Research Article

Dried Blood Spot Sampling with LC-MS Analysis for Routine Therapeutic Caffeine Monitoring in Neonates

Table 1

Intraday and interday accuracy (RE%) and precision (CV%) results for the LC-MS caffeine DBS method .

Nominal conc. (ng/mL)Intraday
500100050008000150002000025000

run 1Mean conc. (ng/mL)441.81034.25346.78325.015138.920061.824087.7
SD75.020.2698.2339.6668.41668.9140.1
RE%−11.83.26.83.90.80.1−3.8
CV%17.02.013.14.14.48.34.7

run 2Mean conc. (ng/mL)562.01113.45237.27820.415204.017404.821902.6
SD38.392.9182.4202.22112.4604.31519.3
RE% 12.411.34.7−2.21.4−13.0−12.4
CV%6.88.33.52.613.93.56.9

run 3Mean conc. (ng/mL)503.6971.54970.37387.815702.921355.224908.2
SD60.138.9341.1318.01466.41766.72160.4
RE%0.7−2.8−0.6−7.74.76.8−0.4
CV%12.44.06.94.39.38.38.7

Interday

Mean conc. (ng/mL)
502.4
1039.7
5184.7
7844.4
15348.619607.323632.8
SD75.177.6456.9469.61485.72165.52070.2
Average RE% 0.43.93.6−2.02.3−2.0−5.5
Overall CV%14.97.58.86.09.711.08.8