Research Article

Accuracy of Dose Delivery in Multiple Breath-Hold Segmented Volumetric Modulated Arc Therapy: A Static Phantom Study

Table 2

The mean, the SD, the minimum, and the maximum of the segmental MUs for the five cases with the five different beam-on-times between interrupts.

Beam-on-time between interrupts (s)CaseMeanSDMin.Max.

10117.56.68.936
227.94.615.731.1
328.07.010.238.6
424.610.912.655.8
519.15.911.440.3

15130.311.014.554.0
248.89.326.957.6
349.010.734.065.3
443.716.925.183.6
532.68.720.051.2

20142.918.826.591.4
265.120.523.979.0
365.321.912.880.7
460.525.325.8112.1
546.213.129.475.0

30164.319.839.8102.1
297.740.537.4124.3
3117.721.894.2149.2
498.338.063.9171.2
569.322.825.497.7

40185.722.063.7119.7
2130.256.365.4166.3
3147.125.8115.5178.6
4131.153.575.0220.7
5110.817.392.2138.0