Chemical Warfare Agents Analyzer Based on Low Cost, Room Temperature, and Infrared Microbolometer Smart Sensors

Table 2

System performance.

Chemical detection rangeAll the 8 chemicals detected during AT.
SensitivityDuring AT, DMMP was found at LoD <900 ppb and LoI <7 ppm.
LoD and LoI depend on absorption coefficients and on correlation parameters settings. Typical values are few ppm.
SelectivityVery strong ability to recognize targets and interferents.
All the 8 chemicals were correctly identified during AT, by comparing the experimental spectra to the full DB of 400 chemicals.
Response time Few seconds to detect the gas.
False negative rate No false negative during AT.
False positive rate No false positive during AT.
Size and power needs Size  cm; weight  Kg; power < 15 W (sensor head).