorespiratory interactions used to predict volume responsiveness are listed. RV: right ventricle; LV: left ventricle.

It is well recognized that pulsus paradoxus, or an inspiratory fall in systolic blood pressure by more than 10 mm Hg, is seen in critically ill patients with hypovolemia. With hypovolemia, the myocardium is at the steep portion of the Frank-Starling curve, so any minor variation in the preload with inspiration or expiration can cause appreciable changes in CO and blood pressure. Some rather direct and accurate inferences regarding intravascular volume can be made from analysis of arterial pressure waveforms and Doppler analysis of aortic blood flow. Ironically, while positive pressure ventilation is used, the accuracy of pressure-based measures of fluid responsiveness such as CVP is highly debated, while flow-based measurements achieve their highest accuracy. Indices of intravascular fluid and preload assessment derived from positive pressure ventilator-induced arterial blood pressure changes include systolic pressure variability, the respiratory systolic variation test, stroke volume variability, and respiratory changes in arterial pulse pressure.

4.1. Systolic Pressure Variability (SPV)

The range of blood pressure (difference between maximal and minimal systolic BP) during a single positive pressure breath is defined as SPV. The baseline systolic pressure is taken using a short apneic period. SPV has 2 components, delta up and delta down, corresponding to the difference between the baseline and the peak amplitude during early expiration and at end inspiration, respectively. SPV refers to the sum of delta up and delta down or the total amplitude variation (Figure 2).

231493.fig.002
Figure 2: Arterial line tracings showing systolic pressure variation during the respiratory cycle. The pressure at end expiration defines baseline systolic pressure. SPV has two components, delta up and delta down, corresponding to systolic pressure waves reading at peak amplitude of early inspiration (the upward component or delta up) and at end inspiration (the downward component or delta down). The total amplitude variation or the sum of delta up and delta down is thus the SPV, shown with the arrow on the right of the diagram.

Beside intravascular volume status, SPV and its two components can be affected by a multitude of factors which include arrhythmias, chest wall and lung compliance, abdominal pressure, method of ventilation (spontaneous or mechanical), and myocardial function [4346]. All these factors remaining stable, variations in SPV will reflect changes in intravascular volume. With a normal myocardium functioning at the steep portion of the Frank Starling Curve, SPV is mainly due to delta down component, with a decrease in ventricular filling with positive pressure ventilation [44]. On the other hand, SPV is comprised of a delta up component in a patient with a failing myocardium, which is preload insensitive but responsive to afterload reduction with positive pressure ventilation [47]. The increase in pressure surrounding the heart decreases transmural pressure during systole (and hence afterload), thereby improving the ejection fraction. No exact values of SPV for determination of hypovolemia have been established. In a study by Rooke, SPV less than 5 mm Hg and delta down component less than 2 mm Hg showed absence of hypovolemia [48].

Other methods of analyzing fluid responsiveness from respiration-induced changes in arterial pulse waves have been described and validated. The respiratory systolic variation test uses the slope obtained by the minimal systolic BP during a ventilator maneuver involving four successive incremental positive pressure breaths [49]. Greater negative slope was found to correlate with increasing amounts of hemorrhage [50]. Respiratory change in arterial pulse pressure (DPp) is defined as the difference in pulse pressure of the highest (Ppmax) and lowest magnitude (Ppmin) over several respiratory cycles, divided by the mean of the 2 values, expressed as a percentage. In one study, a DPp of 13% differentiated fluid responders from nonresponders [51]. Stroke volume variability is an index that is based on beat-to-beat variability in the arterial pulse during the respiratory cycle [52].

While all of these parameters can be measured manually via printouts from raw arterial pressure waves, a number of commercial systems discussed below provide automated analysis and trending along with means of measuring cardiac output. The PiCCO system (PULSION Medical Systems AG, Munich, Germany) derives volumetric parameters from arterial waveforms via a set of algorithms termed pulse contour analysis. The latter method, first described by Wesseling et al., computes the end-diastolic to end-systolic change in pressure over systemic vascular resistance (SVR) [53]. The beat-to-beat change in the shape of the arterial pressure waveform reflects the changes in the impedance of the aorta. Absolute values of aortic impedance and SVR obtained by thermodilution (using the same monitor) are necessary for calibration of the monitor; from this, values of stroke volume, cardiac output, and vascular resistance are derived. Thus, pulse contour is able to convert pressure-based signals such as pulse pressure and pulse pressure variability into analogous volume-based signals. Validation studies with the PAC showed that pulse contour analysis was able to track increases in stroke volume accurately in volume responsive postcardiac surgery patients [54, 55]. Likewise, through use of SVV as a gauge of volume responsiveness, investigators were able to predict intraoperative fluid requirements for obese patients undergoing bariatric surgery [56].

In principle, the monitor could indicate fluid responsiveness via pulse variability with mechanical ventilation as a stand-alone monitor, but, with calibration, an additional wealth of information is available. Unlike the PAC, the injectate signal is collected at one of the large arteries outside of the thorax (femoral, axillary, or brachial) and thus measures transpulmonary blood flow. Mathematical analysis of the transpulmonary dilution curve can be carried out further to derive actual fluid compartments within the thorax. Two of the derived parameters, extravascular lung water (EVLW) and global end-diastolic volume (GEDV), are of particular interest. Both GEDV and the related intrathoracic blood volume (ITBV) are volumetric preload parameters in contrast to the more commonly used pressure-derived parameters from CVP and PACs. Changes in GEDV have correlated better with changes in stroke volume than with changes in CVP [57]. The ability to measure extravascular lung water carries the hope of being able to differentiate and follow problems such as pulmonary edema and ARDS [58], the latter with a positive survival associated with which excess fluid use could be avoided [38, 59]. An observational study of septic patients by Martin demonstrated a better survival in those with lower EVLW [60]. Further, fluid therapy algorithms guided by GEDV and EVLW rather than clinical assessment can lead to faster resolution of pulmonary edema, shortened requirement for vasopressors, mechanical ventilation, and thereby ICU stays [61, 62].

An investigation of postoperative cardiac surgery patients showed that CO derived by PAC and transpulmonary method is clinically comparable [63]. The advantage of transpulmonary thermodilution over the PAC is that it does not require insertion into a pulmonary artery (reduce risk of arrhythmias and pulmonary artery rupture), though it does require a CVC and an arterial catheter and their attendant complications. Artifactual increases in CO introduced by tricuspid regurgitation in PAC use would not be present with use of transpulmonary blood flow measurement. Inability to measure mixed venous oxygen saturations can be partially compensated by trending central venous saturations via a CVC.

Similar measurements using transpulmonary lithium dilution and detection by an ion selective electrode are the basis of another commercial system providing continuous CO and pulse wave analysis, the PulseCO (LiDCO Ltd, London, UK) [64]. The concentration-time curve for lithium ion calibration is read out on an attachment appended to the tubing on an existing artery after injecting from any intravenous line. The analytic algorithm is different from pulse contour in a few fundamental ways. The resulting “pulse power” computation is not sensitive to changes in pulse pressure amplitude found in distal arteries and is not affected by under- or over-dampening of arterial pressure transducers [65]. With these corrections built into the computational scheme, the LiDCO does not require a central arterial line. The less invasive nature of this system may also present advantages in certain patient populations such as those with limited central access or severe arterial disease. Despite these advantages, the LIDCO system does not give data such as GEDV and ITBV, and calibration measurements are limited by a desire to minimize lithium exposures (the number is, however, not defined by the manufacturer). The accuracy of the system is also low on patients on therapeutic lithium [65]. A more recent iteration of the LiDCO, the “LiDCO rapid,” uses a software-based nomogram to estimate stroke volume without the need for lithium calibration. Accuracy of the latter has been shown to differ from thermodilution-based cardiac output measurements in cardiac surgery patient; the thought being that the recalibration may be required in such dynamic situations where the arterial compliance fluctuates widely through the course of a case (i.e., pre- versus post-cardiopulmonary bypass) [66]. Pulse wave analysis systems including both LiDCO and PiCCO have been used in combination with passive leg raising to accurately predict fluid responsiveness [67].

Another recently introduced system called FloTrac/Vigileo (Edward Lifesciences, Irvine, Calif) has also been designed to function without external calibration. Instead individual demographic data (height, weight, age, and sex) are used with arterial waveform analysis to calculate CO, the principle being that SV is proportional to pulse pressure (PP). In addition, the FloTrac can also calculate SVV, and if a central venous catheter is present, SVR and central venous saturation can be obtained. Like the LiDCO rapid, the FloTrac features ease of use and requires only an arterial catheter [68]. A study comparing PiCCO with FloTrac found FloTrac/Vigileo (2nd generation software) to be as accurate as the PiCCO in predicting fluid responsiveness using SVV, although the threshold value of SVV for the FloTrac/Vigileo was lower (9.6%) than for the PiCCO (12.1%) [69]. A meta-analysis found FloTrac in acceptable agreement with CO derived with thermodilution techniques. However, the authors concluded that, in patients with rapid hemodynamic changes, hyperdynamic circulation, aortic regurgitation and intraaortic balloon pump counterpulsation, its use was questionable [68].

Pressure Recording Analytical Method (PRAM) (MostCare device, Vytech, Padu, Italy) is another device that estimates CO from the area under the arterial pressure wave. No external calibration is required as with FloTrac. A study in hemodynamically stable children comparing Doppler echocardiography found PRAM reliable, though more validation studies are required [70].

5. Echocardiography

From its initial use in outpatient settings in 1970s, echocardiography has found its place in all inpatient settings, ranging from the operating rooms to the intensive care unit (ICU). Different modalities of ultrasound include transthoracic echo being most noninvasive and portable, while the more invasive nature of esophageal Doppler and transesophageal echo is well tolerated in anesthetized patients. Nevertheless, all echographic techniques spare the patient the complications related to vascular access and indwelling devices. Also, in contrast to the PAC which is dependent on pressure measurements to make volume determinations, echocardiography relies on direct visualization of the cardiac anatomy and flow dynamics. Moreover, in patients where causes of circulatory failure overlap, echocardiography provides the ability to evaluate structural abnormalities (pericardial tamponade), contractility (ejection fraction), besides assessment of intravascular volume.

5.1. Transthoracic Echo (TTE)

Improved image quality with portable study echo machines in the last decade has made TTE a popular tool for intravascular fluid assessment in the ICU. More recently, its use has been advocated in the perioperative settings due to its ability to provide quick, noninvasive functional and fluid assessment. Right heart preload can be reliably obtained by direct measurement of the inferior vena cava diameter (IVC) variations with respiration (Figure 3) and also by right and left ventricular end-diastolic volumes. One study showed that a 50% decrease in IVC diameter (caval index), seen by subcostal views with spontaneous breathing, correlated with an RA pressure of less than 10 mm Hg (mean SD 6 ± 5), as measured by CVP measurements [71]. Recent study in emergency department settings found caval index measurement a useful noninvasive tool for initial determination of CVP [72].

231493.fig.003
Figure 3: A subcostal view of the cavoatrial junction and adjacent structures. RA: right atrium; IVC: inferior vena cava; LIV: liver.

In mechanically ventilated patients, IVC variations with respiration (Delta DIVC) of 12% differentiated patients who responded with increased CO to a fluid bolus from nonresponders [73]. In another study with mechanically ventilated septic patients, the CVP and the IVC diameter increase on inspiration (distensibility index (dIVC)) was measured before and after a gelatin fluid challenge of 7 mL/kg. Response was measured as an increase in CI of 15% or more. dIVC greater than 18% predicted fluid responsiveness with a sensitivity and specificity of Ninety percent. CVP, on the other hand, correlated poorly with CI ( , ) and dIVC [74]. In spite of difficulty in visualizing IVC in postabdominal surgeries and obese patients, TTE provides a quick, noninvasive and reliable method of volume status of the patient.

5.2. Transesophageal Echocardiography (TEE)

Besides cardiac anesthesia, TEE is now used routinely in other complex and long surgeries (e.g., liver transplant), especially among patients with known cardiac pathology. In our operating rooms, the quick availability of TEE machines makes this a first-line diagnostic modality for experienced users when unexpected hemodynamic instability is encountered.

For general use, the main focus is on intravascular volume status assessment and overall cardiac function. In the ICU, TEE is recommended when diagnostic information cannot be adequately obtained by TTE [75]. Limitations of TEE include the lack of availability of trained personal and equipment, the requirement of adequate sedation or intubation, and the impracticality for continuous postoperative use. Also, TEE is an invasive procedure and contraindicated in patients with esophageal strictures and malformation.

As with TTE for IVC measurements to determine preload, TEE can be used to measure SVC collapsibility. Volume expansion was found to decrease SVC collapse, decrease RV stroke volume variation, and increase CI in a study in 22 mechanically ventilated ARDS patients with septic shock [76]. A collapsibility index of 36% was found to predict an 11% increase in CI following a fluid bolus in a separate study [77].

Both TTE and TEE allow visual estimation of ventricular volume. Ninety percent of the stroke volume is obtained by ventricular shortening in the short axis [78]. Therefore, measurement of the LV end-diastolic area using mid-transgastric short axis view by TEE can give a reliable estimate of LV end diastolic volume [79]. In practice, a qualitative assessment of LVEDA provides a quicker and quite reliable assessment, as Leung and Levine report that systolic cavity obliteration is 100% sensitive in detecting hypovolemia [80]. Fluid challenge in these cases can be followed by serial echo views of ventricular filling and correlation with the CVP. On the other hand, distended left ventricle with decreased EF would indicate minimal benefit from a fluid challenge. Additionally, a distended right ventricle and empty left ventricle may indicate right ventricular failure possibly from acute pulmonary embolism or pulmonary vascular disease.

5.3. Esophageal Doppler Monitoring (EDM)

Using the Doppler principle to calculate blood flow velocity in the aorta, EDM can be used to provide minimally invasive continuous CO monitoring. EDM comprises of a small, thin, flexible probe with a Doppler transducer at the tip, inserted 35 cm into the esophagus, and has a tolerability similar to a nasogastric tube [81]. A continuous Doppler beam is used to generate flow velocity profile by analyzing reflected ultrasound waves from the moving red blood cells. Figure 4 shows a flow-time tracing generated by EDM. Flow time is defined as the time needed by the left ventricle to eject the SV and is represented on the waveform from the start of the upstroke to its return to baseline. As flow time is also dependant on the heart rate, “corrected flow time” or FTc analogous to QTc in the electrocardiogram (ECG) is used in flow calculations. Area under the flow time curve is called the stroke distance. Combining with cross-sectional area can give the stroke volume. The cross-sectional area of the descending aorta can be obtained using nomograms (based on patients’ height and weight) or measured directly with the M-mode by the EDM. Correction factor of 1.4 is used to adjust for blood lost to the coronaries and upper body circulation [81]. Additional analysis of flow velocity profile (peak velocity and upslope) can be used to evaluate left ventricular sensitivity to afterload and contractility.

231493.fig.004
Figure 4: Esophageal Doppler tracings from descending aorta blood flow. Flow time and peak velocity are indicated by dashed and vertical arrows, respectively. Parameters derived from the latter measurements include corrected flow time (FTc), acceleration, and stroke distance. The area under the curve (AUC) is equivalent to stroke distance, the length traveled by an erythrocyte during a single cardiac cycle. Assuming that the flow is via a cylindrical path, the stroke volume (SV) is the product of aortic cross-sectional area and the stroke distance. Aortic blood flow is calculated from the product of heart rate and calculated SV and is not exactly equal to true cardiac output, as approximately 10% of total cardiac output is diverted to subclavian and cerebral arteries upstream to where flow is measured.

While nearly emerging devices for fluid management have sought to first establish themselves as similar to the “gold standard” PAC, the initial studies of EDM are relatively unique in their attempt to associate a positive clinical outcome with device use. In postsurgical patients, hypovolemia is a notorious cause of splanchnic hypoperfusion, prolonged hospital stays, and even increased mortality in moderate- to high-risk surgical patients [8284]. With the use of a fluid management protocol built around using the continuous CO and FTc parameters of EDM monitoring, investigators were able to reduce the time to bowel motility, decrease the time to solid food intake, and shorten hospital stay [82, 85].

An analogous method for evaluation of cardiac output is based on estimation of blood flow via analysis of electrical bioimpedance. With electrical velocimetry (EV), all measurements are made via four ECG electrodes applied along one side of the patient’s body. A small amplitude alternating current is applied to one pair of electrodes, while the other detects the change in electrical signal through the cardiac cycle. When blood moves forward through the ascending aorta during ventricular ejection, red cells tend to align more along the body’s long axis, and create less impedance of electrical signals traveling along the same vector. Thus, from analysis of the electrical signal during the cardiac cycle, velocity and flow time parameters are generated and used to calculate stroke volume and cardiac output.

Comparisons between EV and thermodilution cardiac output measurements show a good level of agreement over a wide range of outputs in adult cardiac surgery patients and critically ill adults [86, 87]. An excellent correlation ( ) was found between EV and cardiac output measurements made by the Fick method (using direct measurements of VO2) in children with congenital heart disease, and EV measurements correlated well with Doppler measurements made via transesophageal echo in cardiac surgery patients [88, 89].

Limitations to EDM use are similar to TEE. In addition, severe aortic regurgitation, severe lung disease, altered thoracic blood flow, and presence of intra-ortic balloon pump can alter accuracy of EDM [81].

6. Conclusion

Resuscitation of the hypotensive patient in the perioperative setting can be challenging. Accurate and timely assessment of fluid status and fluid responsiveness are cornerstones in the management of the hypotensive patient. Both hypovolemia and hypervolemia can be harmful. Goal-directed fluid management can result in the appropriate use of fluids, vasopressors, and inotropes, resulting in improved patient outcomes. Invasive, noninvasive, static, and dynamic monitoring methods have strengths and limitations. Continual monitoring of unstable patients through the resuscitation period utilizing a combination of techniques can improve the utility of hemodynamic information, improve the accuracy of assessments of volume status, and improve patient outcomes.

References

  1. J. Jabot, J. L. Teboul, C. Richard, and X. Monnet, “Passive leg raising for predicting fluid responsiveness: importance of the postural change,” Intensive Care Medicine, vol. 35, no. 1, pp. 85–90, 2009. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  2. S. Préau, F. Saulnier, F. Dewavrin, A. Durocher, and J. L. Chagnon, “Passive leg raising is predictive of fluid responsiveness in spontaneously breathing patients with severe sepsis or acute pancreatitis,” Critical Care Medicine, vol. 38, no. 3, pp. 819–825, 2010. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  3. B. Lamia, A. Ochagavia, X. Monnet, D. Chemla, C. Richard, and J. L. Teboul, “Echocardiographic prediction of volume responsiveness in critically ill patients with spontaneously breathing activity,” Intensive Care Medicine, vol. 33, no. 7, pp. 1125–1132, 2007. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  4. X. Monnet, M. Rienzo, D. Osman et al., “Passive leg raising predicts fluid responsiveness in the critically ill,” Critical Care Medicine, vol. 34, no. 5, pp. 1402–1407, 2006. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  5. J. Boldt, M. Lenz, B. Kumle, and M. Papsdorf, “Volume replacement strategies on intensive care units: results from a postal survey,” Intensive Care Medicine, vol. 24, no. 2, pp. 147–151, 1998. View at Publisher · View at Google Scholar · View at Scopus
  6. S. Magder, “How to use central venous pressure measurements,” Current Opinion in Critical Care, vol. 11, no. 3, pp. 264–270, 2005. View at Publisher · View at Google Scholar · View at Scopus
  7. S. Magder, G. Georgiadis, and T. Cheong, “Respiratory variations in right atrial pressure predict the response to fluid challenge,” Journal of Critical Care, vol. 7, no. 2, pp. 76–85, 1992. View at Publisher · View at Google Scholar · View at Scopus
  8. S. Magder and F. Bafaqeeh, “The clinical role of central venous pressure measurements,” Journal of Intensive Care Medicine, vol. 22, no. 1, pp. 44–51, 2007. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  9. R. P. Dellinger, J. M. Carlet, H. Masur et al., “Surviving Sepsis Campaign guidelines for management of severe sepsis and septic shock,” Critical Care Medicine, vol. 32, no. 3, pp. 858–873, 2004. View at Publisher · View at Google Scholar
  10. J. Desmond and M. Megahed, “Is the central venous pressure reading equally reliable if the central line is inserted via the femoral vein,” Emergency Medicine Journal, vol. 20, no. 5, pp. 467–469, 2003. View at Scopus
  11. E. Rivers, B. Nguyen, S. Havstad et al., “Early goal-directed therapy in the treatment of severe sepsis and septic shock,” New England Journal of Medicine, vol. 345, no. 19, pp. 1368–1377, 2001. View at Publisher · View at Google Scholar · View at Scopus
  12. J. L. Vincent, “The measurement of right ventricular ejection fraction,” Intensive Care World, vol. 7, no. 3, pp. 333–336, 1990. View at Scopus
  13. R. Hines and P. G. Barash, “Intraoperative right ventricular dysfunction detected with a right ventricular ejection fraction catheter,” Journal of Clinical Monitoring, vol. 2, no. 3, pp. 206–208, 1986. View at Scopus
  14. L. Spinelli, M. Petretta, M. L. Vicario et al., “Losartan treatment and left ventricular filling during volume loading in patients with dilated cardiomyopathy,” American Heart Journal, vol. 143, no. 3, pp. 433–440, 2002. View at Publisher · View at Google Scholar · View at Scopus
  15. M. L. Fontes, W. Bellows, L. Ngo, and D. T. Mangano, “Assessment of ventricular function in critically ill patients: limitations of pulmonary artery catheterization. Institutions of the McSPI Research Group,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 13, no. 5, pp. 521–527, 1999. View at Publisher · View at Google Scholar
  16. C. P. Tousignant, F. Walsh, and C. D. Mazer, “The use of transesophageal echocardiography for preload assessment in critically iII patients,” Anesthesia and Analgesia, vol. 90, no. 2, pp. 351–355, 2000. View at Scopus
  17. R. M. Hansen, C. E. Viquerat, and M. A. Matthay, “Poor correlation between pulmonary arterial wedge pressure and left ventricular end-diastolic volume after coronary artery bypass graft surgery,” Anesthesiology, vol. 64, no. 6, pp. 764–770, 1986.
  18. M. L. Cheatham, L. D. Nelson, M. C. Chang, and K. Safcsak, “Right ventricular end-diastolic volume index as a predictor of preload status in patients on positive end-expiratory pressure,” Critical Care Medicine, vol. 26, no. 11, pp. 1801–1806, 1998. View at Scopus
  19. L. Diebel, R. F. Wilson, J. Heins, H. Larky, K. Warsow, and S. Wilson, “End-diastolic volume versus pulmonary artery wedge pressure in evaluating cardiac preload in trauma patients,” Journal of Trauma, vol. 37, no. 6, pp. 950–955, 1994. View at Scopus
  20. F. G. Spinale, A. C. Smith, B. A. Carabello, and F. A. Crawford, “Right ventricular function computed by thermodilution and ventriculography. A comparison of methods,” Journal of Thoracic and Cardiovascular Surgery, vol. 99, no. 1, pp. 141–152, 1990. View at Scopus
  21. P. Urban, D. Scheidegger, J. Gabathuler, and W. Rutishauser, “Thermodilution determination of right ventricular volume and ejection fraction: a comparison with biplane angiography,” Critical Care Medicine, vol. 15, no. 7, pp. 652–655, 1987. View at Scopus
  22. W. Voelker, H. P. Gruber, O. Ickrath, R. Unterberg, and K. R. Karsch, “Determination of right ventricular ejection fraction by thermodilution technique—a comparison to biplane cineventriculography,” Intensive Care Medicine, vol. 14, 2, pp. 461–466, 1988. View at Scopus
  23. M. M. Hoeper, J. Tongers, A. Leppert, S. Baus, R. Maier, and J. Lotz, “Evaluation of right ventricular performance with a right ventricular ejection fraction thermodilution catheter and MRI in patients with pulmonary hypertension,” Chest, vol. 120, no. 2, pp. 502–507, 2001. View at Publisher · View at Google Scholar · View at Scopus
  24. M. Hein, A. B. Roehl, J. H. Baumert, R. Rossaint, and P. Steendijk, “Continuous right ventricular volumetry by fast-response thermodilution during right ventricular ischemia: head-to-head comparison with conductance catheter measurements,” Critical Care Medicine, vol. 37, no. 11, pp. 2962–2967, 2009. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  25. R. De Simone, I. Wolf, S. Mottl-Link et al., “Intraoperative assessment of right ventricular volume and function,” European Journal of Cardio-Thoracic Surgery, vol. 27, no. 6, pp. 988–993, 2005. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  26. W. C. Shoemaker, E. S. Montgomery, E. Kaplan, and D. H. Elwyn, “Physiologic patterns in surviving and nonsurviving shock patients. Use of sequential cardiorespiratory variables in defining criteria for therapeutic goals and early warning of death,” Archives of Surgery, vol. 106, no. 5, pp. 630–636, 1973. View at Scopus
  27. M. Y. Rady, J. D. Edwards, and P. Nightingale, “Early cardiorespiratory findings after severe blunt thoracic trauma and their relation to outcome,” British Journal of Surgery, vol. 79, no. 1, pp. 65–68, 1992. View at Scopus
  28. J. Tuchschmidt, J. Fried, R. Swinney, and O. P. Sharma, “Early hemodynamic correlates of survival in patients with septic shock,” Critical Care Medicine, vol. 17, no. 8, pp. 719–723, 1989. View at Scopus
  29. W. C. Shoemaker, P. L. Appel, H. B. Kram, K. Waxman, and T. S. Lee, “Prospective trial of supranormal values of survivors as therapeutic goals in high-risk surgical patients,” Chest, vol. 94, no. 6, pp. 1176–1186, 1988. View at Scopus
  30. O. Boyd, R. M. Grounds, and E. D. Bennett, “A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients,” Journal of the American Medical Association, vol. 270, no. 22, pp. 2699–2707, 1993. View at Publisher · View at Google Scholar · View at Scopus
  31. M. H. Bishop, W. C. Shoemaker, P. L. Appel et al., “Prospective, randomized trial of survivor values of cardiac index, oxygen delivery, and oxygen consumption as resuscitation endpoints in severe trauma,” Journal of Trauma, vol. 38, no. 5, pp. 780–787, 1995. View at Publisher · View at Google Scholar · View at Scopus
  32. L. Gattinoni, L. Brazzi, P. Pelosi et al., “A trial of goal-oriented hemodynamic therapy in critically ill patients,” New England Journal of Medicine, vol. 333, no. 16, pp. 1025–1032, 1995. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  33. M. Yu, M. M. Levy, P. Smith, S. A. Takiguchi, A. Miyasaki, and S. A. Myers, “Effect of maximizing oxygen delivery on morbidity and mortality rates in critically ill patients: a prospective randomized, controlled study,” Critical Care Medicine, vol. 21, no. 6, pp. 830–838, 1993. View at Scopus
  34. J. Tuchschmidt, J. Fried, M. Astiz, and E. Rackow, “Elevation of cardiac output and oxygen delivery improves outcome in septic shock,” Chest, vol. 102, no. 1, pp. 216–220, 1992. View at Scopus
  35. M. A. Hayes, A. C. Timmins, E. H. Yau, M. Palazzo, C. J. Hinds, and D. Watson, “Elevation of systemic oxygen delivery in the treatment of critically ill patients,” New England Journal of Medicine, vol. 330, no. 24, pp. 1717–1722, 1994. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  36. J. D. Sandham, R. D. Hull, R. Frederick Brant et al., “A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients,” New England Journal of Medicine, vol. 348, no. 1, pp. 5–14, 2003. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  37. C. Richard, J. Warszawski, N. Anguel et al., “Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial,” Journal of the American Medical Association, vol. 290, no. 20, pp. 2713–2720, 2003. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  38. A. P. Wheeler, G. R. Bernard, B. T. Thompson et al., “Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury,” New England Journal of Medicine, vol. 354, no. 21, pp. 2213–2224, 2006. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  39. T. D. Coulter and H. P. Wiedemann, “Complications of hemodynamic monitoring,” Clinics in Chest Medicine, vol. 20, no. 2, pp. 249–267, 1999. View at Publisher · View at Google Scholar · View at Scopus
  40. M. J. Jacka, M. M. Cohen, T. To, J. H. Devitt, and R. Byrick, “Pulmonary artery occlusion pressure estimation: how confident are anesthesiologists?” Critical Care Medicine, vol. 30, no. 6, pp. 1197–1203, 2002. View at Scopus
  41. M. F. Roizen, D. L. Berger, R. A. Gabel et al., “Practice guidelines for pulmonary artery catheterization: an updated report by the american society of anesthesiologists task force on pulmonary artery catheterization,” Anesthesiology, vol. 99, no. 4, pp. 988–1014, 2003. View at Publisher · View at Google Scholar · View at Scopus
  42. M. R. Shah, V. Hasselblad, L. W. Stevenson et al., “Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials,” Journal of the American Medical Association, vol. 294, no. 13, pp. 1664–1670, 2005. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  43. R. Pizov, Y. Ya'ari, and A. Perel, “Systolic pressure variation is greater during hemorrhage than during sodium nitroprusside-induced hypotension in ventilated dogs,” Anesthesia and Analgesia, vol. 67, no. 2, pp. 170–174, 1988. View at Scopus
  44. R. Pizov, Y. Ya'ari, and A. Perel, “The arterial pressure waveform during acute ventricular failure and synchronized external chest compression,” Anesthesia and Analgesia, vol. 68, no. 2, pp. 150–156, 1989. View at Scopus
  45. A. Szold, R. Pizov, E. Segal, and A. Perel, “The effect of tidal volume and intravascular volume state on systolic pressure variation in ventilated dogs,” Intensive Care Medicine, vol. 15, no. 6, pp. 368–371, 1989. View at Scopus
  46. M. Takata, R. A. Wise, and J. L. Robotham, “Effects of abdominal pressure on venous return: abdominal vascular zone conditions,” Journal of Applied Physiology, vol. 69, no. 6, pp. 1961–1972, 1990. View at Scopus
  47. M. R. Pinsky, G. M. Matuschak, and J. M. Itzkoff, “Respiratory augmentation of left ventricular function during spontaneous ventilation in severe left ventricular failure by grunting. An auto-EPAP effect,” Chest, vol. 86, no. 2, pp. 267–269, 1984. View at Scopus
  48. G. A. Rooke, H. A. Schwid, and Y. Shapira, “The effect of graded hemorrhage and intravascular volume replacement on systolic pressure variation in humans during mechanical and spontaneous ventilation,” Anesthesia and Analgesia, vol. 80, no. 5, pp. 925–932, 1995. View at Scopus
  49. A. Perel, L. Minkovich, S. Preisman, M. Abiad, E. Segal, and P. Coriat, “Assessing fluid-responsiveness by a standardized ventilatory maneuver: the respiratory systolic variation test,” Anesthesia and Analgesia, vol. 100, no. 4, pp. 942–945, 2005. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  50. A. M. L. Perel, M. Abiad, P. Coriat, and P. Viars, “Respiratory Systolic Variation Test-a new method for assessing preload,” British Journal of Anaesthesia, vol. 74, 1995.
  51. F. Michard, S. Boussat, D. Chemla et al., “Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure,” American Journal of Respiratory and Critical Care Medicine, vol. 162, no. 1, pp. 134–138, 2000. View at Scopus
  52. D. A. Reuter, T. W. Felbinger, E. Kilger, C. Schmidt, P. Lamm, and A. E. Goetz, “Optimizing fluid therapy in mechanically ventilated patients after cardiac surgery by on-line monitoring of left ventricular stroke volume variations. Comparison with aortic systolic pressure variations,” British Journal of Anaesthesia, vol. 88, no. 1, pp. 124–126, 2002. View at Publisher · View at Google Scholar · View at Scopus
  53. K. H. Wesseling, J. A. P. Weber, and N. T. Smith, “A simple device for the continuous measurement of cardiac output. Its model basis and experimental verification,” Advanced Cardiovascular Physiology, vol. 5, pp. 16–52, 1983.
  54. T. W. Felbinger, D. A. Reuter, H. K. Eltzschig, J. Bayerlein, and A. E. Goetz, “Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with arterial pulse contour analysis,” Journal of Clinical Anesthesia, vol. 17, no. 4, pp. 241–248, 2005. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  55. D. A. Reuter, T. W. Felbinger, C. Schmidt et al., “Stroke volume variations for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery,” Intensive Care Medicine, vol. 28, no. 4, pp. 392–398, 2002. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  56. A. K. Jain and A. Dutta, “Stroke volume variation as a guide to fluid administration in morbidly obese patients undergoing laparoscopic bariatric surgery,” Obesity Surgery, vol. 20, no. 6, pp. 709–715, 2010. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  57. F. Michard, S. Alaya, V. Zarka, M. Bahloul, C. Richard, and J. L. Teboul, “Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock,” Chest, vol. 124, no. 5, pp. 1900–1908, 2003. View at Publisher · View at Google Scholar · View at Scopus
  58. D. M. Berkowitz, P. A. Danai, S. Eaton, M. Moss, and G. S. Martin, “Accurate characterization of extravascular lung water in acute respiratory distress syndrome,” Critical Care Medicine, vol. 36, no. 6, pp. 1803–1809, 2008. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  59. W. Isakow and D. P. Schuster, “Extravascular lung water measurements and hemodynamic monitoring in the critically ill: bedside alternatives to the pulmonary artery catheter,” American Journal of Physiology—Lung Cellular and Molecular Physiology, vol. 291, no. 6, pp. L1118–L1131, 2006. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  60. G. S. Martin, S. Eaton, M. Mealer, and M. Moss, “Extravascular lung water in patients with severe sepsis: a prospective cohort study,” Critical Care, vol. 9, no. 2, pp. R74–R82, 2005. View at Scopus
  61. M. S. Goepfert, D. A. Reuter, D. Akyol, P. Lamm, E. Kilger, and A. E. Goetz, “Goal-directed fluid management reduces vasopressor and catecholamine use in cardiac surgery patients,” Intensive Care Medicine, vol. 33, no. 1, pp. 96–103, 2007. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  62. P. R. Eisenberg, J. R. Hansbrough, D. Anderson, and D. P. Schuster, “A prospective study of lung water measurements during patient management in an intensive care unit,” American Review of Respiratory Disease, vol. 136, no. 3, pp. 662–668, 1987. View at Scopus
  63. F. Mielck, W. Buhre, G. Hanekop, T. Tirilomis, R. Hilgers, and H. Sonntag, “Comparison of continuous cardiac output measurements in patients after cardiac surgery,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 17, no. 2, pp. 211–216, 2003. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  64. D. A. Reuter, C. Huang, T. Edrich, S. K. Shernan, and H. K. Eltzschig, “Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives,” Anesthesia and Analgesia, vol. 110, no. 3, pp. 799–811, 2010. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  65. A. Rhodes and R. Sunderland, “Arterial pulse power analysis: the LiDCO plus system,” in Functional Hemodynamic Monitoring, M. Pinsky and D. Payen, Eds., Springer, Heidelberg, Germany, 2005.
  66. O. Broch, J. Renner, J. Höcker et al., “Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery,” Critical Care, vol. 15, p. R76, 2011. View at Publisher · View at Google Scholar · View at PubMed
  67. B. Geerts, R. De Wilde, L. Aarts, and J. Jansen, “Pulse contour analysis to assess hemodynamic response to passive leg raising,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 25, no. 1, pp. 48–52, 2011. View at Publisher · View at Google Scholar · View at PubMed
  68. J. Mayer, J. Boldt, R. Poland, A. Peterson, and G. R. Manecke, “Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 23, no. 3, pp. 401–406, 2009. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  69. C. K. Hofer, A. Senn, L. Weibel, and A. Zollinger, “Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac and PiCCOplus system,” Critical Care, vol. 12, no. 3, p. R82, 2008. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  70. M. Calamandrei, L. Mirabile, S. Muschetta, G. F. Gensini, L. De Simone, and S. M. Romano, “Assessment of cardiac output in children: a comparison between the pressure recording analytical method and doppler echocardiography,” Pediatric Critical Care Medicine, vol. 9, no. 3, pp. 310–312, 2008. View at Publisher · View at Google Scholar · View at PubMed
  71. B. J. Kircher, R. B. Himelman, and N. B. Schiller, “Noninvasive estimation of right atrial pressure from the inspiratory collapse of the inferior vena cava,” American Journal of Cardiology, vol. 66, no. 4, pp. 493–496, 1990. View at Publisher · View at Google Scholar
  72. A. D. Nagdev, R. C. Merchant, A. Tirado-Gonzalez, C. A. Sisson, and M. C. Murphy, “Emergency department bedside ultrasonographic measurement of the caval index for noninvasive determination of low central venous pressure,” Annals of Emergency Medicine, vol. 55, no. 3, pp. 290–295, 2010. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  73. M. Feissel, F. Michard, J. P. Faller, and J. L. Teboul, “The respiratory variation in inferior vena cava diameter as a guide to fluid therapy,” Intensive Care Medicine, vol. 30, no. 9, pp. 1834–1837, 2004. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  74. C. Barbier, Y. Loubières, C. Schmit et al., “Respiratory changes in inferior vena cava diameter are helpful in predicting fluid responsiveness in ventilated septic patients,” Intensive Care Medicine, vol. 30, no. 9, pp. 1740–1746, 2004. View at Publisher · View at Google Scholar · View at PubMed
  75. D. M. Thys, Z. Hillel, M. E. Goldman, B. P. Mindich, and J. A. Kaplan, “A comparison of hemodynamic indices derived by invasive monitoring and two-dimensional echocardiography,” Anesthesiology, vol. 67, no. 5, pp. 630–634, 1987.
  76. A. Vieillard-Baron, R. Augarde, S. Prin, B. Page, A. Beauchet, and F. Jardin, “Influence of superior vena caval zone condition on cyclic changes in right ventricular outflow during respiratory support,” Anesthesiology, vol. 95, no. 5, pp. 1083–1088, 2001. View at Scopus
  77. A. Vieillard-Baron, K. Chergui, A. Rabiller et al., “Superior vena caval collapsibility as a gauge of volume status in ventilated septic patients,” Intensive Care Medicine, vol. 30, no. 9, pp. 1734–1739, 2004. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  78. J. S. Rankin, P. A. McHale, C. E. Arentzen, et al., “The three dimensional dynamic geometry of the left ventricle in the conscious dog,” Circulation Research, vol. 39, no. 3, pp. 304–313, 1976. View at Scopus
  79. F. M. Clements, D. H. Harpole, T. Quill, R. H. Jones, and R. L. McCann, “Estimation of left ventricular volume and ejection fraction by two-dimensional transoesophageal echocardiography: comparison of short axis imaging and simultaneous radionuclide angiography,” British Journal of Anaesthesia, vol. 64, no. 3, pp. 331–336, 1990.
  80. J. M. Leung and E. H. Levine, “Left ventricular end-systolic cavity obliteration as an estimate of intraoperative hypovolemia,” Anesthesiology, vol. 81, no. 5, pp. 1102–1109, 1994. View at Scopus
  81. M. Singer, “Esophageal Doppler monitoring of aortic blood flow: beat-by-beat cardiac output monitoring,” International Anesthesiology Clinics, vol. 31, no. 3, pp. 99–125, 1993. View at Scopus
  82. D. H. Conway, R. Mayall, M. S. Abdul-Latif, S. Gilligan, and C. Tackaberry, “Randomised controlled trial investigating the influence of intravenous fluid titration using oesophageal Doppler monitoring during bowel surgery,” Anaesthesia, vol. 57, no. 9, pp. 845–849, 2002. View at Publisher · View at Google Scholar · View at Scopus
  83. M. G. Mythen and A. R. Webb, “Perioperative plasma volume expansion reduces the incidence of gut mucosal hypoperfusion during cardiac surgery,” Archives of Surgery, vol. 130, no. 4, pp. 423–429, 1995. View at Scopus
  84. S. Sinclair, S. James, and M. Singer, “Intraoperative intravascular volume optimisation and length of hospital stay after repair of proximal femoral fracture: randomised controlled trial,” British Medical Journal, vol. 315, no. 7113, pp. 909–912, 1997.
  85. T. J. Gan, A. Soppitt, M. Maroof et al., “Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery,” Anesthesiology, vol. 97, no. 4, pp. 820–826, 2002.
  86. S. Suttner, T. Schöllhorn, J. Boldt et al., “Noninvasive assessment of cardiac output using thoracic electrical bioimpedance in hemodynamically stable and unstable patients after cardiac surgery: a comparison with pulmonary artery thermodilution,” Intensive Care Medicine, vol. 32, no. 12, pp. 2053–2058, 2006. View at Publisher · View at Google Scholar · View at PubMed
  87. N. Zoremba, J. Bickenbach, B. Krauss, R. Rossaint, R. Kuhlen, and G. Schälte, “Comparison of electrical velocimetry and thermodilution techniques for the measurement of cardiac output,” Acta Anaesthesiologica Scandinavica, vol. 51, no. 10, pp. 1314–1319, 2007. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  88. K. Norozi, C. Beck, W. A. Osthaus, I. Wille, A. Wessel, and H. Bertram, “Electrical velocimetry for measuring cardiac output in children with congenital heart disease,” British Journal of Anaesthesia, vol. 100, no. 1, pp. 88–94, 2008. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus
  89. C. Schmidt, G. Theilmeier, H. Van Aken et al., “Comparison of electrical velocimetry and transoesophageal Doppler echocardiography for measuring stroke volume and cardiac output,” British Journal of Anaesthesia, vol. 95, no. 5, pp. 603–610, 2005. View at Publisher · View at Google Scholar · View at PubMed · View at Scopus