Research Article
A New Analytical Q-Absorbance Ratio Method Development and Validation for Simultaneous Estimation of Lamivudine and Isoniazid
Table 4
Summary of regression characteristics and validation parameters.
| Parameters | LAM | INH | 246 nm | 272 nm | 246 nm | 272 nm |
| Beer’s law limit (g/mL) | 5–30 | 5–30 | 5–30 | 5–30 | Absorptivity | 287 | 435 | 288 | 276 | Regression equation () | | | | | Slope () Intercept () | 0.0279 0.0079 | 0.0405 0.0191 | 0.0267 0.0118 | 0.0266 0.0065 | Correlation coefficient () | 0.9994 | 0.9989 | 0.999 | 0.999 | Standard deviation (SD) | 0.0009 | 0.0023 | 0.0006 | 0.0017 | LOD (g/mL) | 0.106 | 0.186 | 0.078 | 0.211 | LOQ (g/mL) | 0.321 | 0.563 | 0.238 | 0.639 | Sandell’s sensitivity (g/cm2) | 0.0347 | 0.0229 | 0.0348 | 0.0356 |
|
|