Research Article

A Stability-Indicating High Performance Liquid Chromatographic Assay for the Simultaneous Determination of Pyridoxine, Ethionamide, and Moxifloxacin in Fixed Dose Combination Tablets

Table 2

Data representing accuracy and precision studies for three consecutive days.

APIsPyridoxineEthionamideMoxifloxacin
Actual concentration ( g/mL)57.51012.515507510012515080120160200240

Measured concentration5.037.4910.0512.4915.1049.9374.93100.31124.86150.2178.54122.88159.05199.33240.25
St. deviation0.010.020.010.010.010.200.250.390.260.340.270.270.100.110.51
Precision (%)0.240.260.100.070.070.400.340.390.210.230.340.220.060.060.21
Accuracy (%)100.6499.84100.5099.92100.6799.8699.91100.3199.89100.1498.18102.4099.4099.66100.10
Number of samples999999999999999