Research Article

Development and Validation of Stability Indicating LC-PDA Method for Mycophenolate Mofetil in Presence of Mycophenolic Acid and Its Application for Degradation Kinetics and pH Profile Study

Table 2

Results of method sensitivity, linearity, precision, system suitability, and recovery of analytes.

Parameter/analytes ( )MPA (5.9)MMF (6.8)DP3 (7.9)

Method sensitivity
 Limit of detection (LOD, g/mL)0.0260.1210.02
 Limit of detection (LOQ, g/mL) 0.0780.3660.06

Concentration range ( g/mL)0.08–160.4–1200.06–12
Regression equation ( ) data

 Intercept ( )1975.0−2289.8−2253.7
 Slope ( )70583.244428.286070.0
(error in intercept) 4786 ± 9299.26789.6 ± 12628.6 1731.9 ± 3281.9
(error in slope ) 734 ± 1426.3118.3 ± 286.4286.4 ± 542.8
Correlation coefficient ( )0.9990.9990.999

Precision data, % RSD
 Repeatability, 0.290.150.34
 Intraday precision, times0.40.350.43
 Interday precision, days0.450.420.52

System suitability data (USP-NF 29 section 621, pp. 2135)
 Number of theoretical plates (±SD)3450 ± 324255 ± 263865 ± 50
 USP tailing factor (±SD)1.01 ± 0.011.01 ± 0.011.00 ± 0.042
 USP resolution5.9c2.22.25

Recovery study at the level (% recovery, % RSD),
 50%,99.35, 0.77101.6, 0.5999.56, 1.23
 100%100.5, 0.73100.3, 0.4598.87, 0.76
 150%100.8, 0.8999.98, 0.34100.2, 0.91

= 1.86, 1.895, 1.943 for MMF, DP1, and MPA, respectively ( ; df = 8), astandard deviation (SD) of intercept, bSD of slope, and cresolution with respect to IS HTZ peak.