Research Article

Stability Indicating Method for Simultaneous RP HPLC Determination of Camylofin Dihydrochloride and Nimesulide in Pharmaceutical Preparations

Table 1

Optimization of LC method.

Mobile Phase usedColumn usedFlow rateObservationResult

DI water : acetonitrile (500 : 500)Waters symmetry C18, 250 × 4.6, 5 μ1.0 mL min−1No peaks observedMethod (rejected)
DI water : acetonitrile (500 : 500)Varian Chromosphere 5 C18, 250 × 4.6, 5 μ1.5 mL min−1Long run time and improper peak shapeMethod (rejected)
DI water : methanol (500 : 500)Varian Chromosphere 5 C18, 250 × 4.6, 5 μ1.5 mL min−1Poor resolution and bad peak shapeMethod (rejected)
Buffer solution pH5.0 : methanol (500 : 500)Varian Chromosphere 5 C18, 250 × 4.6, 5 μ1.0 mL min−1Low resolution between first two peaks (<2.0)Method (rejected)
Buffer solution pH5.0 : methanol (500 : 500)Varian Chromosphere 5 C18, 250 × 4.6, 5 μ1.0 mL min−1Good resolution and great peak shapeMethod (accepted)