Research Article

Novel HPLC Analysis of Hydrocortisone in Conventional and Controlled-Release Pharmaceutical Preparations

Table 4

Interday precision of proposed HPLC method showing the mean percentage recovery and RSD of hydrocortisone on different days.

Number of runsPercentage (%)
Day 1Day 2Day 3

1100.5298.52100.16
299.4599.6799.51
3100.0398.3699.72
Mean100.0098.8599.80
% RSD0.540.710.33

of three replicate determinations.