High-Definition X-Ray Fluorescence: Applications
Table 5
Reference and measured values for three polyethylene standards. The experimental uncertainty (at 95% level) includes estimated systematic uncertainty plus Poisson counting statistical error.
Sample
Cr
Hg
Pb
Br
Error
PE_5398_N
Reference
1001
1000
1000
500
4% (95% level)
Measured
1020
950
981
442
10% (95% level)
PE_H_18A
Reference
1000
1100
1200
1100
4% (95% level)
Measured
1020
1150
1220
1090
10% (95% level)
PE_L
Reference
400
200
400
500
4% (95% level)
Measured
401
185
419
508
10% (95% level)