Separation and Quantification of Eight Antidiabetic Drugs on A High-Performance Liquid Chromatography: Its Application to Human Plasma Assay

Table 2

Summary of precision and accuracy from QC samples in plasma ( 𝑛 = 6 ).

Compound spiked concen.IntradayInterday
(ng/mL) M e a n ± S D %RSDNominal (%) M e a n ± S D % RSDNominal (%)

53.7 5 3 . 4 ± 3 . 7 6.9399.4 5 3 . 0 ± 3 . 6 6.7998.6
ROS489.2 4 8 1 . 6 ± 1 9 . 9 4.1398.4 4 8 1 . 9 ± 2 0 . 7 4.2998.5
1531.0 1 5 2 0 . 5 ± 1 2 9 . 5 8.5199.3 1 5 2 1 . 7 ± 1 1 7 . 8 7.7499.3

48.9 4 8 . 6 ± 3 . 5 7.2099.4 4 8 . 1 ± 2 . 9 6.0298.3
PIO491.2 4 9 0 . 5 ± 1 6 . 2 3.3099.8 4 9 2 . 0 ± 1 5 . 9 3.23100.1
1513.8 1 4 9 2 . 3 ± 1 3 2 . 9 8.9098.5 1 4 9 7 . 0 ± 1 2 5 . 6 8.3998.8

50.6 5 2 . 1 ± 3 . 3 6.33102.9 5 0 . 9 ± 4 . 0 7.85100.6
GLP492.3 5 0 4 . 7 ± 1 8 . 9 3.74102.5 4 9 9 . 3 ± 1 9 . 9 3.98101.4
1498.5 1 5 0 6 . 1 ± 1 0 9 . 2 7.25100.5 1 5 0 0 . 1 ± 1 3 0 . 2 8.67100.1

46.8 4 8 . 2 ± 3 . 9 8.09102.9 4 8 . 2 ± 4 . 1 8.50102.9
GLC486.7 4 9 2 . 9 ± 1 5 . 8 3.20101.2 4 9 0 . 8 ± 1 8 . 5 3.76100.8
1467.0 1 4 7 9 . 0 ± 1 1 9 . 2 8.05100.8 1 4 7 5 . 6 ± 1 2 5 . 6 8.51100.5

49.2 5 1 . 1 ± 2 . 9 5.67103.8 4 8 . 9 ± 3 . 8 7.7799.3
REP493.4 4 8 6 . 8 ± 1 5 . 9 3.2698.6 4 8 7 . 1 ± 1 7 . 3 3.5598.7
1502.7 1 4 8 9 . 4 ± 1 2 3 . 7 8.3099.1 1 4 8 6 . 9 ± 9 1 . 7 6.1698.9

52.9 5 1 . 7 ± 3 . 1 5.9997.7 5 1 . 5 ± 3 . 0 5.8297.4
NGL478.6 4 9 2 . 5 ± 1 7 . 2 3.49102.9 4 9 0 . 2 ± 1 6 . 9 3.44102.4
1439.6 1 4 5 2 . 4 ± 1 3 4 . 1 9.23100.9 1 4 4 2 . 3 ± 1 2 1 . 3 8.41100.1

45.7 4 5 . 4 ± 3 . 2 7.0499.3 4 4 . 9 ± 4 . 1 9.1398.2
GLB478.5 4 8 1 . 3 ± 2 8 . 8 5.98100.6 4 7 9 . 6 ± 2 0 . 9 4.35100.2
1452.0 1 4 5 0 . 3 ± 1 1 0 . 1 7.5899.8 1 4 5 1 . 9 ± 9 7 . 9 6.7499.9

48.6 4 9 . 1 ± 2 . 9 5.90101.0 4 8 . 5 ± 3 . 5 7.2199.7
GLM490.5 4 9 9 . 6 ± 2 5 . 6 5.12101.8 4 9 2 . 8 ± 2 4 . 3 4.93100.4
1477.5 1 4 7 9 . 0 ± 1 2 1 . 1 8.18100.1 1 4 7 0 . 8 ± 1 1 5 . 9 7.8899.5