Research Article

Quasistatic Solutions versus Full-Wave Solutions of Single-Channel Circular RF Receive Coils on Phantoms of Varying Conductivities at 3 Tesla

Table 1

Total noise (Ohms).

Circular loop diameter3 cm4 cm5 cm6 cm7 cm8 cm9 cm10 cm

Quasistatic (Q) calculations
0.3 S/m1.511.772.212.883.825.056.638.56
0.6 S/m1.652.163.024.336.188.6311.815.6
0.9 S/m1.792.543.825.788.5412.216.922.6

Q/B ratios
0.3 S/m3.021.360.840.610.460.390.310.27
0.6 S/m2.581.581.090.890.780.730.670.65
0.9 S/m2.491.591.171.070.990.950.890.92

Full-wave (FW) calculations
0.3 S/m0.481.132.394.527.9913.220.931.9
0.6 S/m0.531.212.494.487.5211.817.825.6
0.9 S/m0.621.412.865.048.2612.618.525.8

FW/B ratios
0.3 S/m0.960.870.910.970.971.010.991.00
0.6 S/m0.830.880.90.920.950.991.011.07
0.9 S/m0.860.880.870.930.960.980.971.04

Benchtop (B) measurements
0.3 S/m0.501.302.634.678.2613.121.231.9
0.6 S/m0.641.372.764.867.9311.917.724.0
0.9 S/m0.721.603.275.428.6012.819.024.7