BioMed Research International
Volume 2013 (2013), Article ID 937936, 9 pages
http://dx.doi.org/10.1155/2013/937936
Research Article

Anti-Inflammatory Effects of Arsenic Trioxide Eluting Stents in a Porcine Coronary Model

1Shanghai Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai 200032, China
2Key Laboratory for Ultrafine Materials of Ministry of Education, School of Materials Science and Engineering, East China University of Science and Technology, Shanghai 200237, China
3Department of Cardiology, Sichuan Provincial People’s Hospital, Chengdu 610072, China
4Department of Cardiology, The First Affiliated Hospital of Harbin Medical University, Harbin, Heilongjiang 150001, China

Received 8 November 2012; Accepted 26 December 2012

Academic Editor: George Perry

Copyright © 2013 Li Shen et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Previous research from our group has demonstrated arsenic trioxide eluting stents significantly reduced neointimal area and thickness compared with bare metal stents. In the present study, the anti-inflammatory effects of arsenic trioxide in vitro and arsenic trioxide eluting stents in a porcine coronary model have been explored. Sixty-five pigs underwent placement of 139 oversized stents in the coronary arteries with histologic analysis, endothelial function analysis, and immunohistochemical and western blot analyses. Arsenic trioxide eluting stents effectively inhibited local inflammatory reactions, while no significant difference in endothelialization and endothelial function between arsenic trioxide eluting stents and bare metal stents was observed. Arsenic trioxide eluting stents favorably modulate neointimal formation due to less augmentation of early inflammatory reactions, and quick endothelialization of the stent surface, which might contribute to long-term safety and efficacy of drug eluting stents.

1. Introduction

Polymer-based sirolimus and paclitaxel drug eluting stents (DESs) have revolutionized the treatment of coronary artery diseases through remarkable reduction of angiographic target lesion revascularization [1, 2]. In real world clinical settings, however, concerns about long-term safety and efficacy have emerged [36]. The first-generation sirolimus and paclitaxel eluting stents are both associated with incomplete neointimal coverage [7, 8], impaired endothelial cell function [9], thrombosis [10], hypersensitivity reactions [11], and incomplete stent apposition [12]. Currently used drugs (sirolimus and paclitaxel) and nondegradable polymer coatings are mostly considered to be responsible for late adverse events and pathologic reactions, including the late stent thrombosis, inflammation, and hypersensitivity [13, 14].

DESs are designed to reduce in-stent neointimal growth through the elution of agents that arrest the cycle of cell proliferation. An ideal DES drug should suppress excessive neointimal growth while maintaining the proliferation of endothelial cells and functioning endothelium [15]. Arsenic trioxide (As2O3) has been used in the treatment of patients with acute promyelocytic leukemia (APL) [16, 17] and other types of malignant cancers [1822]. As2O3 acts on cells through a variety of mechanisms, influencing numerous signal transduction pathways and resulting in a vast range of cellular effects that include apoptosis induction, growth inhibition, promotion or inhibition of differentiation, and angiogenesis inhibition [23, 24]. In a recent report [25], a heparin-immobilized copolymer of L-lactide (LA) and 5-methyl-5-benzyloxycarbonate-1,3-dioxan-2-one (MBC) has been synthesized as a biodegradable coating material for local As2O3 delivery. This copolymer was found to reduce thrombosis, localized hypersensitivity, and inflammation, which belong to late-stage adverse events.

The pathological process of in-stent restenosis (ISR) is characterized by an inflammatory healing response after stretch and damage of the vessel wall [2634]. Stent restenotic reduction of DESs can be mainly ascribed to its antiproliferative activity, but anti-inflammation may also be an important explanation for controlling inflammation-triggered proliferative process [35, 36]. As2O3 eluting stents (AESs) effectively reduce neointimal thickening in rabbit and porcine models through inhibiting the cell cycle and inducing vascular smooth muscle cell (VSMC) apoptosis [25, 37]. The present study was performed to evaluate the behavior of inflammatory cells to As2O3 and effect of AESs in a porcine coronary model, thus reveal the mechanism of AESs in reducing ISR.

2. Materials and Methods

2.1. Animals

Pigs (~20 kg weight) were obtained from the Shanghai Animal Administration Center and received daily oral antiplatelet medication until termination. All animal experiments were approved by the Animal Care and Use Committee of Fudan University and were in compliance with the “Guide for the Care and Use of Laboratory Animals” published by the National Academy Press (NIH Publication No. 85-23, revised in 1996).

2.2. Preparation and Surface Morphology of AESs

Heparinized polymer solution [37] (0.1 wt.% in dichloromethane), As2O3 aqueous solution (1 wt.%), and heparinized polymer solution (0.1 wt.% in dichloromethane) were sequentially sprayed onto the surface of BMSs (  mm, Beijing Amsinomed Medical Company, China) to form the basecoat ( μg), drug layer ( μg), and topcoat ( μg), respectively. Polymer-coated stents (PCSs) were prepared by directly spraying the polymer solution onto the stent surface and the quantity of polymer was μg per stent.

2.3. Stent Implantation

On the procedure day, thirty male pigs were anesthetized with ketamine (20 mg/kg intramuscularly) and xylazine (2 mg/kg intramuscularly). BMSs ( , 7 for one week, 5 for two weeks, and 9 for four weeks), PCSs ( , 5 for one week, 5 for two weeks, and 6 for four weeks), and AESs ( , 9 for one week, 11 for two weeks, and 12 for four weeks) were implanted in 2-3 coronary arteries per pig by random assignment to anatomic location. The resulting stent-to-artery ratio was about 1.2-1.3 : 1 by quantitative coronary angiography analysis. The animals were anesthetized with ketamine (20 mg/kg) and xylazine (2 mg/kg) for follow-up angiography in the same orthogonal views before death with 20 mL of potassium chloride intracoronary injection. The stented arteries were carefully dissected from the myocardium and cut into two pieces, each about 9 mm long for cross sections preparation and SEM imaging.

2.4. Measurements of As2O3 Levels in Tissues and Serum

In additional animals at 0, 4, 24, 72 h, and 7 days after deployment of As2O3 eluting stents ( , 6 stents per time point) in the coronary arteries, the delivery of As2O3 from stents to tissues and serum was evaluated using hydride generation reaction interfaced with atomic fluorescence spectrometry assay and HPLC, respectively [37]. In this experiment, fifteen pigs were used and each pig received two As2O3 eluting stents.

2.5. Endothelialization of Stented Arteries and Evaluation of Endothelial Function

The endothelialization of stented arteries was examined using SEM 1, 2, and 4 weeks after stent implantation. Endothelial function after stent implantation was estimated by measuring the coronary vasomotor reactivity in response to acetylcholine (Ach, 60 mg, performed at an infusion rate of 1 mL/min) infusion within 6-month followup. Ten pigs receiving two stents each and 20 stents (6 BMSs, 7 PCSs, and 7 AESs) were used in the experiment. End-diastolic images for each segment were chosen and analyzed with the automated edge detection program (FD-10, Philips, Best, The Netherland). Two orthogonal views with less foreshortening or without overlapping of side branches were selected and averaged for biplane assessment by two experts blinded to stent type. About 5 mm distal to the site of stenting was chosen for analysis. Changes in coronary diameter in response to Ach coronary infusion were expressed as percent changes versus baseline angiograms.

2.6. Immunohistochemical and Western Blot Analyses

For immunohistochemical and western blot analyses, ten pigs were used and each pig randomized received two stents (7 BMSs, 6 PCSs, and 7 AESs). After 4 weeks, the stented arteries were harvested and cut into two pieces. The stent struts were carefully removed from the coronary arteries. Serial sections from the stented arteries were fixed in phosphate buffered saline (PBS) containing 10% formalin, dehydrated by treatment with solutions of increasing alcohol content (70%, 85%, 95%, and 100%), followed by xylene and then embedded in paraffin. Immunohistochemical studies were performed using primary antibodies against interleukin-6 (IL-6), monocyte chemoattractant protein-1 (MCP-1), CD3, and S100. Briefly, the sections were deparaffinized, heated at 95°C for 30 min in antigen retrieval solution, fixed in 0.3% hydrogen peroxide for 30 min, and then air dried. Nonspecific antibody binding sites were blocked by a 30 min incubation in normal horse serum (10%, 0.1% Triton X-100, 0.1 M PBS, pH 7.4). The slide was then incubated in the primary antibodies, respectively. (1 : 100, 0.1% Triton X-100, 0.1 M PBS) (R&D systems, Minneapolis, MN, USA) overnight at 4°C. After three washes in PBS, the slides were again blocked in normal horse serum as described above and exposed to the horseradish peroxidase-conjugated secondary antibody (1 : 400, 0.1% Triton X-100, 0.1 M PBS) (Beyotime, China), followed by the addition of 3,30-diaminobenzidine tetrahydrochloride (DAB) solution. Nuclei were also counterstained with hematoxylin. The total tissue area of the immunostained sections of each specimen was outlined manually and measured in mm2. The tissue areas of the immunostained sections occupied by positive cells were measured automatically using greyscale detection with a fixed threshold and the ratios of immunopositive areas were calculated as percentages of the total tissue area.

Vessel wall expression of inflammatory-associated proteins (IL-6 and MCP-1) was determined by western blot analysis. Briefly, protein extracts (50 μg) were size fractionated on SDS-polyacrylamide gels, transferred to nitrocellulose membrane. Positive control for each target was run on the same gel. Membranes were incubated with an affinity purified polyclonal antibody to IL-6 and MCP-1, respectively, washed, and incubated with secondary antibody. Signals were detected by the ECL chemiluminescence detection system. Glyceraldehyde-3-phosphate dehydrogenase (GAPDH) was used as an internal control to ensure equal amount of protein extract in each sample.

2.7. Statistical Analysis

Numerical data are presented as mean ± standard error of the mean. Continuous variables were compared by ANOVA ( -test with Bonferroni correction), and categorical variables were compared by χ2 test. A value of ≤ 0.05 was considered as a significant difference.

3. Results

3.1. Surface Morphology Examination of AESs

Figure 1 shows the stereoscopic and SEM images of an As2O3 eluting stent before and after expansion. It can be clearly observed that As2O3 crystals were dispersed between the basecoat and topcoat, while the coatings were still uniform. No delamination or destruction of the coatings on the stent can be observed, indicating the drug does not impair integrity of the coatings.

937936.fig.001
Figure 1: Digital photographs (A) and SEM images (B) of As2O3 eluting stent before (1) and after (2) expansion.
3.2. In Vivo As2O3 Levels after Stent Implantation

The levels of As2O3 measured in arterial tissue, serum, and main organs after AESs implantation are presented in Table 1. At 4 h after stent deployment, the mean concentration of As2O3 within the adjacent arterial tissue was μg/g, whereas the value in serum was μg/g, nearly one hundred times higher than that in the serum. The concentration of As2O3 in stented artery tissue reached a maximum level at 24 h, and decreased thereafter, whereas the values in serum, heart, liver, and kidney progressively declined after 4 h. Even at day 7 after stent implantation, the As2O3 concentration in the arterial wall was μg/g, while no As2O3 can be detected in the serum, heart, liver, and kidney.

tab1
Table 1: As2O3 levels in serum and tissues.
3.3. Morphological Evaluation of the Stented Arteries

The typical morphologies of the stented arteries 1, 2, and 4 weeks after stent implantation are shown in Figure 2. In the first two weeks, BMSs and AESs had different appearance with fibrin-platelet deposition and acute inflammatory cells. At one week, a relatively thick layer of matrix composed of many inflammatory cells and fibrin-associated foreign body reactions was present over the stent struts for BMSs, while a much thinner layer of such matrix was observed in the AES group. The percent area stenosis for BMSs and AESs at this time point was % and %, respectively. After two weeks, neointima consisting of amorphous material and inflammatory cells formed. The amorphous material adjacent to the stent strut was also less for AESs as compared with the controlled BMSs. After 4 weeks, neointima consists of SMCs and matrix proteoglycans formed. The density of SMCs in neointima was much lower in the AES group than in the BMS group. There was a trend that the percent area stenosis at four weeks was lower in the AES group compared to BMS, although the difference was not significant ( ).

937936.fig.002
Figure 2: Typical high-power micrographs 1 week, 2 and 4 weeks after the placement of BMSs and AESs in normal porcine coronary arteries. Insert: Digital camera photographs of the inner wall of the stented arteries. At the first week, a thick layer of matrix consists of a lot of inflammatory cells and fibrin associated with foreign body reactions is present over the stent struts for BMSs. A much thinner layer of such matrix can be observed for AESs. At 2 weeks, the neointima consisting of amorphous material and inflammatory cells formed. This amorphous material adjacent to the stent strut was never observed 2 weeks after AESs implantation. At 4 weeks, the neointima consists of smooth muscle cells and matrix proteoglycans. The density of smooth muscle cells in neointima is much lower for AESs than BMSs.
3.4. Endothelialization of the Stented Arteries and Evaluation of Endothelial Function

The progress of endothelialization of stented arteries was examined using SEM at 1, 2, and 4 weeks after stent implantation as shown in Figure 3(a). At the first week, both AES and BMS had quite low endothelial cells attachment. AES showed much less fibrin-platelet deposition than BMS, which was consistent with the morphological results in Figure 2. For the second week of development, endothelial cell attachment could be found to occur on the BMS surface with a relative low density. Although AESs were reported to mildly delay endothelialization in a rabbit iliac artery injury model [25], no difference was observed in a porcine coronary model. After 4 weeks, there was no difference in endothelialization for AES and BMS. The arteries treated with BMSs or AESs were fully endothelialized; the lumen surface of the vessel wall and the stent struts were covered by confluent endothelial cells.

fig3
Figure 3: Typical SEM images (magnification: ×500, ×100 for insets) of the inner wall of stented arteries 1, 2, and 4 weeks after implantation (a) and relaxation responses of coronary segments distal to stents within 6-month followup (b). At the first week, both AESs and BMSs showed quite few endothelial cells attachment. For the second week of development, the endothelialization of AESs was slightly higher than BMSs ( ). After 4 weeks, the arteries treated with BMS or AESs were fully endothelized. The diameter changes between the BMSs, PCSs, and AESs had no significant differences of vasoreactivity in response to the Ach infusion in sites distal to the stents within 6-month followup.

In this study, the diameter changes in response to the Ach infusion in sites distal to the stents between BMSs, PCSs, and AESs had no significant differences within 6-month followup (Figure 3(b)).

3.5. Immunohistochemical and Western Blotting Analyses

Figure 4 shows the immunohistochemical staining and western blotting results 4 weeks after stent implantation. Stent-based As2O3 delivery significantly ( ) inhibited MCP-1, IL-6, and CD3 expression. The ratios of MCP-1, IL-6, and CD3 positive area for BMS, PCS, and AES were , , %; , , %; and , , %, respectively, in Figure 4(a). Immunohistochemical analysis failed to detect any difference in the expression of S100 between BMS, PCS, and AES 4 weeks after stent implantation. We further used antibodies for MCP-1 and IL-6 in western blotting analysis as demonstrated in Figure 4(b). Vascular segments treated with PCSs demonstrated a slight increase in MCP-1 and IL-6 compared with the BMSs control group ( ). The expression levels of MCP-1 and IL-6 in the arteries treated with AESs were significantly lower than those treated with BMS and PCS ( ).

fig4
Figure 4: Immunohistochemical staining (a) and western blotting analysis (b) of arteries treated with BMSs, PCSs, and AESs for 4 weeks after stent implantation. The stent-based As2O3 delivery significantly reduced the expression of proteins MCP-1, IL-6, and CD3 ( ). In western blotting analysis, the expression of MCP-1 and IL-6 in the arteries treated with AESs was significantly lower than BMSs and PCSs ( ).

4. Discussion

It is well known that ISR is a consequence of inflammation, SMC proliferation, and migration [38]. The effects of As2O3 as an antiproliferative agent for drug eluting stents at clinically achievable concentrations were investigated without severe side effects. In the studies done in rabbit iliac artery model and porcine coronary model, AESs effectively reduce neointimal thickening through inhibiting the cell cycle and inducing VSMC apoptosis. No in-stent thrombosis can be observed within six-month followup in porcine coronary arteries [25, 37], suggesting that AESs are a feasible, safe, and efficient DES. In the present study, anti-inflammatory effect of As2O3 in vitro and effect of AESs in a porcine coronary model have been evaluated.

AESs were also found to significantly decrease the protein expression of MCP-1, IL-6, and CD3 compared with BMSs and PCSs. But unfortunately, we failed to detect any decrease of S100 expression, a marker of DCs 4 weeks after AESs implantation. Indeed, the expression of S100 in BMS, PCS, and AES treated arteries was low and the difference was not significant at this time point. MCP-1 is the prototype of the C-C chemokine-β subfamily and exhibits its most potent chemotactic activity toward monocytes and T lymphocytes. In addition to promoting the transmigration of circulating monocytes into tissue, MCP-1 exerts various other effects on monocytes, including superoxide anion induction, cytokines production, and adhesion molecule expression [39]. IL-6 is a cytokine that has been proven to be crucially involved in ischemic heart disease. This substance possesses pleiotropic functions that are mostly in favor of inflammation. Patients with a higher release of IL-6 in the coronary sinus after stent deployment have a higher risk of restenosis. Suzuki and coworkers [40] showed in an experimental model that sirolimus- and dexamethasone-covered stents induce a reduction in release of MCP-1 and IL-6 compared with BMSs, thus decreasing local inflammation.

Endothelium plays an integral role in maintaining vascular homeostasis. This role is not limited to the modulation of vascular tone but also for regulation of inflammation, platelet activation, and thrombosis [41, 42]. Drug eluting stents substantially reduce in-stent restenosis in patients with coronary artery disease and also lead to disruption of the endothelial layer and leaves a thrombogenic metallic surface exposed to the blood stream. Early DES devices have been associated with late stent thrombosis (LST) in patients, a rare though life-threatening event that has become an increasingly controversial issue for interventional cardiology. Delayed or incomplete re-endothelialization is an important predictor of LST, which is increasingly apparent for DES in real-world applications. The early establishment of a functional endothelial layer after vascular injury has been shown to assist in prevention of neointimal proliferation and thrombus formation [43, 44]. In our previous report, stent thrombosis did not occur within 6-month followup for the AESs in a porcine coronary artery model [37]. Stent-based local delivery of sirolimus (Cypher) and paclitaxel (Taxus) profoundly inhibited neointima formation but caused vasomotor dysfunction in distal conduit vessel segments [44]. As revealed in previous reports, the segments proximal and distal to the stents were both more strongly constricted to Ach infusion in the SES group compared with the BMS group [4547]. There was no significant difference in extensive endothelialization and endothelial function between AESs and BMSs after 4 weeks, suggesting improved safety of AES than SES and PES. The early restoration of endothelial function after AES implantation might be interpreted as the result of rapid elution of the loaded drug, which was completely released within 7 days, and thus minimized local toxicity [37]. In contrast, the loaded drug for sirolimus and paclitaxel eluting stents is released up to 60 days after stent implantation. Another reasonable explanation is related to the biodegradable coating material that can induce mild foreign body reactions and quick endothelialization [37].

5. Conclusions

In a porcine coronary artery model AES reduced the neointimal hyperplastic response to injury through inhibition of cell cycle and induction of apoptosis of VSMC [27, 48], less augmentation of early inflammatory reactions, and quick endothelialization of the stent surface. Although further studies are necessary to confirm the long-term efficacy and safety of this novel medical device, our data highlight the necessity to improve our understanding regarding the performance of AESs in reducing ISR.

Conflict of Interests

The authors declare no conflict of interests.

Authors’ Contribution

L. Shen, F. Gong, and W. Tian have contributed equally to this study.

Acknowledgments

The authors thank J. Jianguo and Y. Ruiming for their excellent technical assistance. This work is supported by National Basic Research Program of China (2011CB503905); the Science and Technology Commission of Shanghai, China (12DZ1940604) National Natural Science Foundation of China (30700318, 30870622, 81101133); and the Young Scientific “Phosphor” (Tracking)Project from the Shanghai Science and Technology Development (no. 11QH1400500).

References

  1. J. E. Sousa, M. A. Costa, A. G. M. R. Sousa et al., “Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries,” Circulation, vol. 107, no. 3, pp. 381–383, 2003. View at Publisher · View at Google Scholar · View at Scopus
  2. G. W. Stone, S. G. Ellis, D. A. Cox et al., “A Polymer-based, paclitaxel-eluting stent in patients with coronary artery disease,” The New England Journal of Medicine, vol. 350, no. 3, pp. 221–231, 2004. View at Publisher · View at Google Scholar · View at Scopus
  3. A. Kastrati, J. Mehilli, J. Pache et al., “Analysis of 14 trials comparing sirolimus-eluting stents with bare-metal stents,” The New England Journal of Medicine, vol. 356, no. 10, pp. 1030–1039, 2007. View at Publisher · View at Google Scholar · View at Scopus
  4. B. Lagerqvist, S. K. James, U. Stenestrand, J. Lindbäck, T. Nilsson, and L. Wallentin, “Long-term outcomes with drug-eluting stents versus bare-metal stents in Sweden,” The New England Journal of Medicine, vol. 356, no. 10, pp. 1009–1019, 2007. View at Publisher · View at Google Scholar · View at Scopus
  5. W. H. Maisel, “Unanswered questions: drug-eluting stents and the risk of late thrombosis,” The New England Journal of Medicine, vol. 356, no. 10, pp. 981–984, 2007. View at Publisher · View at Google Scholar · View at Scopus
  6. A. Farb, D. K. Weber, F. D. Kolodgie, A. P. Burke, and R. Virmani, “Morphological predictors of restenosis after coronary stenting in humans,” Circulation, vol. 105, no. 25, pp. 2974–2980, 2002. View at Publisher · View at Google Scholar · View at Scopus
  7. J. I. Kotani, M. Awata, S. Nanto et al., “Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings,” Journal of the American College of Cardiology, vol. 47, no. 10, pp. 2108–2111, 2006. View at Publisher · View at Google Scholar · View at Scopus
  8. G. J. Wilson, G. Nakazawa, R. S. Schwartz et al., “Comparison of inflammatory response after implantation of sirolimus- and paclitaxel-eluting stents in porcine coronary arteries,” Circulation, vol. 120, no. 2, pp. 141–149, 2009. View at Publisher · View at Google Scholar · View at Scopus
  9. S. H. Hofma, W. J. van der Giessen, B. M. van Dalen et al., “Indication of long-term endothelial dysfunction after sirolimus-eluting stent implantation,” European Heart Journal, vol. 27, no. 2, pp. 166–170, 2006. View at Publisher · View at Google Scholar · View at Scopus
  10. A. V. Finn, M. Joner, G. Nakazawa et al., “Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization,” Circulation, vol. 115, no. 18, pp. 2435–2441, 2007. View at Publisher · View at Google Scholar · View at Scopus
  11. A. V. Finn, G. Nakazawa, M. Joner et al., “Vascular responses to drug eluting stents: importance of delayed healing,” Arteriosclerosis, Thrombosis, and Vascular Biology, vol. 27, no. 7, pp. 1500–1510, 2007. View at Publisher · View at Google Scholar · View at Scopus
  12. J. Ako, Y. Morino, Y. Honda et al., “Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis,” Journal of the American College of Cardiology, vol. 46, no. 6, pp. 1002–1005, 2005. View at Publisher · View at Google Scholar · View at Scopus
  13. M. Joner, A. V. Finn, A. Farb et al., “Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk,” Journal of the American College of Cardiology, vol. 48, no. 1, pp. 193–202, 2006. View at Publisher · View at Google Scholar · View at Scopus
  14. T. F. Lüscher, J. Steffel, F. R. Eberli et al., “Drug-eluting stent and coronary thrombosis: biological mechanisms and clinical implications,” Circulation, vol. 115, no. 8, pp. 1051–1058, 2007. View at Publisher · View at Google Scholar · View at Scopus
  15. J. Ako, H. N. Bonneau, Y. Honda, and P. J. Fitzgerald, “Design criteria for the ideal drug-eluting stent,” The American Journal of Cardiology, vol. 100, no. 8, pp. S3–S9, 2007. View at Publisher · View at Google Scholar · View at Scopus
  16. P. Zhang, S. Y. Wang, and L. H. Hu, “Treatment of acute promyelocytic leukemia with intravenous arsenic trioxide,” Chinese Journal of Hematology, vol. 17, pp. 58–60, 1996.
  17. G. Q. Chen, J. Zhu, X. G. Shi et al., “In vitro studies on cellular and molecular mechanisms of arsenic trioxide (As2O3) in the treatment of acute promyelocytic leukemia: As2O3 induces NB4 cell apoptosis with downregulation of Bcl-2 expression and modulation of PML- RARα/PML proteins,” Blood, vol. 88, no. 3, pp. 1052–1061, 1996. View at Scopus
  18. H. Pelicano, L. Feng, Y. Zhou et al., “Inhibition of mitochondrial respiration: a novel strategy to enhance drug-induced apoptosis in human leukemia cells by a reactive oxygen species-mediated mechanism,” Journal of Biological Chemistry, vol. 278, no. 39, pp. 37832–37839, 2003. View at Publisher · View at Google Scholar · View at Scopus
  19. Y. H. Kang, M. J. Yi, M. J. Kim et al., “Caspase-independent cell death by arsenic trioxide in human cervical cancer cells: reactive oxygen species-mediated poly(ADP-ribose) polymerase-1 activation signals apoptosis-inducing factor release from mitochondria,” Cancer Research, vol. 64, no. 24, pp. 8960–8967, 2004. View at Publisher · View at Google Scholar · View at Scopus
  20. Z. Diaz, M. Colombo, K. K. Mann et al., “Trolox selectively enhances arsenic-mediated oxidative stress and apoptosis in APL and other malignant cell lines,” Blood, vol. 105, no. 3, pp. 1237–1245, 2005. View at Publisher · View at Google Scholar · View at Scopus
  21. M. Baumgartner, S. Sturlan, E. Roth, B. Wessner, and T. Bachleitner-Hofmann, “Enhancement of arsenic trioxide-mediated apoptosis using docosahexaenoic acid in arsenic trioxide-resistant solid tumor cells,” International Journal of Cancer, vol. 112, no. 4, pp. 707–712, 2004. View at Publisher · View at Google Scholar · View at Scopus
  22. X. H. Zhu, Y. L. Shen, Y. K. Jing et al., “Apoptosis and growth inhibition in malignant lymphocytes after treatment with arsenic trioxide at clinically achievable concentrations,” Journal of the National Cancer Institute, vol. 91, no. 9, pp. 772–778, 1999. View at Scopus
  23. W. H. Miller Jr., H. M. Schipper, J. S. Lee, J. Singer, and S. Waxman, “Mechanisms of action of arsenic trioxide,” Cancer Research, vol. 62, no. 14, pp. 3893–3903, 2002. View at Scopus
  24. G. J. Roboz, S. Dias, G. Lam et al., “Arsenic trioxide induces dose- and time-dependent apoptosis of endothelium and may exert an antileukemic effect via inhibition of angiogenesis,” Blood, vol. 96, no. 4, pp. 1525–1530, 2000. View at Scopus
  25. W. Yang, J. B. Ge, H. L. Liu et al., “Arsenic trioxide eluting stent reduces neointima formation in a rabbit iliac artery injury model,” Cardiovascular Research, vol. 72, no. 3, pp. 483–493, 2006. View at Publisher · View at Google Scholar · View at Scopus
  26. R. Kornowski, M. K. Hong, F. O. Tio, O. Bramwell, H. Wu, and M. B. Leon, “In-stent restenosis: contributions of inflammatory responses and arterial injury to neointimal hyperplasia,” Journal of the American College of Cardiology, vol. 31, no. 1, pp. 224–230, 1998. View at Publisher · View at Google Scholar · View at Scopus
  27. J. J. Li, J. Li, J. L. Nan et al., “Coronary restenotic reduction of drug-eluting stenting may be due to its anti-inflammatory effects,” Medical Hypotheses, vol. 69, no. 5, pp. 1004–1009, 2007. View at Publisher · View at Google Scholar · View at Scopus
  28. A. Gaspardone and F. Versaci, “Coronary stenting and inflammation,” The American journal of cardiology., vol. 96, no. 12 A, pp. 65L–70L, 2005. View at Scopus
  29. J. J. Li, S. P. Nie, C. Y. Zhang, Z. Gao, X. Zheng, and Y. L. Guo, “Is inflammation a contributor for coronary stent restenosis?” Medical Hypotheses, vol. 68, no. 5, pp. 945–951, 2007. View at Publisher · View at Google Scholar · View at Scopus
  30. A. Dibra, J. Mehilli, S. Braun et al., “Inflammatory response after intervention assessed by serial C-reactive protein measurements correlates with restenosis in patients treated with coronary stenting,” The American Heart Journal, vol. 150, no. 2, pp. 344–350, 2005. View at Publisher · View at Google Scholar · View at Scopus
  31. A. Farb, G. Sangiorgi, A. J. Carter et al., “Pathology of acute and chronic coronary stenting in humans,” Circulation, vol. 99, no. 1, pp. 44–52, 1999. View at Scopus
  32. P. H. Grewe, T. Deneke, A. Machraoui, J. Barmeyer, and K. M. Müller, “Acute and chronic tissue response to coronary stent implantation: pathologic findings in human specimen,” Journal of the American College of Cardiology, vol. 35, no. 1, pp. 157–163, 2000. View at Publisher · View at Google Scholar · View at Scopus
  33. C. Glover, X. Ma, Y. X. Chen et al., “Human in-stent restenosis tissue obtained by means of coronary atherectomy consists of an abundant proteoglycan matrix with a paucity of cell proliferation,” The American Heart Journal, vol. 144, no. 4, pp. 702–709, 2002. View at Publisher · View at Google Scholar · View at Scopus
  34. I. M. Chung, H. K. Gold, S. M. Schwartz, Y. Ikari, M. A. Reidy, and T. N. Wight, “Enhanced extracellular matrix accumulation in restenosis of coronary arteries after stent deployment,” Journal of the American College of Cardiology, vol. 40, no. 12, pp. 2072–2081, 2002. View at Publisher · View at Google Scholar · View at Scopus
  35. F. G. Welt and C. Rogers, “Inflammation and restenosis in the stent era,” Arteriosclerosis, Thrombosis, and Vascular Biology, vol. 22, no. 11, pp. 1769–1776, 2002. View at Publisher · View at Google Scholar · View at Scopus
  36. A. Colombo and G. Sangiorgi, “The Monocyte: the key in the lock to reduce stent hyperplasia?” Journal of the American College of Cardiology, vol. 43, no. 1, pp. 24–26, 2004. View at Publisher · View at Google Scholar · View at Scopus
  37. F. R. Gong, X. Y. Cheng, S. F. Wang, Y. C. Zhao, Y. Gao, and H. B. Cai, “Heparin-immobilized polymers as non-inflammatory and non-thrombogenic coating materials for arsenic trioxide eluting stents,” Acta Biomaterialia, vol. 6, no. 2, pp. 534–546, 2010. View at Publisher · View at Google Scholar · View at Scopus
  38. J. H. Bräsen, A. Kivelä, K. Röser et al., “Angiogenesis, vascular endothelial growth factor and platelet-derived growth factor-BB expression, iron deposition, and oxidation-specific epitopes in stented human coronary arteries,” Arteriosclerosis, Thrombosis, and Vascular Biology, vol. 21, no. 11, pp. 1720–1726, 2001. View at Scopus
  39. N. A. Scott, “Restenosis following implantation of bare metal coronary stents: pathophysiology and pathways involved in the vascular responses to injury,” Advanced Drug Delivery Reviews, vol. 58, no. 3, pp. 358–376, 2006. View at Publisher · View at Google Scholar · View at Scopus
  40. T. Suzuki, G. Kopia, S. I. Hayashi et al., “Stent-based delivery of sirolimus reduces neointimal formation in a porcine coronary model,” Circulation, vol. 104, no. 10, pp. 1188–1193, 2001. View at Scopus
  41. A. Blum, D. J. Schneider, B. E. Sobel, and H. L. Dauerman, “Endothelial dysfunction and inflammation after percutaneous coronary intervention,” The American Journal of Cardiology, vol. 94, no. 11, pp. 1420–1423, 2004. View at Publisher · View at Google Scholar · View at Scopus
  42. J. S. Li, R. Jabara, L. Pendyala et al., “Abnormal vasomotor function of porcine coronary arteries distal to sirolimus-eluting stents,” Journal of American College of Cardiology, vol. 1, no. 3, pp. 279–285, 2008. View at Publisher · View at Google Scholar · View at Scopus
  43. N. Werner, S. Junk, U. Laufs et al., “Intravenous transfusion of endothelial progenitor cells reduces neointima formation after vascular injury,” Circulation research, vol. 93, no. 2, pp. e17–e24, 2003. View at Scopus
  44. D. Kong, L. G. Melo, A. A. Mangi et al., “Enhanced inhibition of neointimal hyperplasia by genetically engineered endothelial progenitor cells,” Circulation, vol. 109, no. 14, pp. 1769–1775, 2004. View at Publisher · View at Google Scholar · View at Scopus
  45. J. W. Kim, S. Y. Suh, C. U. Choi et al., “Six-month comparison of coronary endothelial dysfunction associated with Sirolimus-eluting stent versus paclitaxel-eluting stent,” Journal of American College of Cardiology, vol. 1, no. 1, pp. 65–71, 2008. View at Publisher · View at Google Scholar · View at Scopus
  46. J. W. Kim, H. S. Seo, J. H. Park et al., “A Prospective, randomized, 6-month comparison of the coronary vasomotor response associated with a zotarolimus- versus a sirolimus-eluting stent,” Journal of the American College of Cardiology, vol. 53, no. 18, pp. 1653–1659, 2009. View at Publisher · View at Google Scholar · View at Scopus
  47. B. Dominik and G. Peter, “Endothelial function: from vascular biology to clinical applications,” The American Journal of Cardiology, vol. 90, no. 10, pp. 40L–48L, 2002. View at Scopus
  48. E. Grube, S. Sonoda, F. Ikeno et al., “Six- and twelve-month results from first human experience using everolimus-eluting stents with bioabsorbable polymer,” Circulation, vol. 109, no. 18, pp. 2168–2171, 2004. View at Publisher · View at Google Scholar · View at Scopus