Determination of Glycerin from a Marketed Personal Care Product Using Gas Chromatography

Table 4

Robustness of the developed analytical procedure**.

Assay
(mL/100 mL)
AssayAnalystsInstrumentsStorage conditionChromatographic condition
123III20°C30°C45°CColumn temperature (±5°C)Detector temperature
(±5°C)
Carrier gas flowrate#

9.629.619.639.669.609.629.639.629.599.669.639.61

Each analysis was carried out in three replicates.
A 5% change in the carrier gas flowrate results in no appreciable change in the retention time of the analyte ().