Research Article

Development and Validation of an HPLC Method for the Analysis of Chlorpropham and 3-Chloroaniline in Potato Extracts

Table 2

LOD and LOQ values based on the repeatability injections ( = 10) of 0.05 g/mL of a mixed standard solution of CIPC, IPC, and 3-CA and the statistical data for the calibration curve in the range 0.02–0.1 g/mL.

Assessed approachLOD (g/mL)LOQ (g/mL)
CIPCIPC3-CACIPCIPC3-CA

Calibration curve (0.02–0.1 g/mL)0.0060.0140.0030.0190.0480.010
Injection repeatability (0.05 g/mL)0.0110.0010.0130.0360.0020.042