Research Article

High-Performance Liquid Chromatography Determination of Meloxicam and Piroxicam with Ultraviolet Detection

Table 3

Intra-assay accuracy, precision, and assay linearity for meloxicam in milk.

Intra-assay variability
Concentration added (ng/mL)Concentration measured (ng/mL) (mean ± SD)RSD (%)

1515 ± 0.412.7
750727 ± 273.7
12501149 ± 302.6
30003000 ± 692.3
75007450 ± 3995.3

Assay linearity
Mean ± SDRSD (%)

-intercept−0.0079 ± 0.00045.06
Slope0.0017 ± 0.0000130.76
0.999 ± 0.0030.30

SD: standard deviation; : number of samples; RSD: relative standard deviation.