Research Article

Stability-Indicating RP-TLC/Densitometry Determination of Raloxifene Hydrochloride in Bulk Material and in Tablets

Table 4

Summery of validation parameter.

Parameter dataRLX

Linearity range (ng per band)100–600
Correlation coefficient ( )0.9969
Limit of detection (ng)9.27
Limit of quantification (ng)27.10
Recovery 99.74–100.57

Ruggedness (% RSD)
Analyst-I 1.53
Analyst-II 1.33

Precision (% RSD)
Repeatability of application 0.96–1.49
Interday 1.33–1.37
Intraday 0.41–1.27
RobustnessRobust
SpecificitySpecific