Capacitive versus Overlap Decoupling of Adjacent Radio Frequency Phased Array Coil Elements: An Imaging Robustness Comparison When Sample Load Varies for 3 Tesla MRI

Table 2

Simulated (S) versus measured (M) S-parameters of the 5-ch CD PA coil and the 5-ch OD PA coil.

El#2345

S/m0.3 (0.6) 0.90.3 (0.6) 0.90.3 (0.6) 0.90.3 (0.6) 0.9
1 S-CD−14.3 (−13.4) −13.4−5.30 (−7.90) −9.70−12.1 (−16.0) −18.8−12.9 (−17.6) −21.5
S-OD−12.1 (−10.5) −10.3−4.10 (−6.00) −7.50−10.1 (−13.3) −15.6−10.8 (−15.1) −18.6
M-CD−16.8 (−15.3) −14.4−6.80 (−9.30) −10.7−13.3 (−17.6) -20.0−15.0 (−19.4) −21.9
M-OD−11.2 (−11.4) −9.70−5.70 (−7.80) −9.00−12.7 (−17.5) −20.1−9.10 (−13.5) −15.8

2 S-CD−14.5 (−13.6) −13.5−5.50 (−8.10) −10.0−12.1 (−16.0) −18.8
S-OD−12.5 (−10.6) −10.2−4.00 (−5.90) −7.30−10.0 (−13.3) −15.6
M-CD−20.4 (−17.0) −16.6−7.70 (−10.4) −12.0−15.0 (−18.1) −20.6
M-OD−14.7 (−12.3) −11.9−5.80 (−8.60) −10.2−12.4 (−16.6) −19.1

3 S-CD−14.5 (−13.6) −13.5−5.40 (−7.90) −9.70
S-OD−12.6 (−10.6) −10.2−4.00 (−6.00) −7.50
M-CD−17.0 (−16.0) 15.7−7.80 (−9.70) −10.9
M-OD−15.2 (−13.0) −12.1−5.60 (−7.70) −8.60

4 S-CD−14.2 (−13.4) −13.4
S-OD−12.2 (−10.5) −10.3
M-CD−16.7 (−15.4) −14.0
M-OD−11.0 (−11.8) −9.60