Monte Carlo Simulations of a Scanning System Based on a Panoramic X-Ray Tube with a Conical Anode

1University of Bergen, Allegaten 55, P.O. Box 7803, 5020 Bergen, Norway
2Christian Michelsen Research AS, Fantoftveien 38, P.O. Box 6031, 5892 Bergen, Norway
3Visuray AS, Strandbakken 10, 4070 Randaberg, Norway

Received 9 May 2014; Accepted 30 July 2014; Published 21 August 2014

Academic Editor: Zhiqiang Mao

Copyright © 2014 Andrii Sofiienko et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Monte Carlo simulations were used to study photon production in a panoramic X-ray tube with a conical tungsten target to determine the optimal characteristics of the target shape and electron beam configuration. Several simulations were performed for accelerating potentials equal to 250 kV, 300 kV, and 500 kV with electron beams of various radii and anode sizes. The angular distribution of the photon intensity was analysed by numerical calculations for an assembly composed of an X-ray tube and an external collimator with a cylindrical hole to simulate a panoramic scanning system with an X-ray pencil beam.