Research Article

HPLC-UV Method Development and Validation of Potato Sprout Inhibitor 1,4-Dimethylnaphthalene Using Different Systems

Table 4

LOD and LOQ values on each HPLC system based on the statistical data for the calibration curve in the range 0.02–0.1 g mL−1 for 2-MeN and 1,4-DMN.

HPLC systemInjectorLOD (g mL−1)LOQ (g mL−1)
2-MeN1,4-DMN2-MeN1,4-DMN

Hitachi DADAutosampler0.0140.0170.0460.057
SpectraSERIES UV100Manual0.0070.0080.0240.028
SpectraSERIES UV100Autosampler0.0040.0040.0130.013