Reproducibility of Retinal Nerve Fiber Layer Measurements with Manual and Automated Centration in Healthy Subjects Using Spectralis Spectral-Domain Optical Coherence Tomography

Table 3

Overview published studies comparing RNFLT in SD-OCT.

AuthorSD-OCT usedStudy population (eyes)Results

Leung et al. [6]Cirrus and stratus83 glaucoma
97 controls
Cirrus: (95% CI)
 (i) repeatability 5.12 (3.87–6.37)
  (ii) CV 1.89% (1.43–2.36)
  (iii) ICC 0.975 (0.945–0.991)
Stratus:
  (i) repeatability 11.10 (8.34–13.87)
  (ii) CV 3.58% (2.69–4.48)
  (iii) ICC 0.866 (0.741–0.933)

Vizzeri et al. [8]Cirrus78 glaucoma
32 controls
Cirrus (95% CI)
  (i) Sw 1.3
  (ii) CV 1.5%
  (iii) ICC 0.96 (0.94–0.98)

Kim et al. [9]Cirrus and stratus27 controlsCirrus (95% CI)
 ICC 0.984
Stratus
 ICC 0.894

Gonzalez-Garcia et al. [10]RTVue-10076 glaucoma
60 controls
RTVue (95% CI)
  (i) Sw 0.01
  (ii) CV 0.27%
  (iii) ICC 0.99 (0.98–0.99)

Seibold et al. [11]Cirrus, Spectralis, RTVue-100, stratus80 controlsCirrus (95% CI)
  (i) Reproducibility 8.89 (5.75–12.02)
  (ii) CV 3.03% (1.47–4.03)
  (iii) ICC 0.92 (0.87–0.97)
Spectralis
  (i) Reproducibility 11.72 (9.31–14.13)
  (ii) CV 3.91% (2.81–4.77)
  (iii) ICC 0.90 (0.85–0.95)
RTVue-100
  (i) Reproducibility 6.59 (5.59–7.59)
  (ii) CV 2.09% (1.71–2.41)
  (iii)ICC 0.97 (0.96 to 0.98)
Stratus
  (i) Reproducibility 8.83 (7.21–10.45)
  (ii) CV 2.86% (2.17–3.42)
  (iii) ICC 0.94 (0.91 to 0.97)

Arthur et al. [12]Spectralis30 controlsSpectralis (95% CI)
 ICC 0.96 (0.94–0.99)
Stratus
 ICC 0.86 (0.77–0.94)

Langenegger et al. [14]Spectralis56 controls
47 glaucoma
Spectralis (95% CI)
  (i) CV 1.0% with retest function
  (ii) CV 1.6% without retest function
  (iii) ICC 0.99 (0.98)

Serbecic et al. [15]Spectralis62 ControlsSpectralis (95% CI)
 CV 0.545–3.97%

Current studySpectralis39 controlsSpectralis (95% CI)
  (i) Reproducibility 1.30
  (ii) CV 0.63%
  (iii) ICC 0.983 (0.97–0.99)