Research Article
A Comparative Validation Study of Fluconazole by HPLC and UPLC with Forced Degradation Study
Table 4
Comparison of HPLC and UPLC methods by validation results.
| Validation parameter | HPLC method | UPLC method |
| Accuracy (% Recovery) | 99-100 | 99-100 | Linearity | | | Regression equation | | | Corelation coefficient () | 0.984 | 0.998 | Precision | | | Intermediate precision | (% RSD) 0.61 | (% RSD) 1.18 | Method precision | (% RSD) 0.71 | (% RSD) 0.70 | Robustness | Highly robustness method | Highly robustness method | Specificity | No interference found to analyte peak | No interference found to analyte peak | Limit of detection | 0.5 µg/mL | 0.3 µg/mL | Limit of quantification | 1.5 µg/mL | 1.4 µg/mL | Total analysis time | 7 minutes | 3 minutes |
|
|