Abstract

In this paper we present systems for dual modality imaging, combining fluorescence-enhanced diffuse optical tomography and X-ray computed tomography. Fluorescence diffuse optical tomography is carried out in a cylindrical geometry, which ensures optimal sampling and a straight forward integration with the X-ray modality. Specific acquisition protocols and reconstruction software have been developed to this end. The X-ray computed tomography serves two purposes. First, it provides the anatomical information in the registered dual modality images. Second, it provides the actual shape and boundaries of the animal as a priori input to the fluorescence reconstruction algorithm. To evaluate the performance of the optical imaging system, experiments have been conducted on phantoms, mice with inserted fluorescing capillaries, and finally on mice bearing tumors, ex-vivo and in-vivo. Experiments on mice with capillaries inserted in different region of interest, allow estimating the detection limits of fluorophore concentrations. The fluorescence reconstructions are shown to be geometrically consistent with the X-ray images. Finally we demonstrate the capability of the bimodal system to localize real tumours in mice in-vivo. These results show that dual modality fluorescence-enhanced diffuse optical tomography and X-ray computed tomography imaging in cylindrical geometry has a high potential for small animal tumour evolution studies.

1. Introduction

Three-dimensional near-infrared fluorescence-enhanced Diffuse Optical Tomography (fDOT) has been proven to be an efficient noninvasive tool for preclinical cancer research [1, 2]. With it, the biodistribution of fluorescent probes targeting molecular markers of tumor development can be quantitatively estimated. Moreover, fDOT can be used to assess the influence of anti-cancer treatments on the molecular level, which makes it a relevant technique not only in fundamental research but also in drug development [35]. fDOT, as a molecular imaging modality analog to PET, does not readily provide anatomical information. However, structure can be obtained by combining fDOT with a second modality providing morphological information, such as MRI [6, 7], ultrasound [8, 9], or XCT [1013]. In this way, obtaining in vivo anatomical and molecular information simultaneously becomes possible.

In the present work, XCT has been chosen as the second imaging modality because it provides high resolution, is relatively cheap, and its integration with optical tomography in a single instrument is straightforward. The dual-modality-approach facilitates the interpretation of the measured data as it allows the coregistration of molecular data with anatomical information. This way, the animal shape and boundaries from the XCT can be used as a priori input to the fDOT reconstruction algorithm.

The data presented in this paper were obtained on two instruments. The first one is a laboratory setup, which physically integrates the XCT and the fDOT modality. The second one is an industrial prototype, based on the technology tested on the laboratory setup. In contrast to the laboratory setup, it features a horizontal rotation axis and the XCT and fDOT acquisition chains are physically separated. In both cases, fDOT scanning is performed in a cylindrical geometry. This approach allows sampling of a larger part of the animal surface as compared to a planar tomography setup. This results in superior imaging quality, as was shown previously [14, 15]. To provide high reconstruction accuracy, even in very heterogeneous tissue, the fDOT reconstruction algorithm takes optical heterogeneities into account [16].

In previous work, the laboratory system was characterized in terms of linearity, repeatability, detection limit, and resolution [14]. These studies were based on measurements taken on phantoms. In the present paper, measurements are presented that validate the systems and method on sacrificed mice and on mice in vivo. The fDOT reconstructions are evaluated in terms of the minimum fluorophore quantity required to overcome autofluorescence or nonspecific fluorescence, that is, injected fluorescence nontargeted to tumors, in the localization error of the reconstructed inclusions, and the presence of artefacts.

In the first part of this paper, we describe the optical imaging systems with details given about the instrumentation describing the two purposes of the XCT modality. First, it provides the actual shape and boundaries of the measured animal as a priori input to the fDOT reconstruction algorithm. Second, it provides anatomical information in the dual-modality images registered. The second part is devoted to the description of the method and algorithm used for the fDOT reconstruction. The third part presents results obtained on mice. In a first study, dead mice with a capillary containing fluorophore inserted in the trachea or in the abdomen were imaged on the laboratory setup. The results validate the registration between the XCT and fDOT images. In particular, the error in localizing the fluorescent inclusion was found to be smaller than 1 mm. This study also allowed determining the minimum fluorophore quantity necessary to overcome the signal due to natural or nonspecific fluorescence. Finally, experiments on mice bearing different types of tumors are presented. Two independent studies are discussed. The first one was conducted on the laboratory setup on dead mice bearing lung tumors. The second one was carried out on the industrial setup in vivo on live mice bearing subcutaneous tumors in the pelvic area.

2. Bimodal Instrumentation

2.1. Laboratory Setup

The cylindrical geometry is the most suitable geometry to deal with both modalities since XCT instruments are common in this geometry. Moreover, it allows a uniform scan of the entire object’s surface, which favors the excellent performances of the system in terms of sensibility and isotropic resolution along all axes [15]. Previous studies have led to the development of a planar fDOT continuous wave (CW) tomograph [1619]. To cope with this new geometry, we developed a cylindrical laboratory bench demonstrator as described on Figure 1.

The position of the illumination relative to the detection system, defining the optical axis, is fixed, mounted on an optical table. The optical axis is horizontal (in the 𝑥, 𝑦 plane), and the observed animal is placed vertically on a rotation stage with a vertical rotation axis (around 𝑧), see Figure 2(a). Note that this geometry is dual from the one used on classical XCT systems where the source-detector pair is rotating. In order to scan vertically (along 𝑧) the examined object, the optical chain is mounted on synchronized translation stages and controlled by a computer.

The XCT axis defined by the relative position of the X-ray detector and the X-ray generator has been placed perpendicular to the optical axis in the 𝑥, 𝑦 plane. The optical chain bulk has been optimized according to X-ray constraints: geometric constraints, protection against radiation, against reflections. A geometrical calibration of the XCT and fDOT system by the use of a dedicated calibration object (cylindrical phantom with inserted reference balls that are visible both in XCT and fDOT modalities) is performed prior to acquisition and analysis of data.

The fDOT instrument chain consists of a continuous wave laser source (diode laser HEML, Frankfort Laser Compagny 685 nm, 60 mW), a high-sensitivity CCD camera (Andor DV438) and a combination of filters (coloured RG9, band-pass filter Semrock 775/46, and interference filter ALP 770) for fluorescence detection. The excitation and emission wavelengths of the system and the fluorophores are chosen in the visible range, close to NIR, that is, the so-called therapeutic window to optimize transmission through the whole animal. A custom-made, easy to use holder presented in Figures 2(b)2(d) has been designed for use with both modalities, that ensures the animal is immobile and in an almost “natural” position even though it is vertical.

2.2. Industrial Prototype

Based on the fDOT system described above, an fDOT industrial prototype scanner (see Figure 3(a)) was developed at SCANCO Medical after industrial transfer of the technology [20]. This scanner features a 360° projection, transillumination geometry, just as the lab-setup described in Section 2.1. However, the rotation axis is horizontal, which allows a more natural position of the animal and easier anesthesia setup. The change in scanning-geometry is achieved by mounting the illumination system and the detection system on a platform that can be rotated around the animal. The animal is placed on a holder that can be translated along the rotation axis of the platform, which allows for uniform sampling of the animal surface in two directions. An additional advantage of the horizontal scanning geometry is its increased compatibility with other imaging technologies. Because the described geometry is the preferred one in many existing small animal imaging systems, moving the animal between systems and thus registering of images from different modalities is simplified. The scanner features two diode lasers emitting at 650 nm and 690 nm. A fiber switch allows selection of the excitation wavelength. Fluorescence emission bands are selected by means of a filter changer comprising interference filters optimized for Alexa Fluor 680 and Alexa Fluor 750. Transmission measurements at the excitation wavelength are performed with an ND filter (OD 2.25) in the detection path. The whole illumination and detection system is enclosed by a lightproof housing that allows measurements to be taken under typical workplace lighting conditions.

While the first lab setup integrates the microCT and an fDOT system on a single measurement platform, the present scanner does not yet. MicroCT measurements are performed on a separate system (vivaCT, SCANCO Medical AG, Switzerland). The link between the two modalities is provided by a dedicated animal holder that can be moved from one system to the other (Figure 3(b)). This holder is fixed to an adapter on either of the two scanners in a precise and repeatable manner. As a result, the position of the animal inside the fDOT system can be determined by a microCT scan, provided an appropriate geometrical calibration has been performed once. The geometrical calibration is achieved prior to the measurements, in a similar manner as it is done for the laboratory system, that is to say by means of a calibration object, which is visible in both, the XCT and fDOT modalities. The transformation matrix that relates the coordinate systems to each other is calculated automatically from the optical and the microCT images by software developed by the authors.

3. Method

3.1. Effective Coupling of XCT/fDOT

The effective coupling of XCT/fDOT modalities occurs at different levels of the process. First, the XCT image is used as a priori information for the fDOT algorithm (external shape of the mouse); second, it gives anatomical information to help with fDOT image interpretation.

The X-ray acquisition protocol defines the number of acquired projections and the voltage, current, and filtering of the X-ray generator. The reconstruction of the volume of interest is then conducted using, for example, the classical Feldkamp algorithm [21].

The optical acquisition protocol defines the reference, diffusion, and fluorescence data acquired according to a set of angular positions and Z-positions. A dark noise and optical vignetting effects correction is performed as preprocessing of the data before they are used in the fDOT algorithm for reconstruction. (i)Reference data are acquired on a reference object (with known optical coefficients) at the excitation wavelength, for each source-camera position to calibrate the system for quantification purpose [22]. This calibration is done once, prior to each series of experiments.(ii)Diffusion and fluorescence data are acquired on the object to be reconstructed at both excitation and emission wavelengths, for each source-camera position.

The method used in the system is described in six Steps in Figure 4 using the example of a phantom mimicking the mouse shape and with similar optical properties (𝜇𝑎=0.24 cm−1, 𝜇𝑠=10 cm−1). A solid fluorescent inclusion (1 mm3) visible in XCT modality has been inserted in it. After positioning the object or the animal on the bench (Step  1) and acquisition and reconstruction of XCT data according to the X-ray acquisition protocol (Step  2), the external shape of the object is extracted from the XCT volume (Step  3). The XCT reconstructed volume is then discretized to form a mesh (finite volumes) and used by the fDOT algorithm to solve the light propagation equation taking into account actual object boundaries. Figure 5 shows the surface extracted from the XCT volume (color) superimposed on the XCT image (grey level) for a mouse imaged on the lab setup.

The laser source is approximated thanks to the diffusion approximation by an isotropic point source—called excitation point source—located inside the object at 1/𝜇𝑠 from the boundary. For each of the “detector-source” positions defined in the optical acquisition protocol, the intersection of the excitation light beam with the object’s surface is determined. The intersection point is calculated with a ray-driven method ((a) in Figure 6). This method calculates the intersection point of the light ray from the excitation source with the object’s surface. In the case of an object with a complex shape, several intersection points may be found for a single light ray. Hence, the algorithm has to be capable of selecting the relevant intersection point. The nearest point to the laser source is selected (entering point) ((b) in Figure 6). The same approach is used to calculate the projections of the detectors onto the object’s surface. We finally obtain the geometry used in fDOT reconstruction described in (c) in Figure 6, with the mesh displayed as black dots. The excitation point sources obtained with the ray-driven method described above are represented as red dots and the detector positions are shown in blue.

The laser scanning and image acquisition is performed according to the optical acquisition protocol as illustrated on Figure 4 Step  4. The 3D reconstruction of the fluorescence yield is then achieved (Step  5) by using the method developed in our laboratory [23] and described in the next section. Finally, visualization and quantification of the fluorescence is performed on a merged volume of both modality images (Step  6).

3.2. Reconstruction Method

This method allows the reconstruction of fluorescence even in highly attenuating media such as the liver or lungs. A map of optical heterogeneities is reconstructed first from diffusion data. It compounds the intrinsic tissue heterogeneity and the border effects due to the complex shape of the turbid medium. Then fluorescence is reconstructed from fluorescence data. In this study, the propagation of light modeled as a scalar field 𝜙(𝐫) in highly diffusive media (with 𝐫 position in the studied medium belonging to a domain Ω with boundary 𝑑Ω) is assumed to follow the diffusion approximation equation:𝐷(𝐫)Φ(𝐫)+𝜇𝑎(𝐫)Φ(𝐫)=𝑞0(𝐫)(inΩ),(1) where 𝑞0(𝐫) is the source term and 𝐷(𝐫)=[3𝜇𝑎(𝐫)(𝜇𝑎(𝐫)+𝜇𝑠(𝐫))]1 is the diffusion coefficient, 𝜇𝑎(𝐫) is the absorption coefficient and 𝜇𝑠(𝐫) is the reduced diffusion coefficient. A simplification of (1) leads to the simplified Helmotz equation as explained in [24]2𝑞𝜓+𝜂𝜓=0𝛿𝐫𝐫𝑠𝐷𝐫𝑠with𝜂=2𝐷(𝐫)+𝜇𝐷(𝐫)𝑎(𝐫)𝐷(𝐫),𝜓=𝐷𝜙,(2)𝛿 is the Dirac function to modelize a source positioned at 1/𝜇𝑠(𝐫) inside the medium. 𝜂(𝐫) is a scalar map we call the “heterogeneities map,” which compounds the attenuation and reduced scattering coefficients, and can also take into account effects of diffusion variations. The Helmotz equation (2) has a solution given by the Green function 𝐺 and leads to  (3)𝑞𝜓=0𝐷𝐫𝑠𝐺𝐫𝑠,𝐫.(3)

So, given an 𝜂 map, we can compute the propagation function 𝐺(𝐫1,𝐫2) of light in the medium between any points 𝐫1 and 𝐫2:2𝐺𝐫𝜂𝟏,𝐫𝟐𝐫=𝛿𝟏𝐫𝟐.(4)

By modelling the laser source 𝑠 as a Dirac delta function position at 𝐫𝐬, (1) and (4) give the excitation density wave Φ𝑥(𝐫) at any point 𝐫 of the medium:Φ𝑥𝐫𝐬𝐫,𝐫=𝜆𝐺𝐬,𝐫.(5)

𝜆 is a conversion factor related to the laser source intensity and the detection efficiency of the camera and is determined by calibration on the reference acquisition data. It is such also dependant on the diffusion of the reference phantom 𝐷0

A discrete version of (4) is solved on the mesh extracted from the X-CT volume. Using its external shape, we complete the equation system with the “Robin conditions” [25], see (6), where Φbnd(𝐫) stands for the field at the boundary and 𝑛(𝐫) for the outward normal vector:Φbnd(𝐫)+𝛼𝐷Φbnd𝑛(𝐫)=0.(6)

The global reconstruction process is then performed in two main Steps. We first compute the heterogeneities 𝜂(𝐫) map by comparison of the Green’s functions (light propagating between the source 𝑠 and the detector 𝑑) 𝐺(𝐫𝐬,𝐫𝐝) to the excitation measurements. This first Step (“Heterogeneities” method) is itself an iterative process initialized with a homogeneous map 𝜂0(𝐫). The second Step is the reconstruction of the fluorescence yield from fluorescent measurements using the heterogeneity map obtained with the diffusion data, as explained in [22].

As said before, Φ𝑥(𝐫) denoted the excited light absorbed by the fluorophore at point 𝐫, so the emitted light equals to 𝑆𝑓(𝐫)=Φ𝑥(𝐫)𝑋(𝐫) with 𝑋(𝐫) proportional to the fluorophore concentration in 𝐫. The detected signal in detector 𝑑 is given in 𝐫𝑑. byΦ𝑚𝐫𝑑=𝑞0𝐺𝐫𝑠,𝐫𝐷𝐫𝑠𝑋(𝐫)𝐺𝐷(𝐫)𝐫,𝐫𝑑𝐷𝐫𝑑𝑑𝐫.(7)

Note that we assume that the medium optical properties remain constant at the excitation and emission wavelength. Thus we reconstruct the quantity 𝑋(𝐫)/𝐷(𝐫), that is, the fluorescence concentration divided by the local diffusion coefficient 𝐷(𝐫). As demonstrated in [24], unless 𝐷(𝐫) is known, there is no unique solution to the fluorescence yield reconstruction.

Convergence of the fluorescence reconstruction algorithm is achieved within 15 iterations by using a classical iterative ART algorithm with a relaxation parameter of 0.1. We have shown on different mice experiments that further iterations with this relaxation factor induce nonsignificant variations of the measuresreprojections norm.

4. Measurements and Results

This section presents different kinds of experiments conducted on mice. First experiments were done to determine the minimum fluorophore quantity required to overcome of natural or nonspecific fluorescence. These experiments were conducted on the laboratory setup on sacrificed mice in which capillaries were inserted (mice no. 1 to no. 4). Afterwards, experiments were done on the laboratory setup to validate the feasibility of imaging tumor-bearing mice (mice from no. 5 to no. 7). As no anaesthetic system is installed on the laboratory setup, the mice were sacrificed just before measuring with the laboratory setup. Finally, in vivo experiments were conducted on the industrial setup (mice from no. 8 to no. 10).

Mice used on the laboratory setup were on a chlorophyll-free diet before the measurements were performed, whereas mice used on the industrial setup were not. All the animals used in this study were treated and handled according to the animal experimentation ethical committee Cometh38 for the laboratory setup and in strict adherence to the Swiss Law for Animal Protection for the industrial setup.

4.1. Measurements on Dead Mice Containing a Fluorescent Inclusion

Four experiments were conducted on mice either on the thorax or on the abdominal area. A capillary tube—filled with a solution of Alexa 750—was inserted either in the lungs via the trachea or in the abdomen of sacrificed healthy nude mice. In order to determine the concentration of fluorophore required to overcome natural or nonspecific fluorescence for each region of interest, we compare the fDOT reconstruction to the reconstruction of the same experiment with the mice imaged without the capillaries inside.

The X-ray generator is a monoblock X-ray source, manufactured by Gilardoni (Italy) with a CW power of 500 W. The anodic voltage is tunable from 40 kV to 140 kV allowing X-ray emission with an energy spectrum spreading from about 5 keV to 140 keV. The flux of the radiation emitted by the tungsten anode is controlled by the electrode current intensity adjustable from 1 mA to 7 mA. In our case, X-ray data are acquired according to a fast protocol with 400 projections at 40 kV, 6 mA, and a 50 μm Tin filtering. The Focal spot is approximately 1 mm. The X-ray detector is from Hamamatsu, Japan: C7942-CA with a 1 mm aluminium window. A preliminary Step of 4× binning (560×592 pixels² images) reduces the size of the acquired data and is sufficient for providing external shape information for fDOT. The reconstructed volume consists of 300×300×420 voxels @ 150×150×150μm³ (𝑥, 𝑦, 𝑧).

fDOT data are acquired according to 40×12 sources positions (12 vertical translations and 40 rotations). A rectangular detection area is selected onto the camera images according to the projection of the field of view and binned to form a grid of 40×12 detectors. Some views are removed to avoid the artefacts induced by the mouse holder (i.e., the views for which the optical axis intersects the mouse holder). The total reconstructed volume is composed of 17×13×26 voxels of 2×2×2 mm3; the integration time was 3 s per frame for fluorescence images and 400 ms per frame for diffusion images.

On the laboratory setup, not optimized for acquisition times, the total X-ray acquisition takes about 5 minutes, the DOT acquisition 20 minutes, and the fDOT acquisition 40 minutes, so about 1 hour total acquisition time per mouse. The reconstruction time for XCT is about 10 minutes and 10 minutes for fDOT on a 2.6 Ghz PC (pentium 4).

The coregistered fDOT and XCT images are presented on Figures 7(a) and 7(b) (two view angles) for the lungs area experiment on mouse no. 1. The capillary is filled with 2.5 μL of Alexa 750 at a concentration of 2 μmol/L; that is, with a 5 pmol quantity. The results in Figure 7 highlight clearly the position of the capillary tube visible also in XCT modality on the left image (arrow). We localized the fluorophore in the morphology of the animal with a tolerance of 1 mm.

This experiment was repeated with a quantity of 2 pmol in the capillary Figure 8. This was the minimal quantity necessary to overcome the natural fluorescence in this area. We localized the fluorophore in the morphology of the animal with a tolerance of 1 mm.

The result for the abdominal area experiment on mouse no. 3 is presented on Figure 9. The capillary was filled with a quantity of 5 pmol as for experiment on mouse no. 1. The autofluorescence in the abdomen is much higher than in the lungs, despite the chlorophyll-free diet. Therefore, it was impossible to distinguish specific fluorescence from autofluorescence. For this reason, we performed the same kind of study with a fourth mouse filling the capillary with 2.5 μL of Alexa 750 at a concentration of 4 μmol/L; that is, with a quantity of 10 pmol. The result for the abdominal area experiment on mouse no. 4 is presented in Figure 10. It is now possible to distinguish specific fluorescence from natural fluorescence; the result is comparable to experiment on mouse no. 1 on lungs.

In this first series of experiments, we determined the minimal detectable fluorophore quantities for different areas of the mouse. Specifically, we found the minimum quantity for the lungs to be 2 pmol and for the abdomen 10 pmol. This difference is both due to the different absorption properties of the two types of tissue (absorbent region for the lungs and very heterogeneous region presenting intrinsic fluorescence for the abdomen), but also due to the different levels of autofluorescence present. Although we found a sensitivity limit of 2 pmol for the lungs, we now routinely use 5 pmol to ensure detection of capillary tubes despite variable absorption across different mice and depending on the position of the tube (more difficult when positioned near the liver).

For each one of these experiments, the use of the bimodal system leads to accurate results in terms of localisation with a tolerance of 1 mm depending of the size of the mesh (here 2 mm), independent of the zone of interest. It opens the field for further experiments of diseased mice and registration of fluorescence sources (targeted tumors) in the anatomy of the mouse.

4.2. Measurements on Tumor-Bearing Dead Mice

For tumor bearing mice, we define the optical acquisition protocol with 28×6 sources positions (6 vertical translations and 28 rotations, that is, we suppress views of the mouse holder at the acquisition Step) and the reconstructed volume with 29×32×48 voxels of 1×1×1 mm3 to have a better resolution, the integration time was adapted according to the position.

We used CreLox inducible p53−/−rb−/− transgenic mice. Adenovirus encoding Cre Recombinase is administered intratracheally, which induces lung tumor formation in192±36days. We image the mice 244 days after infection: for that 10 nmol of Angiolone (Fluoptics) marked with AlexaFluor 700 (Invitrogen) were injected intravenously 4 h before imaging. It has been shown that this probe targets the αvβ3 integrin, which is overexpressed in tumors.

The fluorophore quantity is far higher than that used for capillaries inserted into the mouse. This is calculated based on the fact that in the animal, 1% of the injected dose is targeted per gram of tumor, that is, 100 pmol/g tumor. Moreover, highly diseased lungs weigh around 1 g so the maximum quantity for any tumor is 1 g.

We show in Figure 11 the results from two mice; the distribution obtained (small distinct tumoral sites in mouse no. 5 and a unique large site in mouse no. 6) was confirmed by the subsequent autopsy. As these tumours are not visible in XCT, only qualitative localization accuracy can be concluded. The fluorescence level has been quantified by FRI on the lungs after excision and confirms the reconstruction results. To verify that the fluorescence observed was specific, a control healthy mouse, mouse no. 7, was measured and the results are presented in Figure 12 on the same scale as Figure 11. The detected fluorescence is far lower than for tumour-bearing mice no. 5 and no. 6.

4.3. In Vivo Measurements on Tumor-Bearing Mice

Three in vivo measurements were conducted on nude mice bearing tumor xenografts using the industrial prototype. C51 tumor cells were implanted subcutaneously close to the base of the spine or at the thigh, ten days prior to measuring. Two of the mice were injected with two doses of 1 nmol Prosense 680 (VisEn Medical) via the tail vein. Prosense 680 is a probe that is activated by proteases and in particular by cathepsin, which is overexpressed in the tumor cells used here. The first dose of 1 nmol was injected approximately 48 hours and the second 24 hours before imaging. The third mouse served as a control and did not receive any injection of contrast agent. Before performing the fDOT measurements, food was withdrawn for four hours.

fDOT measurements were performed at an excitation wavelength of 650 nm. The power applied was about 1 mW at all source positions and the integration time was 200 ms per frame.

We performed microCT measurements on a vivaCT (SCANCO Medical) immediately after the fDOT measurements. The images obtained were segmented and used to generate the mesh for fDOT reconstruction (see (c) in Figure 6), as described above in Section 3.1. After reconstruction, the images obtained with the two modalities were segmented and registered.

Figure 13 to Figure 15 show views of the registered microCT and fDOT data obtained from the three in vivo experiments. Postprocessing of the microCT images was performed to visualize bone and skin. In all three cases the tumor masses form ulcers that are readily discernable on the microCT images. The reconstructed fluorophore distributions are visualized as iso-surfaces, showing the concentration at 50% of the reconstructed maxima. In all three cases, some fluorescence is reconstructed that is not readily attributable to the tumor tissue. Nonspecific fluorescence was reconstructed in the abdomen and might stem from fluorescent substances found in food, as those animals were not placed in a chlorophyrle-free diet, also due to autofluorescence or nonspecific fluorescence (i.e., fluorescence not targeted to tumors).

Mouse no. 8 bears one tumor (arrow) close to the spine (Figure 13), which is clearly visible in both microCT and fDOT modalities. Here, the maximum reconstructed fluorophore concentration is about six times above the maximum detected for the control case no. 10 (Figure 15). This mouse had a second, smaller tumor close to its left knee. An ulcer is visible in the microCT image but is not detected by the fDOT modality. The location of the second tumor made the measurements more challenging, as most of the source and detector positions are at a relatively far distance from the tumor site. Tissue at this specific location is not optimally sampled, which might lead to a decreased detection sensitivity at this particular location.

In mouse no. 9 (Figure 14) grew one, large tumor at the left thigh (arrow). The tumor is clearly reconstructed by the fDOT algorithm as a torus-like shape. Some nonspecific fluorescence was reconstructed in the abdomen.

Figure 15 shows the distribution of fluorophore for the control case, which is a noninjected diseased mouse (mouse no. 10). No fluorescence activity was detected in the tumor tissue (the ulcer is visible close to the lower spine, arrow), but some fluorescence was located in the lower abdomen. However, the reconstructed maximum concentration is at least a factor of 2 below the threshold applied in the two cases discussed above.

5. Conclusion

Two systems, a laboratory setup and an industrial prototype, dedicated to the multimodal optical and X-ray measurements, have been developed. Experiments have been successfully conducted on sacrificed mice with a capillary tube containing Alexa750, inserted either in the trachea or the abdomen. We have thus determined the limit of sensitivity of the laboratory system to detect fluorescence signal over natural or nonspecific fluorescence. For lung area, this limit is 5 pmol; for the abdomen it is 10 pmol. The fluorophore localization is in accordance with the X-ray reconstruction. Some experiments on mice bearing tumors showed that the reconstructed fluorescence in the case of mice without tumors is far lower than the quantity reconstructed for mice bearing tumors. In these reconstructions, artefacts and nonspecific fluorescence are reconstructed. Artefacts can be due to areas that are slightly thicker at certain angles of view or to the mouse holder itself. The optical acquisition protocol has been adapted to suppress these views. Measurements with the industrial prototype confirm and extend that the proposed method can be used for in vivo imaging of tumors in mice. Increased protease activity was visualized in the tissue of subcutaneous tumor xenografts. Autofluorescence seems not to be an issue in this kind of experiments even if the mice are not on a chlorophyll-free diet prior to measuring. The systems were successfully applied to different regions of the mouse anatomy that is, trachea, abdomen, pelvis, and for deeply sited tumors/fluorophore inclusions as well as for subcutaneous tumors.

In this paper, the optical reconstructions were performed with no prior knowledge on the position of the fluorophore. The use of XCT as prior knowledge offers others advantages [26, 27]. It allows, for example, to speed up the computation by introducing some spatial constraints and prior knowledge [28]. Future work will address the use of a priori as a regularization factor in the optical reconstruction scheme and particularly in a better estimation of the “heterogeneities map” increasing the accuracy of the reconstructions of the parameters extracted from optical measurements more accurate. Furthermore, knowing the position of the organs from the CT reconstruction, and constraining the optical reconstruction algorithm, the optical reconstruction of the fluorescence would be also more accurate [29].

The precise coregistering of fDOT with X-ray μCT data/images allows evaluation of molecular information from marked labelled tissue within its anatomical context.

Acknowledgments

This paper was funded by a grant from SCANCO MEDICAL AG, Fabrikweg 2, 8306 Bruttisellen Switzerland. The animal model used at LETI is provided by Ralph Meuwissen, Centre Recherche INSERM U823 Equipe 2, Institut Albert Bonniot 38706 La Tronche Cedex, Grenoble, France. The animal model used at SCANCO is provided by group Markus Rudin, Inst Biomed. Engineering, UZH/ETHZ, Switzerland.