Research Article
Performance of LED Fluorescence Microscopy for the Diagnosis of Pulmonary Tuberculosis in HIV Positive Individuals in Addis Ababa, Ethiopia
Table 3
Performance of smear microscopy using Ziehl-Neelsen and LED-FM in TB suspected HIV positive individuals per patient analysis.
| | ZN | Kappa value | LED-FM | Kappa value | | Direct | Concentrated | Direct | Concentrated |
| Sensitivity % (90% CI) | 29.2 (26.9–32.1) | 45.8 (41.2–50.4) | 0.767 | 62.5 (56.25–68.75) | 62.5 (56.25–68.75) | 0.743 | Positive (%) | 7 (3.9) | 11 (6.2) | | 15 (8.4) | 15 (8.4) | | Specificity % (90% CI) | 100 (90–100) | 100 (90–100) | | 100 (90–100) | 100 (90–100) | | PPV % | 100 | 100 | | 100 | 100 | | NPV % | 90.1 | 92.2 | | 94.5 | 94.5 | |
|
|