Summary of fatigue failure tests (all values before failure): max. load, vertical and lateral stiffness and number of cycles. The Contour Lock plate shows better performance in all investigated parameters.

SpecimenMaximal load [N]Vertical Stiffness [N/mm]Lateral Stiffness [N/mm]Number of cyclesFailure type

TomoFix 3128022002000>600002, 3
TomoFix 4128017501500>600002, 3
TomoFix 5176020002300>1200001, 2
Contour Lock 3240021004400>2000002
Contour Lock 4176023002400>1200002
Contour Lock 5240027002600>2000001, 2, 3