Research Article

Novel HPLC Analysis of Hydrocortisone in Conventional and Controlled-Release Pharmaceutical Preparations

Table 2

Accuracy of the HPLC method showing the mean percentage recovery of hydrocortisone at three concentration levels.

Concentration levelsAmount taken (mg/ml)Amount added (mg/ml)Amount recoveredPercentage recovery (%)

80%0.100.080.0782097.75
0.100.080.0784098.01
0.100.080.0783697.92
100%0.100.100.10073100.73
0.100.100.10120101.20
0.100.100.10052100.52
120%0.100.120.1199899.98
0.100.120.12062100.05
0.100.120.1193499.45