Research Article

Is Subcutaneous Route an Alternative to Intravenous Route for Mouse Contrast-Enhanced Magnetic Resonance Imaging at 1.5 T?

Table 1

Acquisition parameters for dynamic CE-MRI imaging.

ParametersValues

Sequence typeSpoiled gradient recalled echo-2D
TR/TE/FA25ms / 4ms / 90°
Slice thickness1.75 mm
OrientationCoronal
FOV dimensions (read x phase)80 x 40 mm2
Reconstructed matrix size128 x 64
Reconstructed pixel size0.625 x 0.625 mm2
Readout BW30 kHz
Nex6
Image acquisition time4.8 s
Total image number200
Acquisition rate for images 1-5012.5 images per min
Acquisition rate for image 51-2001 image per min
Total acquisition duration2h 34min