Capacitive versus Overlap Decoupling of Adjacent Radio Frequency Phased Array Coil Elements: An Imaging Robustness Comparison When Sample Load Varies for 3 Tesla MRI Table 1 Max (mean) g-factor values (see Figures
5(a) and
5(b) ).
Anterior-posterior Left-right 0.3 S/m ×2 ×3 ×4 ×5 ×2 ×3 ×4 ×5 Simulated CD 1.47 2.96 11.68 65.30 4.43 9.59 37.80 7142.59 (1.15) (1.78) (3.92) (13.74) (1.30) (2.37) (6.76) (43.88) OD 1.55 3.09 11.72 78.04 5.15 11.61 43.39 2775.42 (1.17) (1.87) (4.30) (16.51) (1.36) (2.58) (7.39) (45.69) Measured CD 1.66 3.55 15.14 847.96 5.82 10.80 79.26 2256.00 (1.23) (2.00) (4.21) (15.25) (1.30) (2.49) (6.14) (18.01) OD 1.79 4.09 17.04 506.96 6.34 12.79 82.29 4480.72 (1.28) (2.14) (4.80) (18.21) (1.32) (2.64) (6.62) (22.08) 0.6 S/m ×2 ×3 ×4 ×5 ×2 ×3 ×4 ×5 Simulated CD 1.50 3.29 10.88 60.83 2.28 7.16 31.21 1867.61 (1.14) (1.73) (3.67) (12.10) (1.16) (1.92) (5.46) (39.40) OD 1.50 3.40 11.21 68.30 2.65 8.37 38.72 2665.41 (1.16) (1.83) (4.05) (14.64) (1.21) (2.11) (6.20) (44.39) Measured CD 1.90 4.27 15.05 191.98 2.54 6.48 44.91 2354.87 (1.23) (2.02) (4.13) (14.70) (1.19) (2.07) (5.63) (20.42) OD 2.04 4.81 17.83 183.64 2.57 7.24 53.77 4401.27 (1.28) (2.15) (4.64) (16.64) (1.21) (2.17) (6.24) (30.10) 0.9 S/m ×2 ×3 ×4 ×5 ×2 ×3 ×4 ×5 Simulated CD 1.66 3.94 11.97 59.15 1.73 6.28 29.63 992.37 (1.14) (1.77) (3.73) (11.51) (1.11) (1.76) (4.98) (31.86) OD 1.64 3.93 12.22 61.17 1.80 6.98 34.27 925.98 (1.16) (1.84) (3.98) (13.32) (1.13) (1.86) (5.40) (32.28) Measured CD 2.14 4.86 13.54 167.99 1.89 5.71 67.80 3620.02 (1.23) (2.00) (4.06) (13.04) (1.12) (1.90) (5.37) (24.52) OD 2.25 5.25 16.65 177.03 1.93 6.29 71.22 4881.34 (1.29) (2.13) (4.48) (14.87) (1.14) (1.99) (5.80) (30.53)