Nonconjugated Polyelectrolyte as Efficient Fluorescence Quencher and Their Applications as Biosensors: Polymer-Polymer Interaction

Table 1

Quantitative data of CRP in serum—comparison of the methods.

Sample codeAmount of CRP nephelometry (mg/L)Amount of CRP fluorescence (mg/L)

14.584.73 ± 0.15
215.0414.44 ± 0.09
314.1114.25 ± 0.50
435.0135.63 ± 0.10
528.6630.17 ± 0.21
615.0715.42 ± 0.23
712.8412.76 ± 0.45
87.157.34 ± 0.12
912.0013.45 ± 0.14
10 8.237.79 ± 0.05
118.768.76 ± 0.55
128.448.56 ± 0.09
1333.4534.10 ± 0.11
1418.2618.47 ± 0.41
157.067.45 ± 0.21
169.589.19 ± 0.10
1714.1214.21 ± 0.90
188.569.87 ± 0.15
1910.0710.98 ± 0.22
2012.0412.14 ± 0.55
214.824.97 ± 0.12
228.978.79 ± 0.15
2317.7217.64 ± 0.24
247.947.64 ± 0.11
254.954.25 ± 0.15
2610.1510.45 ± 0.20
2716.2916.45 ± 0.11
2814.3614.12 ± 0.16
2912.4812.24 ± 0.15
3060.4162.45 ± 0.24
3112.9712.21 ± 0.14
3217.916.73 ± 0.09
33Negative sampleMarginal change in fluorescence
34Negative sampleMarginal change in fluorescence
35Negative sampleMarginal change in fluorescence
36Negative sampleMarginal change in fluorescence
37Negative sampleMarginal change in fluorescence
38Negative sampleMarginal change in fluorescence
39Negative sampleMarginal change in fluorescence
40Negative sampleMarginal change in fluorescence