Research Article

Gradient Artefact Correction and Evaluation of the EEG Recorded Simultaneously with fMRI Data Using Optimised Moving-Average

Table 2

Median attenuation in the artefact frequency bins.

EEG data set
Philips dataGE data
Frequency Bin (Hz)OMAAASFrequency bin (Hz)OMAAAS
Attenuation (dB)Attenuation (dB)Attenuation (dB)Attenuation (dB)

2045.375157.526314.0116.961316.1357
4078.957091.808828.0133.673032.4054
6085.600096.020142.0260.065655.6886
8098.2415111.014356.0262.990761.6082
10098.3248110.720770.0347.206540.3737
120110.7477121.144484.0367.697163.3023
140117.0230126.720898.0470.266564.4877
160104.1135110.7799112.0477.209471.3954
180112.6177119.6797126.0576.837964.6367
200120.2490126.1687140.0671.688361.2908
220122.7164127.8049154.0692.428582.1534
240120.7178123.6904168.0773.617957.6112
260122.9039125.0317182.07103.617377.8464
280116.5943117.3843196.08104.840082.8587
300130.2747129.5068210.08111.468181.6893
320141.4297139.5853224.09111.892280.3991
340140.0759136.2834238.10127.104778.2103
360142.5260138.4374252.10143.551685.3269
380144.0613137.4788266.11149.799778.9997
400140.8645132.8269280.11153.859779.6079
420137.6477128.4942294.12174.555984.2324
440142.1547131.7775308.12163.747577.0851
460142.6946131.4058322.13173.499179.8261
480141.9741130.2354336.13173.037182.8779
500131.2361117.6283350.14179.399581.8668
364.15176.036081.8207
378.15160.962779.0052
392.16181.955480.8334
406.16163.759177.9195
420.17175.064379.6067
434.17175.968079.2014
448.18171.180578.3351
462.18171.709178.1379
476.19153.688377.2672
490.20171.556877.0958