Detection of Nitroaromatic and Peroxide Explosives in Air Using Infrared Spectroscopy: QCL and FTIR

Table 2

Validation parameters for the models in the subspectral range of 850–1400 cm−1.

ParameterModulated FTIR (model 1)Unmodulated FTIR (model 2)QCL scan (model 3)
TATPDNTTATPDNTTATPDNT

Wilks’ lambda0.150.040.550.540.050.06
Canonical correlation0.920.980.670.650.980.97
PCCC99.74%99.98%92.00%93.33%100.00%100.00%
PCCCC99.73%99.97%82.00%78.33%100.00%100.00%
Sensibility100.00%100.00%81.00%77.00%100.00%100.00%
Specificity93.33%99.87%84.00%81.00%100.00%100.00%
False alarm0.00%0.00%16.00%19.00%0.00%0.00%
Missed detection6.67%0.13%19.00%23.00%0.00%0.00%
Latent variables (LV)1010101055
value<0.0001<0.0001<0.0001<0.0001<0.0001<0.0001