Research Article

Quantitative Analysis of Resorcinol from Marketed Hair Tonic Using Liquid Chromatographic Technique

Table 3

Results of accuracy study.

% Of nominal value Estimated
Concentration of final solution in (g·mL−1)*
Estimated Recovery
%
Actual concentration of final solution (g·mL−1)*Average assay* 
(mg·100 mL−1)
Actual
Recovery
%
Accuracy
%
%RSD

10050.0045.0345.800.89
80 40.0080.0036.3636.9980.76100.95 0.54
110 56.06112.1250.6351.39112.20100.07
120 60.66121.3254.6955.55121.2999.97

Note: Average from three replicate injections from sample preparations.
The recovery study is carried out over three-concentration range.