Advances in Emergency Medicine
Volume 2015 (2015), Article ID 965368, 5 pages
Research Article

The Accuracy of Point-of-Care Creatinine Testing in the Emergency Department

1Good Samaritan Hospital Medical Center, Department of Emergency Medicine, West Islip, NY 11795, USA
2CityMD, Commack, NY 11725, USA

Received 6 August 2015; Revised 12 November 2015; Accepted 14 November 2015

Academic Editor: Frank Hildebrand

Copyright © 2015 Moises Moreno et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Objective. To determine the accuracy of Point-Of-Care testing (PoCT) creatinine values when compared to standard central laboratory testing (IDMS) and to demonstrate if and how a discrepancy could lead to improper risk stratification for contrast induced nephropathy (CIN). Methods. We conducted a descriptive retrospective chart review of patients seen in the Emergency Department of a single suburban, community, and academic medical center. We included patients who presented to the department between March 2013 and September 2014 who had blood samples analyzed by both PoCT and IDMS. Results. Mean IDMS creatinine values were 0.23 mg/dL higher when compared with i-Stat values. 95% of the time, the IDMS creatinine value was variable and ranged from −0.45 mg/dL to +0.91 mg/dL when compared to the i-Stat creatinine. When using i-Stat creatinine values to calculate GFR, 47 out of 156 patients had risk category variations compared to using the IDMS value. This affected 30.1% of the total eligible sample population (22.9% to 37.3% with 95% CI). Conclusion. We found a significant discrepancy between PoCT and IDMS creatinine values and found that this discrepancy could lead to improper risk stratification for CIN.