Research Article
Liquid Chromatographic Method for the Quantification of Antipsychotic Agent Iloperidone in Pharmaceutical Formulation
Table 1
Regression analysis of the calibration curve and summary of validation parameters.
| Parameters | ILO |
| Linearity range (μg/mL) | 0.1–20 | Slope | 60966 | Standard deviation of slope (SD) | 320.69 | Intercept | 2697.64 | Standard deviation of intercept (SD) | 405.28 | Correlation coefficient | 0.9996 | Detection limit (μg/mL) | 0.022 | Quantitation limit (μg/mL) | 0.067 | Accuracy (%) | 99.26–100.33 | Precision (RSDa, %) | | Intraday precision () | 0.33–1.06 | Interday precision () | 0.41–1.14 | Repeatability study (RSDa, %) | 0.49 |
|
|
a RSD is relative standard deviation and, “n” is the number of determinations.
|