Accuracy of Dose Delivery in Multiple Breath-Hold Segmented Volumetric Modulated Arc Therapy: A Static Phantom Study

Table 1

The monitor units per fraction, the gantry arc angles, and the beam-on-times required for noninterrupted delivery are shown for five VMAT clinical cases. When the beams were interrupted for five seconds with varied beam-on-times between interrupts, the number of VMAT segments was counted and recorded.

CaseMU/fx Gantry arc angle (degree)Beam-on-time (s)Measured number of VMAT segments
Beam-on-time between interrupts (s)
1015203040

1514.419521029171286
2390.716097148643
3588.52301372112964
4786.336021432181386
5544.236019830171286