Optimization of LC method.
| Mobile Phase used | Column used | Flow rate | Observation | Result |
| DI water : acetonitrile (500 : 500) | Waters symmetry C18, 250 × 4.6, 5 μ | 1.0 mL min−1 | No peaks observed | Method (rejected) | DI water : acetonitrile (500 : 500) | Varian Chromosphere 5 C18, 250 × 4.6, 5 μ | 1.5 mL min−1 | Long run time and improper peak shape | Method (rejected) | DI water : methanol (500 : 500) | Varian Chromosphere 5 C18, 250 × 4.6, 5 μ | 1.5 mL min−1 | Poor resolution and bad peak shape | Method (rejected) | Buffer solution pH5.0 : methanol (500 : 500) | Varian Chromosphere 5 C18, 250 × 4.6, 5 μ | 1.0 mL min−1 | Low resolution between first two peaks (<2.0) | Method (rejected) | Buffer solution pH5.0 : methanol (500 : 500) | Varian Chromosphere 5 C18, 250 × 4.6, 5 μ | 1.0 mL min−1 | Good resolution and great peak shape | Method (accepted) |
|
|