Research Article

Feasibility Evaluation of Radioimmunoguided Surgery of Breast Cancer

Table 2

Minimum T:B ± standard deviation to achieve tumour edge delineation. Background activity concentration is a surrogate for administered activity and can be related to relative error in GDP measurements. The combinations of parameters resulting in the optimum overall performance and used to generate the data in this table are as follows: for 111In, a 100 keV energy-window threshold and a 0.5 mm collimator length; for 9 9 m T c , a 90 keV energy-window threshold and 0.0 cm collimation; and for 131I, a 150 keV energy-window threshold and 0.0 cm collimation.

RadionuclideBackground activity concentration
1 kBq/mL2 kBq/mL3 kBq/mL4 kBq/mL5 kBq/mL

9 9 m T c 14.4±1.110.0±0.88.1±0.67.0±0.56.2±0.5
111In11.9±0.98.3±0.76.7±0.66.0±0.65.2±0.4
131I28.3±2.219.4±1.515.6±1.213.4±1.012.0±0.9