Research Article

Utility of Quantitative 99mTc-MAA SPECT/CT for 90yttrium-Labelled Microsphere Treatment Planning: Calculating Vascularized Hepatic Volume and Dosimetric Approach

Table 1

Operator 1 results.

Phantom configurationTrue sphere/
background ratio
SPECT aloneSPECT- CT
Isocontour (%)Measured volume (cm3)Error (%)Isocontour (%)Measured volume (cm3)Error Vol. (%)

18,5 MBq/sphereCylinder 12%6521−2,902%6521−2,90
Sphere 12.842%75,4537,1848%57,224,04
Sphere 27.631%22,356,4327%27,0828,95
Sphere 39.627%1812,5029%16,945,88
Sphere 419.615%14,7284,0019%10,9136,38

37 MBq/sphereCylinder 12%6180−7,981%67350,28
Sphere 1646%44,86−18,4442%49,06−10,80
Sphere 21725%22,587,5229%20,52−2,29
Sphere 323.423%17,559,6927%14,95−6,56
Sphere 443.831%10,4530,6335%9,0813,50

55,5 MBq/sphereCylinder 11%67370,311%67370,31
Sphere 19.944%45,32−17,6031%59,287,78
Sphere 229.633%18,62−11,3331%19,3−8,10
Sphere 338.633%17,328,2535%16,171,06
Sphere 473.131%14,3779,6340%10,5331,63

74 MBq/sphereCylinder 1NANA NANANANA
Sphere 11652%33,72−38,6931%54,86−0,25
Sphere 24521%23,511,9027%19,81−5,67
Sphere 35721%18,0512,8127%15,49−3,19
Sphere 411431%5,95−25,6321%8,324,00

55 MBqCylinder 246%709−8,4025%7952,71
116 MBqCylinder 246%667−13,8235%770−0,52
72 MBqCylinder 333%455−3,8123%5158,88