Research Article

Evaluation of the Utility of Thromboelastography in a Tertiary Trauma Care Centre

Table 1

The derived reference range of thromboelastogram parameters and the reference range of thromboelastogram parameters predefined by the manufacturers.

ParametersPredefined reference range by the manufacturersDerived reference range volunteers

-time (min)5–111.8–14.2
-time (min)1.5–4.50.7–7.3
angle (°)44.5–6627.3–72.3
Maximum amplitude (mm)53–7032.1–87.9