Research Article

Validation of a Diagnostic Microarray for Human Papillomavirus: Coverage of 102 Genotypes

Table 1

Comparison of array performance to blot. values: Fisher’s exact test.

TestBlotArray ( )Array ( )
SequencePositiveNegativePositiveNegativePositiveNegative

Test result +5552631151122163
Test result −802452949010456962
Sensitivity.407.401.713
Specificity.979.987.763
value<.0001<.0001<.0001