Review Article

The Clinical Benefits of Adding a Third Dimension to Assess the Left Ventricle with Echocardiography

Department of Cardiac, Thoracic and Vascular Sciences, School of Medicine, University of Padua, Via Giustiniani 2, 35123 Padua, Italy

Received 6 November 2013; Accepted 23 January 2014; Published 15 May 2014

Academic Editors: L. Agati, A. V. Bruschke, and K. Egstrup

Copyright © 2014 Luigi P. Badano. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Three-dimensional echocardiography is a novel imaging technique based on acquisition and display of volumetric data sets in the beating heart. This permits a comprehensive evaluation of left ventricular (LV) anatomy and function from a single acquisition and expands the diagnostic possibilities of noninvasive cardiology. It provides the possibility of quantitating geometry and function of LV without preestablished assumptions regarding cardiac chamber shape and allows an echocardiographic assessment of the LV that is less operator-dependent and therefore more reproducible. Further developments and improvements for widespread routine applications include higher spatial and temporal resolution to improve image quality, faster acquisition, processing and reconstruction, and fully automated quantitative analysis. At present, three-dimensional echocardiography complements routine 2DE in clinical practice, overcoming some of its limitations and offering additional valuable information that has led to recommending its use for routine assessment of the LV of patients in whom information about LV size and function is critical for their clinical management.

1. Introduction

Quantitation of left ventricular (LV) size, geometry, and function represents the most frequent indication for an echocardiographic study and is pivotal for patient evaluation and management. Indication to cardiac surgery [1], to treatment initiation or suitability for device implantation in systolic heart failure [2], or to discontinuing cardiotoxic medications [3] is among the most critical decisions that rely on an accurate LV geometry and function assessment. Although conventional two-dimensional echocardiography (2DE) is by far the most used imaging modality to assess LV geometry and function in the clinical routine, its accuracy and reproducibility remain suboptimal, particularly when compared to other imaging modalities [4].

The advent of three-dimensional echocardiography (3DE) represents a real breakthrough in cardiovascular ultrasound. Advancements in computer and transducer technology permit the acquisition of 3D data sets with adequate spatial and temporal resolution for assessing most of cardiac pathologies. 3DE enables the visualization of cardiac structures from virtually any perspective, providing a more anatomically sound and intuitive display, as well as an accurate quantitative evaluation of anatomy and function of heart valves [510]. In addition, 3DE overcomes geometric assumptions and enables an accurate quantitative and reproducible evaluation of cardiac chambers [11, 12], thus offering solid elements for patient management [13, 14]. Furthermore, 3DE is the only imaging technique based on volumetric scanning able to show moving structures in the beating heart, in contrast to cardiac magnetic resonance (CMR) or cardiac computed tomography (CT), which are based on postacquisition 3D reconstruction from multiple tomographic images and displaying only 3D rendered snapshots.

Data regarding clinical applications of 3DE are burgeoning and gradually capturing an established place in the noninvasive clinical assessment of anatomy and function of cardiac structures. Recently, joint European Association of Echocardiography and American Society of Echocardiography recommendations have been published, aiming to provide clinicians with a systematic approach to 3D image acquisition and analysis [15]. Among the established indications for the clinical use of 3DE, the assessment of LV geometry and function is the one with more evidence.

This spotlight paper is aimed at summarizing the state-of-the-art 3DE applications to assess LV geometry and function, emphasizing the advantages of 3DE over conventional 2DE and its current limitations (Table 1).

tab1
Table 1: Left ventricular assessment by 3DE.

2. Instrumentation

The milestone in the history of 3DE has been the development of fully sampled matrix array transthoracic transducers based on advanced digital processing and improved image formation algorithms which allowed the operators to obtain on-cart transthoracic (TTE) real-time volumetric imaging with shorter acquisition time, higher spatial and temporal resolution (Figure 1) [16]. Further technological developments (i.e., advances in miniaturization of the electronics and in element interconnection technology) made it possible to insert a full matrix array into the tip of a transesophageal probe and provide transesophageal (TOE) real-time volumetric imaging [17].

897431.fig.001
Figure 1: Two-dimensional echocardiography is a tomographic technique which provides “flat” views of the heart and great vessels whose thickness is fixed and related to the piezoelectric element vertical dimension. Three-dimensional echocardiography is based on real-time volumetric imaging that allows acquisition of pyramidal data sets.
2.1. Comparison between 2DE and 3DE Ultrasound Imaging

A conventional 2D phased array transducer is composed of multiple piezoelectric elements electrically isolated from each other and arranged in a single row. Individual ultrasound wave fronts are generated by firing individual elements in a specific sequence with a delay in phase with respect to the transmit initiation time. Each element adds and subtracts pulses to generate a single ultrasound wave with a specific direction that constitutes a radially propagating scan line (Figure 2). The linear array can be steered in two dimensions—vertical (axial) and lateral (azimuthal)—while the resolution in the axis (elevation) is fixed by the thickness of the tomographic slice, which, in turn, is related to the vertical dimension of piezoelectric elements.

fig2
Figure 2: Schematic drawing of beamforming using a conventional 2D phased array transducer. During transmission (a), focused beams of ultrasound are produced by pulsing each piezoelectric element with precalculated time delays (i.e., phasing). During reception (b), focusing is achieved by applying selective delays at echo signals received by the different piezoelectric elements in order to create isophase signals that will be summed in a coherent way.

Currently, 3DE matrix array transducers are composed of about 3000 independent piezoelectric elements with operating frequencies ranging from 2 to 4 MHz and 5 to 7 MHz for TTE and TOE, respectively. These piezoelectric elements are arranged in a matrix configuration within the transducer (Figure 3(a)) in order to steer the ultrasound beam electronically. The electronically controlled phasic firing of the elements in that matrix generates a scan line that propagates radially and can be steered both laterally (azimuth) and in elevation in order to acquire a volumetric pyramid of data. The main technological breakthrough which allowed manufacturers to develop the matrix transducers has been the miniaturization of electronics that allowed the development of individual electrical interconnections for every piezoelectric require element which could be independently controlled, both in transmission and in reception. On the other hand, the microbeamforming allows the same size of the 2D cable to be used with 3D probes, despite the large number of digital channels required for these fully sampled elements to be connected.

fig3
Figure 3: (a) Schematic drawing of a full matrix array transducer where about 3000 acoustically independent piezoelectric elements are arranged in row and columns and used to steer the beam electronically. This matrix arrangement of piezoelectric elements allows their phasic firing to produce an ultrasound beam that can be steered in vertical (axial), lateral (azimuthal), and anteroposterior (elevation) directions in order to acquire a volumetric (pyramidal) data set. (b) Beamforming with 3D matrix array transducers. To save power and electronic circuitry needs (costs) and reduce the connection cable size the beamforming and steering processes have been split into two: the transducer and the ultrasound machine levels. The transducer contains the piezoelectric elements arranged in a matrix array, interconnection technology and integrated analog circuits (DELAY) to control transmit and receive signals using different subsections of the matrix (patches) to control analog prebeamforming and fine steering. Signals from each patch are summed in order to reduce the number of digital lines in the coaxial cable that connects the transducer to the ultrasound system from 3000 to the conventional size of 128–256 channels. At the ultrasound machine level, analog-to-digital (A/D) convertors amplify, filter, and digitize the elements signals. The resulting digital signals are focused (coarse steering) using digital delay (DELAY) circuitry and summed together ( ) to form the received signal from a desired object.

Beamforming is a technique used to process signals in order to produce directionally or spatially selected signal sent or received from arrays of sensors. In 2DE, all the electronic components for the beamforming (high-voltage transmitters, low-noise receivers, analog-to-digital converter, digital controllers, and digital delay lines) are in the system and consume a lot of power (around 100 W and 1,500 cm2 of personal computer electronics board area). If the same beamforming approach would have been used for matrix array transducers used in 3DE, it would require around 4 kW power consumption and a huge PC board area to accommodate all the needed electronics. To reduce both the power consumption and the size of the connecting cable, several miniaturized circuit boards are incorporated into the transducer, allowing partial beamforming to be performed in the probe (Figure 3). The 3000 channel circuit boards within the transducer control the fine steering by delaying and summing signals within subsections of the matrix, known as patches. This allows reducing the number of the digital channels to be put into the cable that connects the probe to the ultrasound system from 3000 to the conventional 125–256. Coarse steering is controlled by the ultrasound system where the analog-to-digital conversion occurs using digital delay lines (Figure 3).

Additionally, developments in transducer technology have resulted in a reduced transducer footprint, improved side-lobe suppression, increased sensitivity and penetration, and implementation of harmonic capabilities that can be used for both gray-scale and contrast imaging. The most recent generation of matrix transducers are significantly smaller than the previous ones and the quality of 2D and 3D imaging has improved significantly, allowing a single transducer to acquire both 2D and 3DE studies, as well as acquiring the whole left ventricular cavity in a single beat.

3. Image Acquisition and Display

Currently, 3D data set acquisition can be easily implemented into standard echocardiographic examination either by switching among 2D and 3D probes or, with newest all-in-one-probes, by switching between 2D and 3D modalities available in the same probe. The latter probes are also capable of providing single-beat full-volume acquisition, as well as real-time 3D color Doppler imaging.

At present two different methods for 3D data acquisition are available: “real-time” (or “live” 3D mode) and multibeat 3D mode (Figure 4) [18]. In the real-time mode, a thin sector of a pyramidal 3D volume data set is obtained and visualized live, beat after beat as during 2D scanning. Imaging is usually available in several fashions, as narrow volume, zoom, wide-angle (full-volume), and color-Doppler modalities. Heart dynamics is shown in a realistic way, with instantaneous online volume rendered reconstruction. It allows fast acquisition of dynamic pyramidal data structures from a single acoustic view that can encompass the entire heart without the need of reference system and electrocardiographic (ECG) and respiratory gating. Real-time imaging is time-saving for both data acquisition and analysis. Although this acquisition mode overcomes rhythm disturbances or respiratory motion limitations, it still suffers from relatively poor temporal and spatial resolutions.

fig4
Figure 4: Schematic representation of two-dimensional (i.e., tomographic (a)) and single-beat three-dimensional (i.e., volumetric (b)) of the left ventricular short axis at mitral valve level. Volumetric rendering displays many more details and allows better appreciation of spatial relationship among cardiac structures. (c) shows the multibeat acquisition modality: two to six consecutive single-beat subvolumes are stitched together to obtain a full-volume with higher spatial and temporal resolution.

Conversely, multibeat acquisition is realized through sequential acquisitions of narrow smaller volumes obtained from several ECG-gated consecutive heart cycles (from 2 to 6) that are subsequently stitched together to create a single volumetric data set. It provides large data sets with high temporal and spatial resolution, but more prone to artifacts due to patient or respiratory motion or irregular cardiac rhythms. The most appropriate acquisition mode for the specific clinical setting will be chosen in each individual case.

3D data sets can be sectioned in several planes and rotated in order to visualize the cardiac structure of interest from any desired perspective, irrespective of its orientation and position within the heart. This allows the operator to easily obtain unique visualizations that may be difficult or impossible to achieve using conventional 2DE (e.g., en face views of the tricuspid valve or cardiac defects). Two main actions are undertaken by the operator to obtain the desired view from a 3D volumetric data set: cropping and slicing. Similar to what the anatomists or the surgeons do to expose an anatomic structure within a 3DE data set, the operator should remove the surrounding chamber walls. This process of virtually removing the irrelevant neighbouring tissue is called cropping (Figure 5) and can be performed either during or after acquisition. In contrast with 2D images, displaying a cropped image requires also the definition of the viewing perspective (i.e., since the same 3D structure can be visualized en face from either above or below, as well as from any desired view angle) [1]. Slicing refers to a virtual “cutting” of the 3D data set into one or more (up to twelve) 2D (tomographic) grey-scale images (Figure 6).

897431.fig.005
Figure 5: Cropping allows obtaining stereoscopic images of the cardiac structures contained in a pyramidal 3D data set.
897431.fig.006
Figure 6: The 3D data set can be sliced in several 2D cut planes to obtain multiple views of that particular chamber (in the figure a 4–chamber view, upper panel, and an orthogonal view of the left ventricle and the left atrium, lower panel). Simultaneous visualization of orthogonal 2D slices enables the assessment and comparison of wall motion at every segment level of the ventricle, precious for assessing severity and extension of hypertrophy or regional wall motion evaluation both at rest and during stress.

Acquisition of volumetric images generates the technical problem of rendering the depth perception on a flat, 2D monitor. 3D images can be visualized using three display modalities: volume rendering, surface rendering, and tomographic slices (Figure 7). In volume rendering modality, various color maps are applied to convey the depth perception to the observer. Generally, lighter shades (e.g., bronze, Figure 8) are used for structures closer to the observer, while darker shades (e.g., blue, Figure 8) are used for deeper structures. Surface rendering modality displays the 3D surface of cardiac structures, identified either by manual tracing or by using automated border detection algorithms on multiple 2D cross-sectional images of the structure/cavity of interest (Figure 9). This stereoscopic approach is useful for the assessment of shape and for a better appreciation of geometry and dynamic function during the cardiac cycle. Finally, the pyramidal data set can be automatically sliced in several tomographic views simultaneously displayed (Figure 6). Cut planes can be orthogonal, parallel, or free (any given plane orientation), selected as desired by the echocardiographer for obtaining optimized cross-sections of the heart in order to answer specific clinical questions and to perform accurate and reproducible measurements.

897431.fig.007
Figure 7: From the same pyramidal three-dimensional data set, the left ventricle can be visualized using different display modalities: volume rendering, for visualizing morphology and spatial relationships among adjacent structures; surface rendering, for quantitative purposes; and multislice (multiple two-dimensional tomographic views extracted automatically from a single 3D data set) for morphological and functional analysis at different regional levels.
fig8
Figure 8: Examples of volume rendering in mitral stenosis using various depth-encoding colorizations: (a) grey; (c) bronze-blue; (d) copper-blue. (b) illustrates the stereo rendering.
897431.fig.009
Figure 9: Surface rendering display of left ventricular volume. After having identified the endocardial border on the 3 apical slices, the software automatically detects the whole endocardial surface. The accuracy of endocardial border detection on the whole left ventricular circumference can be checked on the transverse slice which can be moved up and down along the left ventricular long axis. Once confirmed by the operator, a 3D cast of the left ventricle is developed and the volume is measured by counting the voxels within the volume.

In the following sections, we will review the current clinical applications of 3DE to assess the geometry and function of the left ventricle and discuss advantages of the technique over conventional 2DE, as well as its limitations.

4. Assessment of Left Ventricular Geometry and Function with 3DE

4.1. Quantification of LV Volumes and Ejection Fraction

To quantitate LV volumes and ejection fraction with 2DE, biplane imaging (i.e., acquisition of both 4- and 2-chamber apical views of the LV) is recommended [19]. However, this approach has several limitations. In many patients, it is difficult to acquire high-quality images in both of these views and to ensure that the imaging planes are truly perpendicular to each other. Moreover, the endocardium should be manually traced in both views and, therefore, LV volume measurement by 2DE is highly dependent on user’s experience. Finally, 2DE uses only partial information contained in 2 predefined cross-sections of the LV to assess global myocardial function and relies on geometrical assumptions about the shape of the LV that may not be necessarily valid in all patients.

The greatest advantage of 3DE in the evaluation of the LV is that, with this technique, the commonest causes of LV volume underestimation with conventional 2DE (i.e., foreshortening of the LV long axis, plane position errors, and geometrical assumptions about LV shape) are no longer real issues (Figure 9) [20, 21]. With 3DE, only one acquisition of the LV is required to obtain volumes and ejection fraction (Figure 10). The acquisition is usually performed from the apical approach and requires that the whole LV is included within the 3D data set. LV data set analysis can be performed using computerized automated or semiautomated endocardial surface detection software, which do not rely on any geometric assumption about LV geometry and require only minimal human intervention, therefore improving measurement reproducibility.

897431.fig.0010
Figure 10: Schematic representation of the biplane disc summation algorithm based on the measurement of left ventricular areas and long axis lengths on 4- and 2-chamber two-dimensional views of the left ventricle. To obtain accurate calculation of left ventricular volume the difference between the long axes measured in 4- and 2-chamber apical views should not be larger than 10%.

3D TTE has been extensively validated against CMR (Table 2) and has been demonstrated to be more time-saving, reproducible, and accurate than conventional 2DE for LV volumes and ejection fraction measurement. In most publications, 3D TTE has been shown to slightly underestimate both LV end-diastolic and end-systolic volumes in comparison with those measured with CMR (Tables 2 and 3). A recent meta-analysis of 23 studies comparing 3D TTE with CMR volumes and ejection fraction demonstrated biases of  mL,  mL, and % for LV end-diastolic and end-systolic volumes, and ejection fraction, respectively [22]. Among the few studies that did not exclude patients for poor image quality, the negative volume biases were slightly larger:  mL,  mL, and % for end-diastolic and end-systolic volumes, and ejection fraction, respectively. As compared to 2D TTE, 3D TTE performed favorably in terms of accuracy and reproducibility. Among the 9 studies including both echocardiographic modalities, negative volume biases were larger for 2D TTE than for 3D TTE: for end-diastolic volume,  mL versus  mL; for end-systolic volume,  mL versus  mL; and for ejection fraction, % versus %.

tab2
Table 2: Differences between left ventricular volumes and ejection fraction assessed by three-dimensional echocardiography and conventional two-dimensional echocardiography in comparison with cardiac magnetic resonance.
tab3
Table 3: Differences between left ventricular mass calculation by three-dimensional echocardiography and conventional two-dimensional echocardiography in comparison with cardiac magnetic resonance.

In another recent study, 36 prospectively enrolled patients underwent 3D TTE, 2D TTE, CMR, CT with contrast, and invasive cine ventriculography [23]. Both 3D TTE and 2D TTE significantly underestimated LV volumes as compared to CMR: end-diastolic volumes were lower by 18 and 26 cc, respectively, while CT slightly overestimated end-diastolic volume by 6 cc. For ejection fraction, both 3D TTE and 2D TTE values correlated equally well with CMR values ( for both), with CT outperforming both ( ). 3D TTE LV ejection fraction values were also less reproducible than those derived from CT. One might speculate that TTE’s apparent disadvantage in this study is related to its operator dependence, during both acquisition and image analysis, given that the CT software automatically determined volumes while the specific 3D TTE software used in this study (4D LV function, TomTec Imaging Systems, Unterschleißheim, D) requires tracings of the endocardial borders at both end-diastole and end-systole.

The systematic underestimation of LV volumes by 3D TTE stems in part from its lower spatial resolution as compared to CMR. By convention, when CMR volumes are traced according to Simpson’s method of disks, the trabeculae are included in the LV cavity. With 3D TTE, it is often difficult to clearly identify the endocardial-trabecular border, and as a result, one may instead trace the blood-trabecular interface. In a systematic review of sources of error, it was shown that agreement between 3D TTE and CMR measurements improved when the trabeculae were excluded from the LV cavity [24]. In this study, agreement between 3D TTE and CMR also improved with increasing investigator experience; experienced observers tended to place their 3D TTE borders as far outward as possible.

A recent meta-analysis of studies evaluating 3D TTE for LV volumes demonstrated that 3D TTE was superior to older techniques requiring transducer rotation and reconstruction [22]. Although single-beat acquisition for quantification of LV volumes is feasible [25, 26], its low temporal resolution results in underestimation of EF as compared to 2- and 4-beat acquisition [11]. This is due to the misidentification of the end-systolic frame at low frame rates. The method of data analysis also affects accuracy. Most current software programs perform semiautomated quantification of volumes, requiring the user to identify anatomical landmarks (usually including the apex and mitral annulus) on one or more longitudinal slices of the LV derived from the 3D dataset. The user is then able to edit the endocardial borders displayed by the program. Several studies have shown that when endocardial borders are traced and/or adjusted on more than 2 cut planes, accuracy of volume quantification is improved [2729]. This is likely due to the fact that fewer geometric assumptions are required. With programs that perform automated border detection, software sensitivity settings also influence volumes. Lower sensitivity causes the border to be drawn closer to the endocardial-trabecular border, rather than at the blood-trabecular interface, increasing volumes and resulting in less bias as compared to CMR [30]. Despite all the advantages of 3D TTE, its use remains limited mostly to academic centers. Acquisition and processing of 3D datasets require specialized skills, which sonographers and physicians may not obtain during their formal training. Moreover, quantification of LV volumes and ejection fraction with currently available 3D software is time consuming. Fortunately, software programs that perform fully automated LV endocardial detection are being developed and appear promising [31]; their ease of use will likely expand the role of 3D TTE in everyday practice (Figure 11).

897431.fig.0011
Figure 11: Single-beat full-volume acquisition and automatic endocardial border detection allow fast assessment of left ventricular volume and ejection fraction with the possibility of averaging results over multiple cardiac beats.

3D TTE is a robust technique for quantifying LV volumes and ejection fraction in patients with abnormal LV geometry. Regional wall motion abnormalities, including those due to myocardial infarction, do not compromise the accuracy of LV volume and ejection fraction measurement [3235]. 3D TTE volumes are generally accurate even in very dilated and aneurysmal ventricles [3638], provided that the sonographer takes care to include the entire ventricle in the pyramidal data set.

In a recent meta-analysis Dorosz et al. [39] found significantly larger biases and limits of agreement for 2D TTE (  mL,  mL, and % for LV volumes and ejection fraction, resp.) than for 3D TTE ( ML,  mL, and %) when both were compared with CMR. These data, together with the superior reproducibility of 3D TTE, should warrant its use as the preferred echo technique to assess LV size and function.

For years, the usefulness of 3D TTE in everyday practice was limited by the absence of reference values for LV chamber volumes and ejection fraction. Recently, several publications [4044] have addressed this gap in the literature (Table 4). Some of the variation in the reference ranges from study to study is likely attributable to differences in echocardiographic equipment and analysis software, as well as heterogeneity in measurement techniques [45]. Despite the fact that the LV volumes obtained in normal subjects are significantly larger with 3DE than with 2DE, the LV ejection fraction is similar [41]. The largest of these studies [44] explicitly stratified subjects according to ethnicity, finding that LV volumes were smaller among Asian Indians than White Europeans, while ejection fraction did not differ. The EchoNORMAL (the echocardiographic normal ranges meta-analysis of the left heart) collaboration study carried out by the University of Aukland (New Zealand) is a meta-analysis of echo measurements obtained from 23301 normal subjects collected from several echo labs around the world and will provide ethnicity specific reference values for most of the conventional and 3D echocardiographic measurements [46]. In most of the studies that measured LV geometry and function in normal subjects [40, 41, 43, 44], weak to moderate negative correlations were seen between age and LV volumes, while LV ejection fraction showed either no change or a significant increase with age [41]. These findings are in keeping with those of prior CMR studies [47, 48]. The ongoing normal reference ranges for echocardiography (NORRE) study has been designed to establish normal values for a variety of 2D and 3D TTE parameters in a population of 1100 Caucasian Europeans, ranging in age from 25 to 75 [49].

tab4
Table 4: Published reference values for three-dimensional echocardiography derived left ventricular geometry and function parameters.
4.2. Assessment of Regional Left Ventricular Function

Unlike 2D TTE, which requires the sonographer to move and rotate the probe in order to acquire images of all LV walls from the parasternal and apical windows, 3D TTE can image all walls in a single volume acquisition. Moreover, foreshortening is not a problem anymore with 3D TTE, and 2D slices obtained from the postprocessing of the 3D data set can be rotated and reoriented at anytime after acquisition. In addition, a number of 2D views (slices) can be simultaneously displayed from the acquired 3D data set, providing a comprehensive visualization of LV endocardial motion and myocardial thickening (Figure 7). For these reasons, 3D TTE is particularly well-suited to assess the location and extent of regional wall motion abnormalities of the LV. Regional LV volumes by 3D TTE (Figure 12) have been shown to correlate well with CMR ( values generally 0.8 and higher) [50]. Thorstensen et al. [51] reported in patients with recent myocardial infarction a reasonable correlation ( ) between wall motion score index by 3D TTE and the extent of delayed gadolinium enhancement by CMR. Among patients with suboptimal 3DTTE image quality, administration of echocardiographic contrast has been shown to improve agreement of wall motion assessment with CMR [52].

897431.fig.0012
Figure 12: The full-volume of the left ventricle can be divided into 16 or 17 regional pyramidal subvolumes whose base is on the endocardium (different colours identify different segments) and the apex is in the center of gravity within the cavity of the left ventricle. The volume changes of individual regional pyramidal subvolume can be tracked throughout the cardiac cycle (see the time-volume curves in the lower part of the figure) and a “regional ejection fraction” can be measured.

In the area of stress echocardiography, 3D TTE is advantageous because it requires only a single apical acquisition at peak stress, whereas 2D TTE requires at least 3 (apical 2-, 3-, and 4-chamber views). This would be particularly useful in patients with rapid heart rate recovery after exercise. Furthermore, during offline postprocessing, it is possible to view images in a multitude of planes and reorient them to determine the extent of wall motion abnormalities and to rule out artifacts. A number of studies have demonstrated that exercise and pharmacologic stress testing with 3D TTE are feasible, with more rapid image acquisition and higher interobserver agreement in wall motion interpretation as compared to 2D TTE [5358]. The sensitivity of 3D TTE to significant coronary artery disease, based on an invasive angiography reference standard, has been found equal to or better than that of 2D TTE in most [53, 55, 57] but not in all studies [59]. Inadequate segment visualization is most common in the anterior and lateral walls [58], but administration of an ultrasound contrast agent may significantly improve image quality [60]. In order to optimize detection of wall motion abnormalities, it is important to maximize temporal resolution (i.e., use a volume rate as high as possible) and to ensure adequate breath holds to avoid artifacts. Specialized software that helps the echocardiographer identify corresponding planes at rest and stress, so that each wall may be evaluated against itself, has been shown to improve interobserver agreement for wall motion analysis [61].

4.3. Quantification of Left Ventricular Mass

For quantification of LV mass, both the endocardial and epicardial surfaces must be outlined to measure LV myocardial volume (Figure 13). By convention, as with CMR tracings, the papillary muscles are part of the LV cavity (i.e., not included in the myocardial volume). LV myocardial volume is then multiplied by the density of cardiac muscle (1.05 g/mL) to calculate LV mass. With 3D TTE datasets, LV mass analysis is typically done offline using semiautomated software (Figure 13).

fig13
Figure 13: Left ventricular mass measurement using three-dimensional echocardiography. Using automated or semiautomated endocardial and epicardial boundary detection endocardial and epicardial volumes are measured (a). By subtracting the left ventricular cavity volume from the epicardial volume, the volume of myocardium is obtained (b). Left ventricular mass is calculated by multiplying myocardial volume by its specific gravity (1.05 g/cm3).

Several studies have compared 3D TTE to 2D TTE and CMR and have generally found that LV mass measured by 3D TTE correlates better with CMR results [6272]. Unlike 2D TTE, 3D TTE does not systematically underestimate mass; likely because foreshortening of the LV long-axis dimension is generally avoided [63]. Furthermore, 3D TTE has better interobserver agreement than 2D TTE [63, 64, 69]. Even in the presence of wall motion abnormalities, abnormal LV geometry secondary to congenital heart disease or hypertrophic cardiomyopathy, 3D TTE-based measurement of LV mass is relatively accurate as compared to CMR [32, 66, 70] although a recent meta-analysis has suggested that LV mass measurements are more likely to be underestimated in patients with cardiac disease than in healthy volunteers [67].

Interestingly, M-mode echocardiography, a method commonly used to diagnose LV hypertrophy in clinical practice, has been shown in some studies to overestimate LV mass in comparison to 3D TTE [41, 7274]. This is likely because M-mode relies heavily on geometric assumptions and is greatly influenced by the acquisition plane. M-mode wall thickness measurements are usually obtained near the base of the heart at the level of the papillary muscles, where the LV wall tends to be thickest, whereas the actual wall thickness decreases gradually from base to apex. In a small study of LV mass in patients with abnormal LV geometry due to congenital heart disease, M-mode correlated poorly with CMR ( ), while 3D TTE performed well ( ) [75].

Two studies have reported reference values for 3D LV mass and LV mass/end-diastolic volume ratio in Japanese and Italian cohorts (Table 3) [40, 41].

4.4. Assessment of LV Dyssynchrony

The availability of cardiac resynchronization therapy (CRT) for refractory heart failure has generated considerable interest in LV intraventricular dyssynchrony assessment. For quantitative dyssynchrony evaluation, the LV is most commonly divided into 16 or 17 segments as per the American Heart Association standard model [76], and the time to minimum systolic volume is determined for each LV segment. The standard deviation of the regional times to minimum systolic volume has been proposed as the dyssynchrony index [77]. Several small studies have suggested that an increased dyssynchrony index prior to implant is predictive of favorable response to CRT [7881] and that placement of the LV lead near the most delayed segment may be associated with favorable LV remodeling [82], but these findings have not been consistent [83, 84]. It is clear that greater degrees of LV dysfunction are associated with higher dyssynchrony indices [77, 8587], but this may be due to noisy, low-amplitude ejection curves as much as true dyssynchrony [21, 85]. Moreover, inter- and intraobserver reproducibility of the 3D dyssynchrony index is hindered by suboptimal image quality [78] and is less robust than that of tissue Doppler imaging [87, 88]. The low temporal resolution of 3D TTE as compared to tissue Doppler imaging is another important consideration in this context [89]. However, the lower temporal resolution does not seem to be a major drawback of second-generation 3DE scanners for LV dyssynchrony assessment [30]. Since 3DE measures regional volume changes and not regional velocities (as tissue Doppler imaging, which requires a high frame rate), it seems likely that echo systems using a temporal resolution of 20 to 30 volumes/s are adequate to sample the usual frequency of regional volume curves which is less than 10 Hz [90].

Finally, a reliable, clinically meaningful cutoff value for the 3D dyssynchrony index has yet to be established [89]. Larger, multicenter studies are needed to establish the role of 3D TTE dyssynchrony analysis in the evaluation of patients prior to CRT.

4.5. Novel (Research) Applications of 3D TTE for Advanced Assessment of the Left Ventricle

3D strain is a novel technology aimed at assessing LV myocardial deformation by analyzing the motion of myocardial speckles within the 3D LV data sets. This technology allows following the motion of speckles in the space without any assumption about the direction of their motion and the measurement of all LV deformation components (longitudinal, radial, and circumferential) plus the computation of the area strain (a composite one which includes the longitudinal and circumferential deformation) from a single apical LV 3D data set. Theoretically, 3D speckle-tracking technology should overcome the main limitations of 2D speckle tracking: the “out-of-plane” motion of speckles due to the rotation and shortening motion of the left ventricle and the need to interpolate the whole LV myocardium from the partial information contained in a limited number of thin, tomographic slices of the LV (Figure 14).

fig14
Figure 14: Three-dimensional speckle-tracking analysis of left ventricular longitudinal myocardial deformation using two different platforms. Results can be displayed as bull’s eye plots and/or time-strain curves.

3D TTE strain acquired using speckle tracking has been studied in several clinical contexts. In patients surviving an acute myocardial infarction, higher baseline global longitudinal strain by 3D TTE is independently predictive of improvement in LV ejection fraction at 6-month follow-up [91]. Impaired circumferential, radial, and longitudinal strains have been associated with LV dilation in the setting of ischemic cardiomyopathy with reduced LV ejection fraction [92]. In a population of diabetic patients with normal LV ejection fraction, elevated hemoglobin A1c correlated with reduced global longitudinal, circumferential, and area strain [93]. A study of heart transplant recipients with preserved LV ejection fraction showed that 3D speckle-tracking global longitudinal and circumferential strain, but not 2D strain, were predictive of New York Heart Association functional class [94]. These findings suggest that 3D speckle-tracking strain could be useful in predicting outcomes and detecting subclinical disease in a variety of cardiac and systemic conditions.

To date, the main issues that are currently limiting the applicability of 3D strain in the clinical arena are (i) lack of reference values for the regional and global strain parameters for the various deformation components; (ii) lack of multicenter outcome studies assessing its additive prognostic value over the well-established 2D strain; and (iii) significant intervendor differences of 3D strain algorithms and values among vendors, preventing its routine applicability for clinical purposes [95].

Another application of 3DE that may open a complete new way to assess prognosis in patients with myocardial and valvular heart diseases is the quantitative analysis of LV shape. 3D TTE is well-suited to portray global and regional shape of the LV [96]. A 3D TTE study by Salgo et al. [97] demonstrated that, in patients with dilated cardiomyopathy, the apical and septal curvatures were diminished as compared to controls, reflecting the ventricle’s overall globular shape. Mannaerts et al. [98] have described a 3D sphericity index that, in a postmyocardial infarction population, is highly predictive of adverse remodeling (progressive LV dilation). In a 3D TTE study of patients with severe mitral regurgitation, sphericity was increased as compared to controls, even in the presence of only mild LV dilation and normal ejection fraction; following successful mitral valve repair, sphericity decreased [99].

5. Conclusions

3DE is a novel imaging technique based on acquisition and display of volumetric data sets in the beating heart. This permits a comprehensive evaluation of LV anatomy and function from a single acquisition and expands the diagnostic possibilities of noninvasive cardiology. It provides the possibility of quantitating geometry and function of LV without preestablished assumptions regarding cardiac chamber shape and allows an echocardiographic assessment of the LV that is less operator-dependent and therefore more reproducible.

Further developments and improvements for widespread routine applications include higher spatial and temporal resolution to improve image quality, faster acquisition, processing and reconstruction, and fully automated quantitative analysis. At present, 3DE complements routine 2DE in clinical practice, overcoming some of its limitations and offering additional valuable information that has led to recommending its use for routine assessment of the LV of patients in whom information about LV size and function is critical for their clinical management.

Abbreviations

2DE:Two-dimensional echocardiography
3DE:Three-dimensional echocardiography
CMR:Cardiac magnetic resonance
CRT:Cardiac resynchronization therapy
CT:Cardiac computerized tomography
ECG:Electrocardiographic
LV:Left ventricle/ventricular
TTE:Transthoracic echocardiography
TOE:Transoesophageal echocardiography.

Conflict of Interests

The author declares that there is no conflict of interests regarding the publication of this paper.

Acknowledgments

The author would like to thank Dr. Denisa Muraru M.D, first, for her thoughtful review of the paper and for having provided the figures and, second, for having run the 3D echo research project at the University of Padua. Without her enthusiasm and commitment the project would have not even started and would have not been so successful.

References

  1. A. Vahanian, O. Alfieri, F. Andreotti, et al., “Guidelines on the management of valvular heart disease (version 2012): The Joint Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS),” European Heart Journal, vol. 33, pp. 2451–2496, 2012.
  2. J. J. McMurray, S. Adamopoulos, S. D. Anker, et al., “ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2012: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESC,” European Heart Journal, vol. 33, pp. 1787–1847, 2012.
  3. D. Bovelli, G. Plataniotis, and F. Roila, “Cardiotoxicity of chemotherapeutic agents and radiotherapy-related heart disease: ESMO clinical practice guidelines,” Annals of Oncology, vol. 21, supplement 5, pp. v277–v282, 2010. View at Publisher · View at Google Scholar · View at Scopus
  4. G. Steg, S. K. James, D. Atar, et al., “Grupo de Trabajo para el manejo del infarto agudo de miocardio con elevacion del segmento ST de la Sociedad Europea de Cardiologia (ESC),” Revista Española de Cardiología, vol. 66, pp. 53.e1–53.e46, 2013.
  5. L. P. Badano, D. Muraru, and M. Enriquez-Sarano, “Assessment of functional tricuspid regurgitation,” European Heart Journal, vol. 34, pp. 1875–1885, 2013.
  6. D. Muraru, L. P. Badano, C. Sarais, E. Soldà, and S. Iliceto, “Evaluation of tricuspid valve morphology and function by transthoracic three-dimensional echocardiography,” Current Cardiology Reports, vol. 13, no. 3, pp. 242–249, 2011. View at Publisher · View at Google Scholar · View at Scopus
  7. D. Muraru, L. P. Badano, M. Vannan, and S. Iliceto, “Assessment of aortic valve complex by three-dimensional echocardiography: a framework for its effective application in clinical practice.,” European Heart Journal Cardiovascular Imaging, vol. 13, pp. 541–555, 2012.
  8. D. Muraru, M. Cattarina, F. Boccalini, et al., “Mitral valve anatomy and function: new insights from three-dimensional echocardiography,” Journal of Cardiovascular Medicine, vol. 14, pp. 91–99, 2013.
  9. D. Muraru, M. F. Tuveri, M. P. Marra, L. P. Badano, and S. Iliceto, “Carcinoid tricuspid valve disease: incremental value of three-dimensional echocardiography,” European Heart Journal Cardiovascular Imaging, vol. 13, no. 4, p. 329, 2012. View at Publisher · View at Google Scholar
  10. L. P. Badano, E. Dall'Armellina, M. J. Monaghan et al., “Real-time three-dimensional echocardiography: technological gadget or clinical tool?” Journal of Cardiovascular Medicine, vol. 8, no. 3, pp. 144–162, 2007. View at Publisher · View at Google Scholar · View at Scopus
  11. F. Maffessanti, D. Muraru, R. Esposito, et al., “Age-, body size-, and sex-specific reference values for right ventricular volumes and ejection fraction by three-dimensional echocardiography: a multicenter echocardiographic study in 507 healthy volunteers,” Circulation. Cardiovascular Imaging, vol. 6, no. 5, pp. 700–10, 2013.
  12. D. Peluso, L. P. Badano, D. Muraru, et al., “Right atrial size and function assessed with three-dimensional and speckle-tracking echocardiography in 200 healthy volunteers,” European Heart Journal Cardiovascular Imaging, vol. 14, no. 11, pp. 1106–1114, 2013.
  13. L. P. Badano, F. Boccalini, D. Muraru, et al., “Current clinical applications of transthoracic three-dimensional echocardiography.,” Journal of Cardiovascular Ultrasound, vol. 20, pp. 1–22, 2012.
  14. L. P. Badano, C. Ginghina, J. Easaw et al., “Right ventricle in pulmonary arterial hypertension: faemodynamics, structural changes, imaging, and proposal of a study protocol aimed to assess remodelling and treatment effects,” European Journal of Echocardiography, vol. 11, no. 1, pp. 27–37, 2010. View at Publisher · View at Google Scholar · View at Scopus
  15. R. M. Lang, L. P. Badano, W. Tsang, et al., “EAE/ASE recommendations for image acquisition and display using three-dimensional echocardiography,” European Heart Journal Cardiovascular Imaging, vol. 13, pp. 1–46, 2012.
  16. M. Jiwa, S. Millett, X. Meng, and V. M. Hewitt, “Impact of the presence of medical equipment in images on viewer's perceptions of the trustworthiness of an individual on-screen,” Journal of Medical Internet Research, vol. 14, article e100, 2012.
  17. H. Dalen, B. O. Haugen, and T. Graven, “Feasibility and clinical implementation of hand-held echocardiography,” Expert Review of Cardiovascular Therapy, vol. 11, pp. 49–54, 2013.
  18. J. M. Duffin, “The cardiology of R.T.H. Laennec,” Medical History, vol. 33, no. 1, pp. 42–71, 1989. View at Scopus
  19. R. M. Lang, M. Bierig, R. B. Devereux, et al., “Recommendations for chamber quantification,” European Journal of Echocardiography, vol. 7, pp. 79–108, 2006.
  20. E. O. Chukwu, E. Barasch, D. G. Mihalatos et al., “Relative importance of errors in left ventricular quantitation by two-dimensional echocardiography: insights from three-dimensional echocardiography and cardiac magnetic resonance imaging,” Journal of the American Society of Echocardiography, vol. 21, no. 9, pp. 990–997, 2008. View at Publisher · View at Google Scholar · View at Scopus
  21. V. Mor-Avi and R. M. Lang, “The use of real-time three-dimensional echocardiography for the quantification of left ventricular volumes and function,” Current Opinion in Cardiology, vol. 24, no. 5, pp. 402–409, 2009. View at Publisher · View at Google Scholar · View at Scopus
  22. Y. J. Shimada and T. Shiota, “A meta-analysis and investigation for the source of bias of left ventricular volumes and function by three-dimensional echocardiography in comparison with magnetic resonance imaging,” The American Journal of Cardiology, vol. 107, no. 1, pp. 126–138, 2011. View at Publisher · View at Google Scholar · View at Scopus
  23. J. Greupner, E. Zimmermann, A. Grohmann, et al., “Head-to-head comparison of left ventricular function assessment with 64-row computed tomography, biplane left cineventriculography, and both 2- and 3-dimensional transthoracic echocardiography: comparison with magnetic resonance imaging as the reference standard,” Journal of the American College of Cardiology, vol. 59, pp. 1897–1907, 2012.
  24. V. Mor-Avi, C. Jenkins, H. P. Kühl et al., “Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error,” JACC: Cardiovascular Imaging, vol. 1, no. 4, pp. 413–423, 2008. View at Publisher · View at Google Scholar · View at Scopus
  25. S. A. Chang, S. C. Lee, E. Y. Kim et al., “Feasibility of single-beat full-volume capture real-time three-dimensional echocardiography and auto-contouring algorithm for quantification of left ventricular volume: Validation with cardiac magnetic resonance imaging,” Journal of the American Society of Echocardiography, vol. 24, no. 8, pp. 853–859, 2011. View at Publisher · View at Google Scholar · View at Scopus
  26. L. Macron, P. Lim, A. Bensaid et al., “Single-beat versus multibeat real-time 3d echocardiography for assessing left ventricular volumes and ejection fraction: a comparison study with cardiac magnetic resonance,” Circulation: Cardiovascular Imaging, vol. 3, no. 4, pp. 450–455, 2010. View at Publisher · View at Google Scholar · View at Scopus
  27. O. I. I. Soliman, B. J. Krenning, M. L. Geleijnse et al., “A comparison between QLAB and tomtec full volume reconstruction for real time three-dimensional echocardiographic quantification of left ventricular volumes,” Echocardiography, vol. 24, no. 9, pp. 967–974, 2007. View at Publisher · View at Google Scholar · View at Scopus
  28. F. Li, Q. Wang, G. H. Yao et al., “Impact of the number of image planes of real-time three-dimensional echocardiography on the accuracy of left atrial and ventricular volume measurements,” Ultrasound in Medicine and Biology, vol. 34, no. 1, pp. 40–46, 2008. View at Publisher · View at Google Scholar · View at Scopus
  29. D. Muraru, L. P. Badano, G. Piccoli et al., “Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography,” European Journal of Echocardiography, vol. 11, no. 4, pp. 359–368, 2010. View at Publisher · View at Google Scholar · View at Scopus
  30. D. Muraru, L. P. Badano, D. Ermacora, G. Piccoli, and S. Iliceto, “Sources of variation and bias in assessing left ventricular volumes and dyssynchrony using three-dimensional echocardiography,” International Journal of Cardiovascular Imaging, vol. 28, pp. 1357–1368, 2012. View at Publisher · View at Google Scholar · View at Scopus
  31. P. Thavendiranathan, S. Liu, D. Verhaert et al., “Feasibility, accuracy, and reproducibility of real-time full-volume 3D transthoracic echocardiography to measure LV volumes and systolic function: a fully automated endocardial contouring algorithm in sinus rhythm and atrial fibrillation,” JACC: Cardiovascular Imaging, vol. 5, no. 3, pp. 239–251, 2012. View at Publisher · View at Google Scholar · View at Scopus
  32. A. C. Pouleur, J. B. le Polain de Waroux, A. Pasquet et al., “Assessment of left ventricular mass and volumes by three-dimensional echocardiography in patients with or without wall motion abnormalities: comparison against cine magnetic resonance imaging,” Heart, vol. 94, no. 8, pp. 1050–1057, 2008. View at Publisher · View at Google Scholar · View at Scopus
  33. J. Chan, C. Jenkins, F. Khafagi, L. Du, and T. H. Marwick, “What is the optimal clinical technique for measurement of left ventricular volume after myocardial infarction? A comparative study of 3-dimensional echocardiography, single photon emission computed tomography, and cardiac magnetic resonance imaging,” Journal of the American Society of Echocardiography, vol. 19, no. 2, pp. 192–201, 2006. View at Publisher · View at Google Scholar · View at Scopus
  34. A. S. Gopal, E. O. Chukwu, D. G. Mihalatos et al., “Left ventricular structure and function for postmyocardial infarction and heart failure risk stratification by three-dimensional echocardiography,” Journal of the American Society of Echocardiography, vol. 20, no. 8, pp. 949–958, 2007. View at Publisher · View at Google Scholar · View at Scopus
  35. K. Arai, T. Hozumi, Y. Matsumura et al., “Accuracy of measurement of left ventricular volume and ejection fraction by new real-time three-dimensional echocardiography in patients with wall motion abnormalities secondary to myocardial infarction,” The American Journal of Cardiology, vol. 94, no. 5, pp. 552–558, 2004. View at Publisher · View at Google Scholar · View at Scopus
  36. L. Sugeng, V. Mor-Avi, L. Weinert et al., “Quantitative assessment of left ventricular size and function: side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference,” Circulation, vol. 114, no. 7, pp. 654–661, 2006. View at Publisher · View at Google Scholar · View at Scopus
  37. N. A. Marsan, J. J. M. Westenberg, S. D. Roes et al., “Three-dimensional echocardiography for the preoperative assessment of patients with left ventricular aneurysm,” Annals of Thoracic Surgery, vol. 91, no. 1, pp. 113–121, 2011. View at Publisher · View at Google Scholar · View at Scopus
  38. J. X. Qin, M. Jones, T. Shiota et al., “Validation of real-time three-dimensional echocardiography for quantifying left ventricular volumes in the presence of a left ventricular aneurysm: in vitro and in vivo studies,” Journal of the American College of Cardiology, vol. 36, no. 3, pp. 900–907, 2000. View at Publisher · View at Google Scholar · View at Scopus
  39. J. L. Dorosz, D. C. Lezotte, D. A. Weitzenkamp, L. A. Allen, and E. E. Salcedo, “. Performance of 3-dimensional echocardiography in measuring left ventricular volumes and ejection fraction: a systematic review and meta-analysis.,” Journal of the American College of Cardiology, vol. 59, pp. 1799–1808, 2012.
  40. S. Fukuda, H. Watanabe, M. Daimon et al., “Normal values of real-time 3-dimensional echocardiographic parameters in a healthy Japanese population: the JAMP-3D study,” Circulation Journal, vol. 76, no. 5, pp. 1177–1181, 2012. View at Publisher · View at Google Scholar · View at Scopus
  41. D. Muraru, L. P. Badano, D. Peluso, et al., “Comprehensive analysis of left ventricular geometry and function by three-dimensional echocardiography in healthy adults,” Journal of the American Society of Echocardiography, vol. 26, pp. 618–628, 2013.
  42. E. Aune, M. Bækkevar, O. Rødevand, and J. E. Otterstad, “Reference values for left ventricular volumes with real-time 3-dimensional echocardiography,” Scandinavian Cardiovascular Journal, vol. 44, no. 1, pp. 24–30, 2010. View at Publisher · View at Google Scholar · View at Scopus
  43. K. Kaku, M. Takeuchi, K. Otani et al., “Age- and gender-dependency of left ventricular geometry assessed with real-time three-dimensional transthoracic echocardiography,” Journal of the American Society of Echocardiography, vol. 24, no. 5, pp. 541–547, 2011. View at Publisher · View at Google Scholar · View at Scopus
  44. N. S. Chahal, T. K. Lim, P. Jain, J. C. Chambers, J. S. Kooner, and R. Senior, “Population-based reference values for 3D echocardiographic LV volumes and ejection fraction,” JACC: Cardiovascular Imaging, vol. 5, pp. 1191–1197, 2012.
  45. L. P. Badano, “Defining normative values for 3D LV volumes: the devil is in the details,” JACC: Cardiovascular Imaging, vol. 6, 530 pages, 2013.
  46. K. K. Poppe, R. N. Doughty, and G. A. Whalley, “Redefining normal reference ranges for echocardiography: a major new individual person data meta-analysis,” European Heart Journal Cardiovascular Imaging, vol. 14, pp. 347–348, 2013.
  47. T. Germans, M. J. W. Götte, R. Nijveldt et al., “Effects of aging on left atrioventricular coupling and left ventricular filling assessed using cardiac magnetic resonance imaging in healthy subjects,” The American Journal of Cardiology, vol. 100, no. 1, pp. 122–127, 2007. View at Publisher · View at Google Scholar · View at Scopus
  48. J. Sandstede, C. Lipke, M. Beer et al., “Age- and gender-specific differences in left and right ventricular cardiac function and mass determined by cine magnetic resonance imaging,” European Radiology, vol. 10, no. 3, pp. 438–442, 2000. View at Scopus
  49. P. Lancellotti, L. P. Badano, R. M. Lang, et al., “Normal Reference Ranges for Echocardiography: rationale, study design, and methodology (NORRE Study),” European Heart Journal Cardiovascular Imaging, vol. 14, pp. 303–308, 2013.
  50. C. Sohns, J. M. Sohns, D. Vollman, et al., “Left atrial volumetry from routine diagnostic work up prior to pulmonary vein ablation is a good predictor of freedom from atrial fibrillation,” European Heart Journal Cardiovascular Imaging, vol. 14, no. 7, pp. 684–691, 2013. View at Publisher · View at Google Scholar
  51. A. Thorstensen, H. Dalen, P. Hala, et al., “Three-dimensional echocardiography in the evaluation of global and regional function in patients with recent myocardial infarction: a comparison with magnetic resonance imaging,” Echocardiography, vol. 30, pp. 682–692, 2013.
  52. C. Corsi, P. Coon, S. Goonewardena et al., “Quantification of regional left ventricular wall motion from real-time 3-dimensional echocardiography in patients with poor acoustic windows: effects of contrast enhancement tested against cardiac magnetic resonance,” Journal of the American Society of Echocardiography, vol. 19, no. 7, pp. 886–893, 2006. View at Publisher · View at Google Scholar · View at Scopus
  53. L. P. Badano, D. Muraru, F. Rigo et al., “High volume-rate three-dimensional stress echocardiography to assess inducible myocardial ischemia: a feasibility study,” Journal of the American Society of Echocardiography, vol. 23, no. 6, pp. 628–635, 2010. View at Publisher · View at Google Scholar · View at Scopus
  54. Y. Matsumura, T. Hozumi, K. Arai et al., “Non-invasive assessment of myocardial ischaemia using new real-time three-dimensional dobutamine stress echocardiography: comparison with conventional two-dimensional methods,” European Heart Journal, vol. 26, no. 16, pp. 1625–1632, 2005. View at Publisher · View at Google Scholar · View at Scopus
  55. S. Kennon, C. P. Price, P. G. Mills et al., “Real-time three-dimensional dobutamine stress echocardiography in assessment of ischemia: comparison with two-dimensional dobutamine stress echocardiography,” Journal of the American College of Cardiology, vol. 37, no. 5, pp. 1303–1309, 2001. View at Publisher · View at Google Scholar · View at Scopus
  56. V. Walimbe, M. Garcia, O. Lalude, J. Thomas, and R. Shekhar, “Quantitative real-time 3-dimensional stress echocardiography: a preliminary investigation of feasibility and effectiveness,” Journal of the American Society of Echocardiography, vol. 20, no. 1, pp. 13–22, 2007. View at Publisher · View at Google Scholar · View at Scopus
  57. C. Aggeli, G. Giannopoulos, P. Misovoulos et al., “Real-time three-dimensional dobutamine stress echocardiography for coronary artery disease diagnosis: validation with coronary angiography,” Heart, vol. 93, no. 6, pp. 672–675, 2007. View at Publisher · View at Google Scholar · View at Scopus
  58. M. Takeuchi, S. Otani, L. Weinert, K. T. Spencer, and R. M. Lang, “Comparison of contrast-enhanced real-time live 3-dimensional dobutamine stress echocardiography with contrast 2-dimensional echocardiography for detecting stress-induced wall-motion abnormalities,” Journal of the American Society of Echocardiography, vol. 19, no. 3, pp. 294–299, 2006. View at Publisher · View at Google Scholar · View at Scopus
  59. C. Jenkins, B. Haluska, and T. H. Marwick, “Assessment of temporal heterogeneity and regional motion to identify wall motion abnormalities using treadmill exercise stress three-dimensional echocardiography,” Journal of the American Society of Echocardiography, vol. 22, no. 3, pp. 268–275, 2009. View at Publisher · View at Google Scholar · View at Scopus
  60. L. P. Badano, “Contrast enhanced real-time three-dimensional echocardiography for quantification of myocardial perfusion: a step forward,” European Journal of Echocardiography, vol. 10, no. 4, pp. 465–466, 2009. View at Publisher · View at Google Scholar · View at Scopus
  61. A. Nemes, K. Y. E. Leung, G. van Burken et al., “Side-by-side viewing of anatomically aligned left ventricular segments in three-dimensional stress echocardiography,” Echocardiography, vol. 26, no. 2, pp. 189–195, 2009. View at Publisher · View at Google Scholar · View at Scopus
  62. E. G. Caiani, C. Corsi, L. Sugeng et al., “Improved quantification of left ventricular mass based on endocardial and epicardial surface detection with real time three dimensional echocardiography,” Heart, vol. 92, no. 2, pp. 213–219, 2006. View at Publisher · View at Google Scholar · View at Scopus
  63. V. Mor-Avi, L. Sugeng, L. Weinert et al., “Fast measurement of left ventricular mass with real-time three-dimensional echocardiography: comparison with magnetic resonance imaging,” Circulation, vol. 110, no. 13, pp. 1814–1818, 2004. View at Publisher · View at Google Scholar · View at Scopus
  64. G. Pacileo, B. Castaldi, G. di Salvo et al., “Assessment of left-ventricular mass and remodeling in obese adolescents: M-mode, 2D or 3D echocardiography?” Journal of Cardiovascular Medicine, vol. 14, pp. 144–149, 2013. View at Publisher · View at Google Scholar · View at Scopus
  65. X. Q. Jian, M. Jones, A. Travaglini et al., “The accuracy of left ventricular mass determined by real-time three-dimensional echocardiography in chronic animal and clinical studies: a comparison with postmortem examination and magnetic resonance imaging,” Journal of the American Society of Echocardiography, vol. 18, no. 10, pp. 1037–1043, 2005. View at Publisher · View at Google Scholar · View at Scopus
  66. S. A. Chang, H. K. Kim, S. C. Lee, et al., “Assessment of left ventricular mass in hypertrophic cardiomyopathy by real-time three-dimensional echocardiography using single-beat capture image,” Journal of the American Society of Echocardiography, vol. 26, pp. 436–442, 2013.
  67. Y. J. Shimada and T. Shiota, “Meta-analysis of accuracy of left ventricular mass measurement by three-dimensional echocardiography,” The American Journal of Cardiology, vol. 110, pp. 445–452, 2012. View at Publisher · View at Google Scholar · View at Scopus
  68. M. Lenstrup, J. Kjaergaard, C. L. Petersen, A. Kjaer, and C. Hassager, “Evaluation of left ventricular mass measured by 3D echocardiography using magnetic resonance imaging as gold standard,” Scandinavian Journal of Clinical and Laboratory Investigation, vol. 66, no. 8, pp. 647–657, 2006. View at Publisher · View at Google Scholar · View at Scopus
  69. S. C. Yap, R. J. M. van Geuns, A. Nemes et al., “Rapid and accurate measurement of LV mass by biplane real-time 3D echocardiography in patients with concentric LV hypertrophy: comparison to CMR,” European Journal of Echocardiography, vol. 9, no. 2, pp. 255–260, 2008. View at Publisher · View at Google Scholar · View at Scopus
  70. A. E. van den Bosch, D. Robbers-Visser, B. J. Krenning et al., “Comparison of real-time three-dimensional echocardiography to magnetic resonance imaging for assessment of left ventricular mass,” The American Journal of Cardiology, vol. 97, no. 1, pp. 113–117, 2006. View at Publisher · View at Google Scholar · View at Scopus
  71. H. Oe, T. Hozumi, K. Arai et al., “Comparison of accurate measurement of left ventricular mass in patients with hypertrophied hearts by real-time three-dimensional echocardiography versus magnetic resonance imaging,” The American Journal of Cardiology, vol. 95, no. 10, pp. 1263–1267, 2005. View at Publisher · View at Google Scholar · View at Scopus
  72. C. Jenkins, K. Bricknell, L. Hanekom, and T. H. Marwick, “Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography,” Journal of the American College of Cardiology, vol. 44, no. 4, pp. 878–886, 2004. View at Publisher · View at Google Scholar · View at Scopus
  73. M. Takeuchi, T. Nishikage, V. Mor-Avi et al., “Measurement of left ventricular mass by real-time three-dimensional echocardiography: validation against magnetic resonance and comparison with two-dimensional and m-mode measurements,” Journal of the American Society of Echocardiography, vol. 21, no. 9, pp. 1001–1005, 2008. View at Publisher · View at Google Scholar · View at Scopus
  74. H. P. Kühl, P. Hanrath, and A. Franke, “M-mode echocardiography overestimates left ventricular mass in patients with normal left ventricular shape: a comparative study using three-dimensional echocardiography,” European Journal of Echocardiography, vol. 4, no. 4, pp. 312–319, 2003. View at Scopus
  75. A. E. van den Bosch, D. Robbers-Visser, B. J. Krenning et al., “Real-time transthoracic three-dimensional echocardiographic assessment of left ventricular volume and ejection fraction in congenital heart disease,” Journal of the American Society of Echocardiography, vol. 19, no. 1, pp. 1–6, 2006. View at Publisher · View at Google Scholar · View at Scopus
  76. M. D. Cerqueira, N. J. Weissman, V. Dilsizian et al., “Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association,” Circulation, vol. 105, no. 4, pp. 539–542, 2002. View at Publisher · View at Google Scholar · View at Scopus
  77. S. Kapetanakis, M. T. Kearney, A. Siva, N. Gall, M. Cooklin, and M. J. Monaghan, “Real-time three-dimensional echocardiography: a novel technique to quantify global left ventricular mechanical dyssynchrony,” Circulation, vol. 112, no. 7, pp. 992–1000, 2005. View at Publisher · View at Google Scholar · View at Scopus
  78. O. I. I. Soliman, B. M. van Dalen, A. Nemes, et al., “Quantification of left ventricular systolic dyssynchrony by real-time three-dimensional echocardiography,” Journal of the American Society of Echocardiography, vol. 22, no. 3, pp. 232–239, 2009. View at Publisher · View at Google Scholar · View at Scopus
  79. N. A. Marsan, G. B. Bleeker, C. Ypenburg et al., “Real-time three-dimensional echocardiography permits quantification of left ventricular mechanical dyssynchrony and predicts acute response to cardiac resynchronization therapy,” Journal of Cardiovascular Electrophysiology, vol. 19, no. 4, pp. 392–399, 2008. View at Publisher · View at Google Scholar · View at Scopus
  80. N. A. Marsan, M. M. Henneman, J. Chen et al., “Real-time three-dimensional echocardiography as a novel approach to quantify left ventricular dyssynchrony: a comparison study with phase analysis of gated myocardial perfusion single photon emission computed tomography,” Journal of the American Society of Echocardiography, vol. 21, no. 7, pp. 801–807, 2008. View at Publisher · View at Google Scholar · View at Scopus
  81. C. Lau, H. M. Abdel-Qadir, I. Lashevsky, M. Hansen, E. Crystal, and C. Joyner, “Utility of three-dimensional echocardiography in assessing and predicting response to cardiac resynchronization therapy,” Canadian Journal of Cardiology, vol. 26, no. 9, pp. 475–480, 2010. View at Scopus
  82. M. Becker, R. Hoffmann, F. Schmitz et al., “Relation of optimal lead positioning as defined by three-dimensional echocardiography to long-term benefit of cardiac resynchronization,” The American Journal of Cardiology, vol. 100, no. 11, pp. 1671–1676, 2007. View at Publisher · View at Google Scholar · View at Scopus
  83. A. Deplagne, S. Lafitte, S. Reuter et al., “Absence of additional improvement in outcome of patients receiving cardiac resynchronization therapy paced at the most delayed left ventricular region,” Archives of Cardiovascular Diseases, vol. 102, no. 8-9, pp. 641–649, 2009. View at Publisher · View at Google Scholar · View at Scopus
  84. C. Miyazaki, M. M. Redfield, B. D. Powell et al., “Original articles dyssynchrony indices to predict response to cardiac resynchronization therapy a comprehensive prospective single-center study,” Circulation: Heart Failure, vol. 3, no. 5, pp. 565–573, 2010. View at Publisher · View at Google Scholar · View at Scopus
  85. C. Sonne, L. Sugeng, M. Takeuchi et al., “Real-time 3-dimensional echocardiographic assessment of left ventricular dyssynchrony: pitfalls in patients with dilated cardiomyopathy,” JACC: Cardiovascular Imaging, vol. 2, no. 7, pp. 802–812, 2009. View at Publisher · View at Google Scholar · View at Scopus
  86. Y. Mu, L. Chen, Q. Tang, and G. Ayoufu, “Real time three-dimensional echocardiographic assessment of left ventricular regional systolic function and dyssynchrony in patients with dilated cardiomyopathy,” Echocardiography, vol. 27, no. 4, pp. 415–420, 2010. View at Publisher · View at Google Scholar · View at Scopus
  87. M. Takeuchi, A. Jacobs, L. Sugeng et al., “Assessment of left ventricular dyssynchrony with real-time 3-dimensional echocardiography: comparison with Doppler tissue imaging,” Journal of the American Society of Echocardiography, vol. 20, no. 12, pp. 1321–1329, 2007. View at Publisher · View at Google Scholar · View at Scopus
  88. C. Russo, M.-P. Jaubert, Z. Jin, S. Homma, and M. R. Di Tullio, “Intra- and interobserver reproducibility of left ventricular mechanical dyssynchrony assessment by real time three-dimensional echocardiography,” Echocardiography, vol. 29, no. 5, pp. 598–607, 2012. View at Publisher · View at Google Scholar · View at Scopus
  89. J. Gorcsan III, T. Abraham, D. A. Agler et al., “Echocardiography for cardiac resynchronization therapy: recommendations for performance and reporting—a report from the American Society of Echocardiography Dyssynchrony Writing Group endorsed by the Heart Rhythm Society,” Journal of the American Society of Echocardiography, vol. 21, no. 3, pp. 191–213, 2008. View at Publisher · View at Google Scholar · View at Scopus
  90. E. A. Bacha, F. J. Zimmerman, V. Mor-Avi et al., “Ventricular resynchronization by multisite pacing improves myocardial performance in the postoperative single-ventricle patient,” Annals of Thoracic Surgery, vol. 78, no. 5, pp. 1678–1683, 2004. View at Publisher · View at Google Scholar · View at Scopus
  91. E. Abate, G. E. Hoogslag, M. L. Antoni, et al., “Value of three-dimensional speckle-tracking longitudinal strain for predicting improvement of left ventricular function after acute myocardial infarction.,” Journal of the American Society of Echocardiography, vol. 110, pp. 961–967, 2012.
  92. A. Lilli, M. T. Baratto, J. Del Meglio et al., “Three-dimensional simultaneous strain-volume analysis describes left ventricular remodelling and its progression: a pilot study,” European Journal of Echocardiography, vol. 12, no. 7, pp. 520–527, 2011. View at Publisher · View at Google Scholar · View at Scopus
  93. X. Zhang, X. Wei, Y. Liang, M. Liu, C. Li, and H. Tang, “Differential changes of left ventricular myocardial deformation in diabetic patients with controlled and uncontrolled blood glucose: a three-dimensional speckle-tracking echocardiography-based study,” Journal of the American Society of Echocardiography, vol. 26, pp. 499–506, 2013.
  94. J. A. Elefteriades and E. A. Farkas, “Thoracic aortic aneurysm clinically pertinent controversies and uncertainties,” Journal of the American College of Cardiology, vol. 55, no. 9, pp. 841–857, 2010. View at Publisher · View at Google Scholar · View at Scopus
  95. L. P. Badano, U. Cucchini, D. Muraru, O. Al Nono, C. Sarais, and S. Iliceto, “Use of three-dimensional speckle tracking to assess left ventricular myocardial mechanics: inter-vendor consistency and reproducibility of strain measurements,” European Heart Journal Cardiovascular Imaging, vol. 14, pp. 285–293, 2013.
  96. V. Mor-Avi, L. Sugeng, and R. M. Lang, “Real-time 3-dimensional echocardiography: an integral component of the routine echocardiographic examination in adult patients?” Circulation, vol. 119, no. 2, pp. 314–329, 2009. View at Publisher · View at Google Scholar · View at Scopus
  97. I. S. Salgo, W. Tsang, W. Ackerman et al., “Geometric assessment of regional left ventricular remodeling by three-dimensional echocardiographic shape analysis correlates with left ventricular function,” Journal of the American Society of Echocardiography, vol. 25, no. 1, pp. 80–88, 2012. View at Publisher · View at Google Scholar · View at Scopus
  98. H. F. J. Mannaerts, J. A. van der Heide, O. Kamp, M. G. Stoel, J. Twisk, and C. A. Visser, “Early identification of left ventricular remodelling after myocardial infarction, assessed by transthoracic 3D echocardiography,” European Heart Journal, vol. 25, no. 8, pp. 680–687, 2004. View at Publisher · View at Google Scholar · View at Scopus
  99. F. Maffessanti, E. G. Caiani, G. Tamborini et al., “Serial changes in left ventricular shape following early mitral valve repair,” The American Journal of Cardiology, vol. 106, no. 6, pp. 836–842, 2010. View at Publisher · View at Google Scholar · View at Scopus
  100. I. Hof, K. Chilukuri, A. Arbab-Zadeh et al., “Does left atrial volume and pulmonary venous anatomy predict the outcome of catheter ablation of atrial fibrillation?” Journal of Cardiovascular Electrophysiology, vol. 20, no. 9, pp. 1005–1010, 2009. View at Publisher · View at Google Scholar · View at Scopus
  101. H. P. Kühl, M. Schreckenberg, D. Rulands et al., “High-resolution transthoracic real-time three-dimensional echocardiography: quantitation of cardiac volumes and function using semi-automatic border detection and comparison with cardiac magnetic resonance imaging,” Journal of the American College of Cardiology, vol. 43, no. 11, pp. 2083–2090, 2004. View at Publisher · View at Google Scholar · View at Scopus
  102. L. E. Hudsmith, A. S. H. Cheng, D. J. Tyler et al., “Assessment of left atrial volumes at 1.5 Tesla and 3 Tesla using FLASH and SSFP cine imaging,” Journal of Cardiovascular Magnetic Resonance, vol. 9, no. 4, pp. 673–679, 2007. View at Publisher · View at Google Scholar · View at Scopus
  103. T. Shiota, P. M. McCarthy, R. D. White et al., “Initial clinical experience of real-time three-dimensional echocardiography in patients with ischemic and idiopathic dilated cardiomyopathy,” The American Journal of Cardiology, vol. 84, no. 9, pp. 1068–1073, 1999. View at Publisher · View at Google Scholar · View at Scopus
  104. M. Cameli, M. Lisi, M. Focardi et al., “Left atrial deformation analysis by speckle tracking echocardiography for prediction of cardiovascular outcomes,” The American Journal of Cardiology, vol. 110, pp. 264–269, 2012. View at Publisher · View at Google Scholar · View at Scopus
  105. N. P. Nikitin, C. Constantin, P. H. Loh et al., “New generation 3-dimensional echocardiography for left ventricular volumetric and functional measurements: comparison with cardiac magnetic resonance,” European Journal of Echocardiography, vol. 7, no. 5, pp. 365–372, 2006. View at Publisher · View at Google Scholar · View at Scopus
  106. J. L. Gutiérrez-Chico, J. L. Zamorano, L. Pérez De Isla et al., “Comparison of left ventricular volumes and ejection fractions measured by three-dimensional echocardiography versus by two-dimensional echocardiography and cardiac magnetic resonance in patients with various cardiomyopathies,” The American Journal of Cardiology, vol. 95, no. 6, pp. 809–813, 2005. View at Publisher · View at Google Scholar · View at Scopus
  107. X. Qi, B. Cogar, M. C. Hsiung et al., “Live/real time three-dimensional transthoracic echocardiographic assessment of left ventricular volumes, ejection fraction, and mass compared with magnetic resonance imaging,” Echocardiography, vol. 24, no. 2, pp. 166–173, 2007. View at Publisher · View at Google Scholar · View at Scopus
  108. L. S. Bicudo, J. M. Tsutsui, A. Shiozaki et al., “Value of real time three-dimensional echocardiography in patients with hypertrophic cardiomyopathy: comparison with two-dimensional echocardiography and magnetic resonance imaging,” Echocardiography, vol. 25, no. 7, pp. 717–726, 2008. View at Publisher · View at Google Scholar · View at Scopus