The Performance of ICDAS-II Using Low-Powered Magnification with Light-Emitting Diode Headlight and Alternating Current Impedance Spectroscopy Device for Detection of Occlusal Caries on Primary Molars

Table 3

Area under the ROC curve (standard error), sensitivity, and specificity at D1 diagnostic threshold.

D1 diagnostic thresholdAC Impedance
Spectroscopy
Low-powered
magnification + LED
Examiner 1Examiner 2Examiner 1Examiner 2

AUC (SE)0.60 (0.14)0.70 (0.13)0.93 (0.04)*0.87 (0.09)*
Sensitivity0.620.570.900.87
Specificity0.680.700.700.75

AUC: area under the curve; SE: standard error.
Statistically significant difference ( ).

We are committed to sharing findings related to COVID-19 as quickly as possible. We will be providing unlimited waivers of publication charges for accepted research articles as well as case reports and case series related to COVID-19. Review articles are excluded from this waiver policy. Sign up here as a reviewer to help fast-track new submissions.