A Cadaveric Comparative Study on the Surgical Accuracy of Freehand, Computer Navigation, and Patient-Specific Instruments in Joint-Preserving Bone Tumor Resections

Table 1

Location accuracy (56 entry and 56 exit cuts) and mean bone cutting time for each technique.

ParametersFreehandCASPSICAS + PSI

Mean location accuracy (mm) (95% CI)9.2 ± 3.3 (8.0; 10.3)3.6 ± 2.1 (2.5; 4.8)1.9 ± 1.1 (0.8; 3.0)2.0 ± 1.0 (0.9; 3.1)

Plane directionEntry cuts (mm)4.3 ± 2.72.1 ± 1.31.2 ± 0.90.8 ± 0.5
Exit cuts (mm)6.8 ± 3.43.1 ± 2.22.3 ± 1.51.6 ± 0.8

Mean bone resection time (minutes)6.8 ± 1.117.0 ± 2.44.8 ± 1.012.8 ± 3.8

Standard deviation is listed for each value. 95% CI is given for the LA.