Chromatography Research International / 2014 / Article / Tab 6 / Research Article
Development and Validation of a Novel RP-HPLC Method for Estimation of Losartan Potassium in Dissolution Samples of Immediate and Sustained Release Tablets Table 6 Results of intermediate and intraday precision.
Replicates Day 1 (area) Day 2 (area) Column I (area) Column II (area) Area 1 4008.94702 4022.12301 4067.21134 4054.66455 Area 2 3986.29590 4006.95633 3956.54367 4036.49724 Area 3 3946.11646 3977.66754 3998.43554 3986.98756 Area 4 4029.71875 3989.07986 4009.46983 3992.43927 Area 5 4052.51440 4044.43532 4032.44356 4049.87234 Area 6 4025.62451 4065.12433 3988.54437 3995.05438 Average 4008.20284 4017.56440 4008.77472 4019.25256 Standard deviation 37.590 33.228 37.999 31.095 RSD 0.94 0.83 0.95 0.77 Average 4012.88362 4014.01364 Standard deviation 32.747 32.157 RSD 0.82 0.80
RSD: relative standard deviation.