Hybrid Modality Fusion of Planar Scintigrams and CT Topograms to Localize Sentinel Lymph Nodes in Breast Lymphoscintigraphy: Technical Description and Phantom Studies

Table 2

Measured exposure and calculated CT dose versus kVp and mAs. In CT scans, the CTDIvol and DLP values are used to convert in-phantom ion chamber measurements to estimated effective dose in the standard CT dosimetry phantoms (μSv).

KVpmAslocationExposure [R]CTDI_ [mGy]DLP [mGy cm]Effective dose [mSv]

8020center0.003950.45318.1370.308
Periphery0.00928
11010center0.005880.57623.0570.392
Periphery0.01137
11020center0.011771.15346.1150.784
Periphery0.02274
13020center0.017911.72068.8181.170
Periphery0.03376
13090center0.080597.742309.6815.265
Periphery0.15192