Table 1

Steep and progressive suction procedures without outer contention. Median and range values of viscoelastic parameters related to one single suction cycle according to the two recording modalities for the two probe aperture diameters (AD 2 mm and AD 4 mm) on normal skin and connective tissue disorders.

Skin conditionMD (mm)RD (mm)
AD 2 mmAD 4 mmAD 2 mmAD 4 mm

Steep modality
 Normal0.280.420.080.14
(0.12–0.33)(0.21–0.59)(0.05–0.16)(0.08–0.22)
 Scleroderma0.190.25**0.100.13
(0.09–0.24)(0.10–0.37)(0.04–0.18)(0.08–0.18)
 Ehlers-Danlos0.300.61*0.030.09*
(0.15–0.50)(0.27–0.89)(0.02–0.12)(0.04–0.13)
 Dermatoporosis0.240.450.130.31*
(0.14–0.35)(0.18–0.63)(0.09–0.23)(0.16–0.35)
 Acromegaly0.180.300.100.20
(0.09–0.30)(0.23–0.39)(0.08–0.20)(0.13–0.17)

Progressive modality
 Normal0.210.380.070.09
(0.11–0.26)(0.28–0.51)(0.05–0.12)(0.07–0.13)
 Scleroderma0.170.23**0.130.17*
(0.05–0.21)(0.10–0.31)(0.07–0.18)(0.06–0.20)
 Ehlers-Danlos0.260.470.060.05**
(0.08–0.43)(0.39–0.87)(0.04–0.15)(0.03–0.08)
 Dermatoporosis0.240.420.21*0.34*
(0.16–0.43)(0.32–0.65)(0.15–0.33)(0.24–0.36)
 Acromegaly0.190.27**0.16*0.21*
(0.13–0.31)(0.17–0.40)(0.14–0.22)(0.12–0.24)

Compared to normal; P < 0.05; P < 0.01.