Clinical Study
Measuring Corneal Thickness with SOCT, the Scheimpflug System, and Ultrasound Pachymetry
Table 3
Coefficient of interoperator reproducibility of CCT and MCT measurements for different instruments and operators.
| Coefficient of reproducibility | Operator | System | RTVue 100 FD | Pentacam HR | Pachette 3 |
| CCT | 1 and 2 | 0.92% | 1.00% | 1.25% | 1 and 3 | 1.08% | 1.27% | 1.05% | 2 and 3 | 0.72% | 1.06% | 1.44% |
| MCT | 1 and 2 | 1.00% | 0.95% | — | 1 and 3 | 1.16% | 1.55% | — | 2 and 3 | 0.58% | 1.31% | — |
|
|