Research Article

Use of Alizarin Red S as an Ion-Pair Reagent for the Spectrophotometric Assay of Fexofenadine in Pharmaceuticals and in Spiked Human Urine

Table 4

Evaluation of intraday and interday precision and accuracy.

MethodFFH taken
(μg mL−1)
Intraday ( )Interday ( )
FFH founda (μg mL−1)%RSDb%REcFFH
founda   (μg mL−1)
%RSDb%REc

A4.004.081.262.114.101.062.65
6.006.141.322.346.170.622.85
8.008.151.911.778.180.592.34

B1.001.011.521.201.022.002.04
2.001.981.880.861.971.741.27
3.002.950.871.622.930.912.05

aMean value of five determinations; brelative standard deviation (%); crelative error (%).