Research Article

The Role of CT-Based Attenuation Correction and Collimator Blurring Correction in Striatal Spect Quantification

Table 1

Average recovery (AR) and root mean squared coefficient of variation (RMSCOV) values for unsmoothed reconstructions, (abbreviations: p0, pc, pd: no, constant, and depth dependent correction for collimator blurring resp.).

Collimator blurringARRMSCOV

CTp031.071.3
Basedpc (method 1)30.863.4
Attenuationpc (method 2)32.563.7
Correctionpd (method 1)30.364.3
pd (method 2)32.563.4

Uniformp029.974.1
Basedpc (method 1)29.367.1
Attenuationpc (method 2)31.166.6
Correctionpd (method 1)28.868.1
pd (method 2)31.166.4