Tab 1
/
Research Article
A Stability Indicating UPLC Method for the Determination of Levofloxacin Hemihydrate in Pharmaceutical Dosage Form: Application to Pharmaceutical Analysis
Table 1
Result of precision, linearity, LOD, and LOQ study.
Parameter
UPLC
Linearity (µg/mL)
0.5–80
Correlation coefficient (
)
0.9994
Slope (m)
46965
Intercept (
c
)
85486
LOD (µg/mL)
0.1
LOQ (µg/mL)
0.5
Accuracy (% RSD)
0.21–0.76
Interday precision (RSD,
)
0.57
Intraday precision (RSD,
)
0.26